AXSOS 702748Recalls & reported problems
Stryker GmbH · Screwdriver · Class 1
All models of AXSOS (184)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04546540713711), received between Aug 18, 2015 and May 9, 2025 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Fracture3
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Known Impact Or Consequence To Patient2
- Insufficient Information1
- No Information1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 9, 2025Break; Fracture; Break; Fracture
AS REPORTED: "WE HAVE HAD AN INCIDENT WHERE TWO SCREWDRIVERS/HEADS HAVE SHEARED OFF DURING USE."
Patient problems: Insufficient Information
MAUDE report 21995486 ↗MalfunctionMar 26, 2021Fracture; Fracture
"THE CUSTOMER REPORTED THAT DURING A FEMORAL PLATING SURGERY, TWO SCREWDRIVERS BROKE INTRA OP WHILE ATTEMPTING TO EXTRACT A LOCKING SCREW PREVIOUSLY INSERTED TO CHANGE HIS POSITION. ADDITIONAL INFORMATION: THE BROKEN PIECES WERE RETRIEVED FROM THE PATIENT. THE LOCKING SCREW INVOLVED IN THE EVENT HAS SUCCESSFULLY BEEN IMPLANTED. THE SURGEON DID NOT APPLIED PRESSURE AND DID NOT WORKED AT AN EXTREME ANGLE."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11575723 ↗MalfunctionAug 12, 2020Fracture; Fracture
AS REPORTED: " THE DOCTOR WAS USING A T20 SCREWDRIVER (703539) TO REMOVE A LOCKING SCREW AND REPLACE IT WITH A DIFFERENT LENGTH. THE SCREW WAS TIGHT, AS THE DOCTOR APPLIED MORE TORQUE TO THE SCREWDRIVER, THE TIP OF THE BLADE SNAPPED OFF. WE RECOVERED THE BROKEN PIECE AND NOTHING ADDITIONAL REMAINED IN THE PATIENT. ADDITIONALLY REPORTED FROM SALES REP, THE ACTUAL ITEM NUMBER FOR THE SCREWDRIVER IS 702748 WITH LOT N00517.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 10399430 ↗MalfunctionSep 10, 2015Break
IT WAS REPORTED THAT THE PRODUCT WAS RETURNED TO STRYKER (B)(4) WITH A NOTE FROM THE HOSPITAL ACCOUNT "BROKEN - TIP IN PATIENT". NO ADDITIONAL INFORMATION WAS PROVIDE BY THE CUSTOMER. FOLLOW UP WITH THE STRYKER SALES REP INDICATED THAT NO PRODUCT COMPLAINT/ISSUE HAD BEEN LOGGED BY THE CUSTOMER. NO INFORMATION ABOUT THE EVENT IS AVAILABLE.
Patient problems: No Information
MAUDE report 5068710 ↗MalfunctionAug 18, 2015Break
SCREWDRIVER WAS BROKEN DURING A SURGICAL PROCEDURE TO EXTRACT AXSOS 5.0. THIS OPERATION WAS THE FIRST WITH THIS INSTRUMENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 5012018 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page