FUJIFILM EC-600LSRecalls & reported problems
Fujifilm Corporation · The FUJIFILM Endoscope Models EC-600LS is intended for the visualization of the lower digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the rectum and large intestine. · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04547410330229), received between Nov 15, 2023 and Dec 25, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break1
- Device Reprocessing Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 25, 2025Device Reprocessing Problem; Device Reprocessing Problem
IT WAS REPORTED THAT THE SCOPE WAS SWABBED FOR STANDARD MICRO TESTING ON 10/28/2025. CULTURE RETURNED POSITIVE 10/31/2025. SCOPE WAS REPROCESSED USING AER ON 10/31/2025 AND REMOVED FROM SERVICE. THE SCOPE WAS CULTURE TESTED AS PART OF THEIR MONTHLY TESTING STANCE, AND IT FAILED. THEY PUT IT IN THE AER AGAIN AND CULTURE TESTED AGAIN, THIS TIME IT PASSED. THE SCOPE WAS SENT IN FOR EVALUATION. THE AREA THAT FAILED THE INITIAL TEST WAS THE DISTAL TIP. THE FACILITY STAFF (REGISTERED NURSE) CONFIRMED THE SCOPE IN QUESTION WAS LAST USED ON A PATIENT (B)(6) 2025. NO MALFUNCTION OCCURRED WHEN THE SCOPE IN QUESTION WAS USED ON A PATIENT. THE SCOPE WAS CULTURE TESTED ON 9/30/2025 AND THE RESULTS WERE NEGATIVE. THE SCOPE WAS SWABBED FOR STANDARD MICRO TESTING ON 10/28/2025. CULTURE RETURNED POSITIVE 10/31/2025. THE SCOPE WAS RE-CULTURED ON 10/31/2025 AND 11/3/2025 WITH NEGATIVE RESULTS. AFTER THE SCOPE WAS MANUALLY CLEANED THROUGH THEIR AER UNIT AND PLACED IN STORAGE, THERE WERE NO ISSUES FOUND AFTER THE CLEANING PROCESS. CUSTOMER HAS NOT CONFIRMED THE POSITIVE CULTURE ORGANISM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23891915 ↗MalfunctionNov 15, 2023Break; Break
ON (B)(6) 2023, FUJIFILM HEALTHCARE AMERICAS CORPORATION WAS INFORMED OF AN EVENT INVOLVING WITH EC-600LS. IT WAS REPORTED THAT THE KNOB BROKE DURING THE PROCEDURE PRIOR TO REACHING CECUM AND THE KNOB WAS UNABLE TO ROTATE. THERE WAS A DELAY TO SWITCH TO A DIFFERENT SCOPE. THERE WAS NO INJURY TO THE PATIENT REPORTED. AS SUCH, THIS REPORT IS BEING SUBMITTED IN AN ABUNDANCE OF CAUTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18139948 ↗
Other Colonoscope and accessories, flexible/rigid devices
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS CF-HQ190LOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls2 deaths6,804 reports
- EVIS EXERA III XENON LIGHT SOURCE OLYMPUS CLV-190Olympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid6,617 reports
- EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-HQ190Olympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls5 deaths6,244 reports
- EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-H190Olympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid2 recalls7 deaths5,572 reports
- EVIS EXERA III COLONOVIDEOSCOPE OLYMPUS PCF-H190DLOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls2,830 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS PCF-H190LOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls2 deaths2,513 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS CF-H190IOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid1 recall1,967 reports
- PENTAX Medical EC38-i10cLHoya Corporation · Colonoscope and accessories, flexible/rigid1,618 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page