FUJIFILM EC-760R-V/LRecalls & reported problems
Fujifilm Corporation · This device is intended for the visualization of the lower digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the rectum and large intestine. · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Fujifilm colonoscope is used to look inside the rectum and large intestine. Reports mostly describe control knobs breaking, suction or water flow problems, and scratches or small tears to the bowel lining. The records include 25 reports that mention this identifier, with 12 injuries and no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 26
- Deaths reported
- 0
- Injuries reported
- 13
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 26 adverse event reports that list this exact device (DI 04547410358681), received between Sep 13, 2019 and Sep 16, 2026 — classified as 13 injuries, 13 malfunctions.
The most frequently reported specific device problem is “Sharp Edges” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Sharp Edges6
- Adverse Event Without Identified Device or Use Problem3
- Improper or Incorrect Procedure or Method3
- Partial Blockage3
- Appropriate Device Problem Term/Code Not Available1
- Blocked Connection1
- Break1
- Device Handling Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions12
- Abrasion8
- Laceration(s)2
- Perforation2
- Bowel Perforation1
- Intraoperative Pain1
Recent adverse event reports
The 25 most recent of 26 reports, in the reporter's own words as published by the FDA.
InjurySep 16, 2026Sharp Edges
On 09/02/2026, FUJIFILM Healthcare Americas Corporation became aware of an event involving EC-760R-V/L.\\It was reported that the distal end is rough, resulting in a small cut in the patient's colon during a diagnostic procedure. \\FUJIFILM Healthcare Americas Corporation is reporting this event, due to cut on colon.\\Ref\FUJIFILM Healthcare Americas Corporation Complaint # (b)(4).
Patient problems: Laceration(s)
MAUDE report 50796662 ↗InjuryJul 23, 2025Material Deformation; Material Deformation
ON JUNE 25TH 2025, FUJIFILM CORPORATION WAS INFORMED OF AN EVENT INVOLVING EC-760R-V/L VIA MEDSUN REPORT, (B)(4) FROM FDA. IT WAS REPORTED THAT, ON (B)(6) 2025, DURING A COLONOSCOPY SCREENING FOR COLORECTAL CANCER, THE PHYSICIAN REALIZED THAT THE BIG WHEEL PART OF THE SCOPE HAD BROKEN MIDWAY THROUGH THE PROCEDURE. THE PHYSICIAN WAS STILL ABLE TO VISUALIZE AND WITHDREW THE SCOPE. THE SCOPEWAS THEN USED TO RETROFLEX IN THE RECTUM, AND AFTER THE RECTUM WAS VISUALIZED, A PERFORATION WAS NOTED. THE SCOPE WAS THEN CHANGED OUT TO AN EGD SCOPE. THE ANGULATION CONTROL KNOB IS BROKEN ON THE DEVICE.
Patient problems: Bowel Perforation
MAUDE report 22583983 ↗MalfunctionJul 3, 2025Partial Blockage; Partial Blockage
IT WAS REPORTED THAT THE JET WASH CHANNEL IS TOO STRONG AND IT DAMAGED TISSUE. THE PATIENT REMAINED STABLE, WAS DISCHARGED, AND FOLLOW-UP CONFIRMED FULL RECOVERY. ISSUE RESOLVED WHEN THE FLOW OF THE ENDO GATOR DESCREASED BY THE CUSTOMER.
Patient problems: Abrasion
MAUDE report 22393716 ↗MalfunctionApr 23, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT THE BLACK STUFF WAS COMING OUT OF THE SCOPE AFTER THE PROCEDURE. AT THE END OF THE CASE, THE TECH OF THE FACILITY SAID THE SCOPE MADE A "WEIRD" NOISE WHEN FLUSHING THE SCOPE. IT WAS THEN TAKEN TO THE SCOPE CLEANING ROOM AND DURING THE BRUSHING OF THE SUCTION CHAMBER, SMALL BLACK PARTICLES CAME OUT OF THE SCOPE. THEY FELT LIKE HARD PLASTIC PIECES OF WHAT THE SCOPE IS MADE OUT OF.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21888060 ↗MalfunctionJun 20, 2024Increase in Pressure; Increase in Pressure
ON MAY 20TH FUJIFILM HEALTHCARE AMERICAS CORPORATION WAS INFORMED OF AN EVENT INVOLVING EC-760R-V/L. IT WAS REPORTED THAT THE DEVICE IS PUSHING TOO MUCH FLUID AND HURTING THE PATIENT. THE PATIENT WAS HURTING HIS COLON/STOMACH. PATIENT IS STABLE AND HAD NO FURTHER COMPLICATIONS POST PROCEDURE. NO MEDICAL INTERVENTION WAS NEEDED. THE SITE WAS NOT ABLE TO RETRIEVE THE POLYP. THE MUCOSAL LINING WAS DAMAGED, SO NO CLEAR BIOPSY COULD BE TAKEN. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: Intraoperative Pain
MAUDE report 19576065 ↗MalfunctionApr 26, 2024Gas/Air Leak; Gas/Air Leak
ON MARCH 29TH 2024 FUJIFILM HEALTHCARE AMERICAS CORPORATION WAS INFORMED OF AN EVENT INVOLVING EC-760R-V/L. IT WAS REPORTED THAT LEAKS WERE IDENTIFIED AT THE CO2 INSUFFLATION CONNECTION POINT AND WOULD NOT ALLOW FOR BOWEL INSUFFLATION. AS SUCH, THIS REPORT IS BEING SUBMITTED IN AN ABUNDANCE OF CAUTION. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19194107 ↗MalfunctionApr 26, 2024Break; Material Fragmentation; Break; Material Fragmentation
ON MARCH 28TH 2024 FUJIFILM HEALTHCARE AMERICAS CORPORATION WAS INFORMED OF AN EVENT INVOLVING EC-760R-V/L. IT WAS REPORTED THAT THE ANGULATION KNOB SNAPPED DURING A DIAGNOSTIC. THERE WAS NO HARM OR INJURY TO THE PATIENT. AS SUCH, THIS REPORT IS BEING SUBMITTED IN AN ABUNDANCE OF CAUTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19192929 ↗MalfunctionJan 30, 2024Suction Problem; Suction Failure; Suction Problem; Suction Failure
ON JANUARY 3RD 2024 FUJIFILM HEALTHCARE AMERICAS CORPORATION WAS INFORMED OF AN EVENT INVOLVING EC-760R-V/L. IT WAS REPORTED THAT DURING A DIAGNOSTIC PROCEDURE THE SUCTIONS WAS NOT WORKING PROPERLY. THE PROCEDURE WAS PROLONGED DUE THE PHYSICIAN WAS UNABLE TO SUCTION PROPERLY. THERE WAS NO HARM OR INJURY TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18610260 ↗MalfunctionJan 26, 2024Partial Blockage; Partial Blockage
ON (B)(6) 2023 FUJIFILM CORPORATION WAS INFORMED OF AN EVENT INVOLVING EC-760R-V/L. IT WAS REPORTED THAT THE SCOPE IS SPRAYING A DOUBLE WATER STREAM OUT OF THE DISTAL TIP, THERE IS DIFFICULTY PUTTING THE FIRST STEP THROUGH THE WATER CHANNEL AND THE BLUE WATER INDICATOR RING AROUND THE WATER BUTTON NEEDS TO BE REPLACED. THE PATIENT RECEIVED MINOR MICRO TEARING OF THE MUCOSA. NO INTERVENTION OR TREATMENT WAS NEEDED. THERE WAS NO PROCEDURE DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18585952 ↗MalfunctionJan 26, 2024Partial Blockage; Partial Blockage
ON (B)(6), 2023, FUJIFILM CORPORATION WAS INFORMED OF AN EVENT INVOLVING EC-760R-V/L. IT WAS REPORTED THAT THERE IS A CLOG IN THE DISTAL CO2 CHANNEL. CO2 IS NOT COMING OUT AND WAS UNABLE TO BE USED DURING PATIENT CARE. ROOM AIR WAS USED INSTEAD. THE CASE TOOK LONGER WHILE ATTEMPTING TO TROUBLESHOOT, AND CARE DEVIATED FROM THE NORMAL PROTOCOL. THE PATIENT REMAINED SAFE AND THE PROCEDURE WAS COMPLETED. THERE WAS NO PATIENT HARM OR INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18585951 ↗MalfunctionJan 4, 2024Blocked Connection; Device Handling Problem; Blocked Connection; Device Handling Problem
ON DECEMBER 5TH, 2023, FUJIFILM HEALTHCARE AMERICAS CORPORATION WAS INFORMED OF AN EVENT INVOLVING EC-760R-V/L. IT WAS REPORTED THAT THERE IS A CLOG IN THE DISTAL CO2 CHANNEL. CO2 IS NOT COMING OUT AND WAS UNABLE TO BE USED DURING PATIENT CARE. ROOM AIR WAS USED INSTEAD. THE CASE TOOK LONGER WHILE ATTEMPTING TO TROUBLESHOOT, AND CARE DEVIATED FROM THE NORMAL PROTOCOL. THE PATIENT REMAINED SAFE AND THE PROCEDURE WAS COMPLETED. THERE WAS NO PATIENT HARM OR INJURY. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18443760 ↗InjuryDec 29, 2023Free or Unrestricted Flow; Free or Unrestricted Flow
ON (B)(6) 2023 FUJIFILM HEALTHCARE AMERICAS CORPORATION WAS INFORMED OF AN EVENT INVOLVING EC-760R-V/L. IT WAS REPORTED THAT THE SCOPE IS SPRAYING A DOUBLE WATER STREAM OUT OF THE DISTAL TIP, THERE IS DIFFICULTY PUTTING THE FIRST STEP THROUGH THE WATER CHANNEL AND THE BLUE WATER INDICATOR RING AROUND THE WATER BUTTON NEEDS TO BE REPLACED. THE PATIENT RECEIVED MINOR MICRO TEARING OF THE MUCOSA. NO INTERVENTION OR TREATMENT WAS NEEDED. THERE WAS NO PROCEDURE DELAY.
Patient problems: Laceration(s)
MAUDE report 18419341 ↗MalfunctionNov 6, 2023No Apparent Adverse Event; No Apparent Adverse Event
ON (B)(6) 2023, FUJIFILM HEALTHCARE AMERICAS CORPORATION WAS INFORMED OF AN EVENT INVOLVING EC-760R-V/L. IT WAS REPORTED THAT THE CABLE SNAPPED DURING THE PROCEDURE. THE DELAY WAS THAT THE CUSTOMER HAD TO CHANGED THE SCOPE OUT DURING THE PROCEDURE. THERE WAS NO MAJOR DELAY, AND THE PATIENT WAS NOT IMPACTED. THERE WAS NO HARM OR INJURY TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18074458 ↗MalfunctionOct 6, 2023Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
ON SEPTEMBER 8TH, 2023, FUJIFILM HEALTHCARE AMERICAS CORPORATION WAS INFORMED OF AN EVENT INVOLVING EC-760R-V/L. IT WAS REPORTED THAT THE LIGHT SOURCE IS VERY DIM. THE CUSTOMER HAS TROUBLE TO SEE IF IT WAS A SCOPE LIGHT FAILURE OR LIGHT SOURCE FAILURE. IT WAS DETERMINED TO BE THE SCOPE LIGHT FAILURE. THERE IS NO SERIOUS INJURY OR DEATH ASSOCIATED WITH THE EVENT. THEREFORE, THIS REPORT IS BEING SUBMITTED IN ABUNDANCE OF CAUTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17886716 ↗MalfunctionSep 7, 2023Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
ON AUGUST 8, 2023, FUJIFILM HEALTHCARE AMERICAS CORPORATION WAS INFORMED OF AN EVENT INVOLVING EC-760R-V/L. IT WAS REPORTED THAT CUSTOMER REQUESTED ADVICE ON A POTENTIAL INFECTION CONTROL ISSUE. THEY HAD AN INEXPERIENCED SCOPE TECH PUT ¼ - 1/2 CUP OF EMPOWER ENZYMATIC SOLUTION INTO ONE OF THEIR MEDIVATORS DSD WASHERS WHERE THE OPA WAS SUPPOSED TO GO, AND TWO OF THE SCOPES THAT WERE RUN THROUGH THERE WERE USED ON PATIENTS BEFORE THEY CAUGHT THE MISTAKE. THERE IS NO SERIOUS INJURY OR DEATH ASSOCIATED WITH THE EVENT. THEREFORE, THIS REPORT IS BEING SUBMITTED IN ABUNDANCE OF CAUTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17701856 ↗InjuryAug 31, 2023Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
ON MAY 22, 2023, FUJIFILM HEALTHCARE AMERICAS CORPORATION WAS INFORMED OF AN EVENT INVOLVING EC-760R-V/L. IT WAS REPORTED THAT CUSTOMER REQUESTED ADVICE ON A POTENTIAL INFECTION CONTROL ISSUE. THEY HAD AN INEXPERIENCED SCOPE TECH PUT ¼ - 1/2 CUP OF EMPOWER ENZYMATIC SOLUTION INTO ONE OF THEIR MEDIVATORS DSD WASHERS WHERE THE OPA WAS SUPPOSED TO GO, AND TWO OF THE SCOPES THAT WERE RUN THROUGH THERE WERE USED ON PATIENTS BEFORE THEY CAUGHT THE MISTAKE. THERE IS NO SERIOUS INJURY OR DEATH ASSOCIATED WITH THE EVENT. THEREFORE, THIS REPORT IS BEING SUBMITTED IN ABUNDANCE OF CAUTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17666059 ↗InjuryAug 31, 2023Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
ON MAY 22, 2023, FUJIFILM HEALTHCARE AMERICAS CORPORATION WAS INFORMED OF AN EVENT INVOLVING EC-760R-V/L. IT WAS REPORTED THAT CUSTOMER REQUESTED ADVICE ON A POTENTIAL INFECTION CONTROL ISSUE. THEY HAD AN INEXPERIENCED SCOPE TECH PUT ¼ - 1/2 CUP OF EMPOWER ENZYMATIC SOLUTION INTO ONE OF THEIR MEDIVATORS DSD WASHERS WHERE THE OPA WAS SUPPOSED TO GO, AND TWO OF THE SCOPES THAT WERE RUN THROUGH THERE WERE USED ON PATIENTS BEFORE THEY CAUGHT THE MISTAKE. THERE IS NO SERIOUS INJURY OR DEATH ASSOCIATED WITH THE EVENT. THEREFORE, THIS REPORT IS BEING SUBMITTED IN ABUNDANCE OF CAUTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17665921 ↗InjuryJul 25, 2023Sharp Edges; Sharp Edges
ON JULY 6, 2023, FUJIFILM CORPORATION WAS INFORMED OF AN EVENT INVOLVING EC-760R-V/L. THIS REPORT ADDRESSES THE THIRD OF THREE EVENTS THAT OCCURRED AT THE USER FACILITY USING THE SAME SCOPE. IT WAS REPORTED THAT DURING A PROCEDURE THE PATIENT EXPERIENCED AN ABRASION OF THE CECUM THAT IS SUSPECTED TO HAVE BEEN CAUSED BY A SHARP POINT ON THE DISTAL END. THE SCOPE WAS SWITCHED CAUSING A SLIGHT DELAY. THEREAFTER, THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT FURTHER INCIDENT. THERE IS NO DEATH ASSOCIATED WITH THE EVENT. AS SUCH, THIS REPORT IS BEING SUBMITTED IN ABUNDANCE OF CAUTION.
Patient problems: Abrasion
MAUDE report 17390525 ↗InjuryJul 25, 2023Sharp Edges; Sharp Edges
ON JUNE 27, 2023, FUJIFILM CORPORATION WAS INFORMED OF AN EVENT INVOLVING EC-760R-V/L. THIS REPORT ADDRESSES THE SECOND OF THREE EVENTS THAT OCCURRED AT THE USER FACILITY USING THE SAME SCOPE. IT WAS REPORTED THAT DURING A PROCEDURE THE PATIENT EXPERIENCED A MINOR COLON ABRASION THAT IS SUSPECTED TO HAVE BEEN CAUSED BY A SHARP POINT ON THE DISTAL END. THE SCOPE WAS SWITCHED CAUSING A SLIGHT DELAY. THEREAFTER, THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT FURTHER INCIDENT. THERE IS NO DEATH ASSOCIATED WITH THE EVENT. AS SUCH, THIS REPORT IS BEING SUBMITTED IN ABUNDANCE OF CAUTION.
Patient problems: Abrasion
MAUDE report 17390505 ↗MalfunctionJul 14, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2023, FUJIFILM CORPORATION WAS INFORMED OF AN EVENT INVOLVING EC-760R-V/L. THIS REPORT ADDRESSES THE FIRST OF THREE EVENTS THAT OCCURRED AT THE USER FACILITY USING THE SAME SCOPE. IT WAS REPORTED THAT DURING A PROCEDURE THE PATIENT EXPERIENCED A MINOR COLON ABRASION THAT IS SUSPECTED TO HAVE BEEN CAUSED BY A SHARP POINT ON THE DISTAL END. THE SCOPE WAS SWITCHED CAUSING A SLIGHT DELAY. THEREAFTER, THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT FURTHER INCIDENT. THERE IS NO DEATH ASSOCIATED WITH THE EVENT.
Patient problems: Abrasion
MAUDE report 17328844 ↗InjuryJul 7, 2023Sharp Edges; Sharp Edges
ON (B)(6) 2023, FUJIFILM HEALTHCARE AMERICAS CORPORATION WAS INFORMED OF AN EVENT INVOLVING EC-760R-V/L. THIS REPORT ADDRESSES THE FIRST OF THREE EVENTS THAT OCCURRED AT THE USER FACILITY USING THE SAME SCOPE. IT WAS REPORTED THAT DURING A PROCEDURE THE PATIENT EXPERIENCED A MINOR COLON ABRASION THAT IS SUSPECTED TO HAVE BEEN CAUSED BY A SHARP POINT ON THE DISTAL END. THE SCOPE WAS SWITCHED CAUSING A SLIGHT DELAY. THEREAFTER, THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT FURTHER INCIDENT. THERE IS NO DEATH ASSOCIATED WITH THE EVENT. DURING A RETROSPECTIVE REVIEW, IT WAS DETERMINED THAT THIS EVENT SHOULD BE REPORTED IN ABUNDANCE OF CAUTION.
Patient problems: Abrasion
MAUDE report 17283348 ↗InjuryJul 7, 2023Sharp Edges; Sharp Edges
ON (B)(6) 2023, FUJIFILM HEALTHCARE AMERICAS CORPORATION WAS INFORMED OF AN EVENT INVOLVING EC-760R-V/L. THIS REPORT ADDRESSES THE SECOND OF THREE EVENTS THAT OCCURRED AT THE USER FACILITY USING THE SAME SCOPE. IT WAS REPORTED THAT DURING A PROCEDURE THE PATIENT EXPERIENCED A MINOR COLON ABRASION THAT IS SUSPECTED TO HAVE BEEN CAUSED BY A SHARP POINT ON THE DISTAL END. THE SCOPE WAS SWITCHED CAUSING A SLIGHT DELAY. THEREAFTER, THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT FURTHER INCIDENT. THERE IS NO DEATH ASSOCIATED WITH THE EVENT. AS SUCH, THIS REPORT IS BEING SUBMITTED IN ABUNDANCE OF CAUTION.
Patient problems: Abrasion
MAUDE report 17283346 ↗InjuryJul 7, 2023Sharp Edges; Sharp Edges
ON (B)(6) 2023, FUJIFILM HEALTHCARE AMERICAS CORPORATION WAS INFORMED OF AN EVENT INVOLVING EC-760R-V/L. THIS REPORT ADDRESSES THE THIRD OF THREE EVENTS THAT OCCURRED AT THE USER FACILITY USING THE SAME SCOPE. IT WAS REPORTED THAT DURING A PROCEDURE THE PATIENT EXPERIENCED AN ABRASION OF THE CECUM THAT IS SUSPECTED TO HAVE BEEN CAUSED BY A SHARP POINT ON THE DISTAL END. THE SCOPE WAS SWITCHED CAUSING A SLIGHT DELAY. THEREAFTER, THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT FURTHER INCIDENT. THERE IS NO DEATH ASSOCIATED WITH THE EVENT. AS SUCH, THIS REPORT IS BEING SUBMITTED IN ABUNDANCE OF CAUTION.
Patient problems: Abrasion
MAUDE report 17283345 ↗InjuryDec 2, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON 06 NOV 2019 FUJIFILM MEDICAL SYSTEMS U.S.A., INC (FMSU) WAS INFORMED THAT A SMALL PERFORATION WAS DISCOVERED IN A PATIENT AFTER ROUTINE COLONOSCOPY ON A PATIENT WITH CROHN'S DISEASE. PATIENT 1: THE PHYSICIAN INDICATED THAT THE COLON WAS TORTUROUS AND FULL OF STOOL AND THERE WAS DIFFICULTY GETTING INTO THE ILEOCECAL VALVE; HE SUCCEEDED AND WITHDREW THE SCOPE. SHORTLY AFTER PROCEDURE THE PATIENT STARTED TO EXHIBIT SOME PAIN AND IT WAS DETERMINED WITH X-RAY THAT HE HAD A SMALL PERFORATION NEAR THE CECUM. DUE TO THE PATIENT HAVING CROHN'S DISEASE AND IT BEING A COMPLICATION, THE SCOPE WAS NOT CONSIDERED TO BE A CAUSE AT THAT TIME. THE PATIENT HAS HAD SURGERY TO CORRECT THE PERFORATION. PATIENT 2: THE FOLLOWING WEEK A COLONOSCOPY PROCEDURE WAS PERFORMED WITH SOME DIFFICULTY ON PATIENT 2, THE PHYSICIAN WAS ABLE TO COMPLETE THE PROCEDURE. THE PHYSICIAN STATED THAT AT ONE POINT OF THE PROCEDURE WHERE HE WAS RUNNING THE TIP OF THE SCOPE CLOSE TO THE MUCOSA, HE NOTICED, WHAT WAS DESCRIBED AS A "RED LINE OR SCRATCH" ON THE MUCOSA. THE PHYSICIAN ASKED THE TECHNICIAN IF THE SCOPE WAS THE SAME SCOPE THAT WAS USED THE WEEK PRIOR WITH THE PERFORATION CASE (PATIENT 1); IT WAS DETERMINED THAT IT
Patient problems: Perforation
MAUDE report 9398650 ↗InjuryDec 2, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON 06 NOV 2019 FUJIFILM CORPORATION WAS INFORMED THAT A SMALL PERFORATION WAS DISCOVERED IN A PATIENT AFTER ROUTINE COLONOSCOPY ON A PATIENT WITH CROHN'S DISEASE. PATIENT 1: THE PHYSICIAN INDICATED THAT THE COLON WAS TORTUROUS AND FULL OF STOOL AND THERE WAS DIFFICULTY GETTING INTO THE ILEOCECAL VALVE; HE SUCCEEDED AND WITHDREW THE SCOPE. SHORTLY AFTER PROCEDURE THE PATIENT STARTED TO EXHIBIT SOME PAIN AND IT WAS DETERMINED WITH X-RAY THAT HE HAD A SMALL PERFORATION NEAR THE CECUM. DUE TO THE PATIENT HAVING CROHN'S DISEASE AND IT BEING A COMPLICATION, THE SCOPE WAS NOT CONSIDERED TO BE A CAUSE AT THAT TIME. THE PATIENT HAD SURGERY TO CORRECT THE PERFORATION. PATIENT 2: THE FOLLOWING WEEK A COLONOSCOPY PROCEDURE WAS PERFORMED WITH SOME DIFFICULTY ON PATIENT 2, THE PHYSICIAN WAS ABLE TO COMPLETE THE PROCEDURE. THE PHYSICIAN STATED THAT AT ONE POINT OF THE PROCEDURE WHERE HE WAS RUNNING THE TIP OF THE SCOPE CLOSE TO THE MUCOSA, HE NOTICED, WHAT WAS DESCRIBED AS A "RED LINE OR SCRATCH" ON THE MUCOSA. THE PHYSICIAN ASKED THE TECHNICIAN IF THE SCOPE WAS THE SAME SCOPE THAT WAS USED THE WEEK PRIOR WITH THE PERFORATION CASE (PATIENT 1); IT WAS DETERMINED THAT IT WAS. THE PROCEDURE ON PATIE
Patient problems: Perforation
MAUDE report 9398632 ↗
FDA clearance and classification
Cleared through 510(k) K172916 on Mar 1, 2018 (decision: Substantially Equivalent). Submitted by Fujifilm Corporation as “FUJIFILM Endoscope Models EG-760R, EG-760Z, EC-760R-V/L and EC-760ZP-V/L, and FUJIFILM Water Tank Model WT-603”. The FDA review took 157 days. FDA record ↗
The FDA regulates product code FDS (gastroscope and accessories, flexible/rigid) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗
Common questions about the FUJIFILM EC-760R-V/L
Has the FUJIFILM EC-760R-V/L been recalled?
No FDA recall record in our data names this exact device identifier (DI 04547410358681), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the FUJIFILM EC-760R-V/L?
The FDA's MAUDE database lists 26 adverse event reports for this device, received between Sep 13, 2019 and Sep 16, 2026. In those reports, the most common specific device problems are “Sharp Edges” (6), “Improper or Incorrect Procedure or Method” (3) and “Partial Blockage” (3); the most common patient problems are “Abrasion” (8), “Laceration(s)” (2) and “Perforation” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the FUJIFILM EC-760R-V/L?
None of the reports are classified as deaths, 13 as injuries and 13 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the FUJIFILM EC-760R-V/L FDA approved?
It was cleared through 510(k) K172916 on Mar 1, 2018. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the FUJIFILM EC-760R-V/L, and is it still sold?
The FDA's device database lists Fujifilm Corporation as the company responsible for it (brand FUJIFILM, model EC-760R-V/L). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Gastroscope And Accessories, Flexible/Rigid devices
- PENTAX Medical EG29-i10cHoya Corporation · Gastroscope And Accessories, Flexible/Rigid1,623 reports
- EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-1TH190Olympus Medical Systems Corp. · Gastroscope And Accessories, Flexible/Rigid4 recalls6 deaths1,018 reports
- EVIS EXERA lll GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-XP190NOlympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid1 recall552 reports
- EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE GIF-Q165Olympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid351 reports
- EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF TYPE H180Olympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid1 death301 reports
- PENTAX EG27-i10Hoya Corporation · Gastroscope and accessories, flexible/rigid261 reports
- EVIS EXERA II COLONOVIDEOSCOPE OLYMPUS CF TYPE H180ALOlympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid2 deaths244 reports
- BALLOON MAJ-1351Olympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid174 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page