TARGET XL M0036121650Recalls & reported problems
Stryker Corporation · 360 Soft Detachable Coil · Class 2 · Implant
All models of TARGET XL (34)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04546540711557), received between Mar 20, 2019 and May 6, 2026 — classified as 2 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Premature Separation” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Premature Separation4
- Fracture2
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 6, 2026Fracture
THE COIL (SUBJECT DEVICE) WAS RETURNED FOR ANALYSIS, AND IT WAS DISCOVERED THAT THE MAIN COIL (SUBJECT DEVICE) WAS SEEN TO BE FRACTURED NEAR THE DETACHMENT ZONE. THE SUBJECT DEVICE WAS REPLACED, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25096170 ↗MalfunctionFeb 13, 2026Premature Separation
IT WAS REPORTED THAT DURING THE PROCEDURE THE SUBJECT COIL WAS INITIALLY ADVANCED INTO THE MICROCATHETER WHILE THE COIL'S SHEATH WASN'T FULLY SEATED AGAINST THE HUB. A FEW LOOPS HAD ALREADY BEEN DEPLOYED IN THE MICROCATHETER. THE SUBJECT COIL WAS CAREFULLY PULLED BACK INTO THE SHEATH. THE SUBJECT COIL WAS REINSERTED INTO THE MICROCATHETER AND ADVANCED. THE SUBJECT COIL WAS VISIBLE UNDER FLUOROSCOPY, BUT IT WAS NOT OUT OF MICROCATHETER. THE SUBJECT COIL WAS FOUND TO BE DETACHED. THE SUBJECT COIL WAS REPLACED, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24338434 ↗InjuryNov 6, 2024Premature Separation; Premature Separation
IT WAS REPORTED DURING THE PROCEDURE THE SUBJECT COIL PREMATURELY DETACHED FROM THE PUSHER WIRE INSIDE THE CATHETER. THE COIL WAS REMOVED FROM THE CATHETER WITH A MICRO SNARE. THERE WAS A SURGICAL DELAY OF UNKNOWN TIME. THE PATIENT WAS NOT AFFECTED. THE DEVICE WAS REPLACED AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20623649 ↗InjuryMay 10, 2024Premature Separation; Premature Separation
IT WAS REPORTED THAT DURING THE TREATMENT OF RIGHT CAROTID-CAVERNOUS FISTULA, THE OPERATOR USED THE MICROCATHETER AND THE GUIDEWIRE TO NAVIGATE TO THE INTERIOR OF THE FISTULOUS ORIFICE. DURING THE POSITIONING OF THE SUBJECT COIL, IT DETACHED PREMATURELY AND THE STENT RETRIEVER WAS USED TO CAPTURE IT. A 30 MINUTES SURGICAL DELAY WAS REPORTED DUE TO THIS EVENT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY, NO FURTHER INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19294350 ↗MalfunctionJan 21, 2022Fracture; Fracture
THE SUBJECT COIL WAS RETURNED FOR ANALYSIS AND IT WAS DISCOVERED THAT THE COIL DELIVERY WIRE BROKEN/FRACTURED DURING USE. THERE WAS NO CLINICAL CONSEQUENCE TO THE PATIENT REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13312408 ↗MalfunctionMar 20, 2019Premature Separation; Premature Separation
IT WAS REPORTED THAT THE SUBJECT COIL REDEPLOYED IN THE CATHETER . NO CLINICAL CONSEQUENCES REPORTED TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8438485 ↗
Other DEVICE, NEUROVASCULAR EMBOLIZATION devices
- POD RBYPODJ60-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION355 reports
- Penumbra System SCH1Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION295 reports
- RUBY Coil RBY4C0860-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION276 reports
- Smart Coil 400SMTHXSFT0208Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION165 reports
- Penumbra System RH1-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION154 reports
- Optima Coil System OPTI0208CSS10Balt USA LLC · Device, Neurovascular Embolization3 recalls3 reports
- POD RBYPOD6-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION148 reports
- POD RBYPODJ30-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION132 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page