ES2 SPINAL SYSTEM 48280310Recalls & reported problems

Stryker Corporation · POLYAXIAL SCREWDRIVER · Class 1

All models of ES2 SPINAL SYSTEM (54)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04546540710475), received between Apr 4, 2017 and Jun 6, 2023 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Residue After Decontamination” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Residue After Decontamination2
  • Adverse Event Without Identified Device or Use Problem1
  • Defective Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Information1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 6, 2023Residue After Decontamination; Residue After Decontamination

    A COMPANY REPRESENTATIVE REPORTED THAT A PROCEDURE WAS NOT ABLE TO BE COMPLETED AS PLANNED DUE TO RESIDUE FOUND IN THE SHAFTS OF TWO ES2 POLYAXIAL SCREWDRIVERS. THE PATIENT WAS CLOSED AND PRESCRIBED ANTIBIOTICS TO TREAT ANY POTENTIAL INFECTION. THE PATIENT WILL BE RESCHEDULED FOR A FUTURE SURGERY. THIS REPORT IS FOR THE SECOND OF TWO ES2 POLYAXIAL SCREWDRIVERS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17070711 ↗
  • InjuryJun 6, 2023Residue After Decontamination; Residue After Decontamination

    A COMPANY REPRESENTATIVE REPORTED THAT A PROCEDURE WAS NOT ABLE TO BE COMPLETED AS PLANNED DUE TO RESIDUE FOUND IN THE SHAFTS OF TWO ES2 POLYAXIAL SCREWDRIVERS. THE PATIENT WAS CLOSED AND PRESCRIBED ANTIBIOTICS TO TREAT ANY POTENTIAL INFECTION. THE PATIENT WILL BE RESCHEDULED FOR A FUTURE SURGERY. THIS REPORT IS FOR THE FIRST OF TWO ES2 POLYAXIAL SCREWDRIVERS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17070710 ↗
  • MalfunctionApr 4, 2017Defective Component; Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT; LOCKING MECHANISM OF THE SCREWDRIVER DEFECTIVE. THEREFORE THE SCREW WAS INSERTED WITH HIGH PRESSURE ON THE SCREWDRIVER. SCREW LOOSENED AND MUST BE REPLACED DURING MEDICAL PROCEDURE. ANOTHER SCREW WAS AVAILABLE.

    Patient problems: No Information; No Known Impact Or Consequence To Patient

    MAUDE report 6458251 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →