Skip to content

PENTAX Medical EG29-i10c (UDI 04961333247387)Recalls & reported problems

Hoya Corporation · PENTAX Medical Video Upper GI Scope EG29-i10c · Class 2

All models of PENTAX Medical (34)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This is the Pentax video upper GI scope, a flexible endoscope used to look inside the stomach. The 1,633 reports often describe blocked or distorted camera images, fluid or blood leaking into the scope, and parts breaking. Some mention a clogged suction channel. The records include no deaths, 2 injuries and 1,631 malfunctions, and no recalls name this device. Reports fell to 46 in the last 12 months from 86 in the 12 months before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
1,623
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1,623 adverse event reports that list this exact device (DI 04961333247387), received between Jul 28, 2021 and Jul 13, 2026 — classified as 2 injuries, 1,621 malfunctions.

The most frequently reported specific device problem is “Optical Obstruction” (406 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 61 reportsDec 2023: 65 reportsJan 2024: 64 reportsFeb 2024: 31 reportsMar 2024: 47 reportsApr 2024: 29 reportsMay 2024: 33 reportsJun 2024: 41 reportsJul 2024: 54 reportsAug 2024: 12 reportsSep 2024: 10 reportsOct 2024: 10 reportsNov 2024: 7 reportsDec 2024: 3 reportsJan 2025: 6 reportsFeb 2025: 2 reportsMar 2025: 10 reportsApr 2025: 11 reportsMay 2025: 6 reportsJun 2025: 12 reportsJul 2025: 12 reportsAug 2025: 13 reportsSep 2025: 2 reportsOct 2025: 2 reportsNov 2025: 4 reportsDec 2025: 4 reportsJan 2026: 7 reportsFeb 2026: 3 reportsMar 2026: 4 reportsApr 2026: 4 reportsMay 2026: 6 reportsJun 2026: 2 reportsJul 2026: 2 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202665
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Optical Obstruction406
  • Fluid/Blood Leak335
  • Break313
  • Erratic or Intermittent Display157
  • Optical Distortion154
  • Obstruction of Flow117
  • No Display/Image92
  • Unintended Movement29

Patient problems

  • No Clinical Signs, Symptoms or Conditions1,619
  • Insufficient Information6
  • Foreign Body In Patient2

Recent adverse event reports

The 25 most recent of 1,623 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 13, 2026Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25791979 ↗
  • MalfunctionJul 1, 2026Fluid/Blood Leak

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(FLUID DAMAGE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25675214 ↗
  • MalfunctionJun 9, 2026No Display/Image

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(BLACKOUT ).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25439756 ↗
  • MalfunctionJun 2, 2026Optical Obstruction

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(CLOUDY ).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25356773 ↗
  • MalfunctionMay 21, 2026Fluid/Blood Leak

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(FLUID DAMAGE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25237980 ↗
  • MalfunctionMay 19, 2026Fluid/Blood Leak

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(CLOUDY).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25212151 ↗
  • MalfunctionMay 14, 2026Fluid/Blood Leak

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(FLUID DAMAGE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25167864 ↗
  • MalfunctionMay 12, 2026Fluid/Blood Leak

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(FLUID DAMAGE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25146003 ↗
  • MalfunctionMay 7, 2026Optical Obstruction

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE (CLOUDY ).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25108337 ↗
  • MalfunctionMay 7, 2026Poor Quality Image

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE (BLACKOUT).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25108326 ↗
  • MalfunctionApr 15, 2026Fluid/Blood Leak

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE (FLUID DAMAGE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24880678 ↗
  • MalfunctionApr 15, 2026No Display/Image

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(BLACKOUT).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24880676 ↗
  • MalfunctionApr 6, 2026Fluid/Blood Leak

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(FLUID DAMAGE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24787277 ↗
  • MalfunctionApr 3, 2026Obstruction of Flow

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. OPERATION/SUCTION CHANNEL CLOGGED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24772958 ↗
  • MalfunctionMar 30, 2026Fluid/Blood Leak

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(CLOUDY).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24724891 ↗
  • MalfunctionMar 30, 2026Fluid/Blood Leak

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(FLUID DAMAGE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24724774 ↗
  • MalfunctionMar 24, 2026Optical Obstruction

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(CLOUDY ).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24678804 ↗
  • MalfunctionMar 23, 2026No Display/Image

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE (BLACKOUT).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24659312 ↗
  • MalfunctionFeb 20, 2026No Display/Image

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE (BLACKOUT ).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24398082 ↗
  • MalfunctionFeb 9, 2026Poor Quality Image

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(BLACKOUT ).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24288761 ↗
  • MalfunctionFeb 9, 2026Optical Obstruction

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(CLOUDY ).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24288731 ↗
  • MalfunctionJan 30, 2026Erratic or Intermittent Display

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24219010 ↗
  • MalfunctionJan 27, 2026No Display/Image

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(BLACKOUT ).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24186258 ↗
  • MalfunctionJan 27, 2026Optical Obstruction

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE (CLOUDY).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24186253 ↗
  • MalfunctionJan 27, 2026Fluid/Blood Leak

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE (FLUID DAMAGE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24186243 ↗

Other Gastroscope And Accessories, Flexible/Rigid devices

All Gastroscope And Accessories, Flexible/Rigid devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page