Skip to content

EVIS EXERA II COLONOVIDEOSCOPE OLYMPUS CF TYPE H180AL (UDI 04953170202339)Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of EVIS EXERA II COLONOVIDEOSCOPE (28)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Olympus EVIS EXERA II colonoscope has 247 reports on file, with 2 deaths and 16 injuries. Many describe the scope failing to clean adequately, contamination, or reprocessing problems, and others describe poor images or blank screens. A few patients mention bleeding, perforation or viral infection. No recalls name this device, and reports rose to 64 in the latest year from 24 in the year before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
252
Deaths reported
2
Injuries reported
16
Recalls naming it
0

Summary

The FDA's MAUDE database contains 252 adverse event reports that list this exact device (DI 04953170202339), received between May 30, 2018 and Sep 30, 2026 — classified as 2 deaths, 16 injuries, 234 malfunctions.

The most frequently reported specific device problem is “Failure to Clean Adequately” (107 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 5 reportsDec 2023: 6 reportsJan 2024: 7 reportsFeb 2024: 9 reportsMar 2024: 12 reportsApr 2024: 9 reportsMay 2024: 2 reportsJun 2024: 8 reportsJul 2024: 1 reportsAug 2024: 5 reportsSep 2024: 4 reportsOct 2024: 4 reportsNov 2024: 2 reportsDec 2024: 1 reportsJan 2025: 1 reportsFeb 2025: 2 reportsMar 2025: 6 reportsApr 2025: 2 reportsMay 2025: 2 reportsJun 2025: 2 reportsJul 2025: 3 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 3 reportsDec 2025: 8 reportsJan 2026: 11 reportsFeb 2026: 5 reportsMar 2026: 5 reportsApr 2026: 15 reportsMay 2026: 6 reportsJun 2026: 5 reportsJul 2026: 3 reportsAug 2026: 2 reportsSep 2026: 6 reportsOct 2026: 0 reports20242025202615
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Clean Adequately107
  • Microbial Contamination of Device37
  • Device Reprocessing Problem34
  • Poor Quality Image15
  • No Display/Image13
  • Adverse Event Without Identified Device or Use Problem9
  • Corroded7
  • Break6

Patient problems

  • No Clinical Signs, Symptoms or Conditions215
  • No Known Impact Or Consequence To Patient12
  • Hemorrhage/Blood Loss/Bleeding4
  • No Consequences Or Impact To Patient4
  • Perforation4
  • Viral Infection4
  • Death2
  • No Patient Involvement2

Recent adverse event reports

The 25 most recent of 252 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026No Display/Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50951833 ↗
  • MalfunctionSep 17, 2026Erratic or Intermittent Display

    The customer reported the colonovideoscope had a separate issue. During the initial evaluation of the device, noise and intermittent image were observed. No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50805519 ↗
  • MalfunctionSep 10, 2026Poor Quality Image

    The customer returned the colonovideoscope, which is an Olympus asset with no reported complaint. During the initial evaluation of the device noise image occurs. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50715484 ↗
  • MalfunctionSep 10, 2026Device Reprocessing Problem

    The customer reported the colonovideoscope had a separate issue. During the initial evaluation of the device, it was observed that c-cover and adhesive on distal sheath had foreign objects. No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50705283 ↗
  • MalfunctionSep 9, 2026Device Reprocessing Problem

    The customer returned the Colonovideoscope for a separate issue. During the initial evaluation of the returned device, it was observed that foreign objects were present on the auxiliary jet channel. No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50684115 ↗
  • MalfunctionSep 2, 2026Poor Quality Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50572332 ↗
  • MalfunctionAug 13, 2026No Display/Image; Failure to Power Up

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50244223 ↗
  • MalfunctionAug 5, 2026Device Reprocessing Problem; Failure to Clean Adequately

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50143340 ↗
  • MalfunctionJul 28, 2026Connection Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50055664 ↗
  • MalfunctionJul 3, 2026Failure to Clean Adequately

    THE CUSTOMER RETURNED THE COLONOVIDEOSCOPE TO THE SERVICE CENTER FOR EVALUATION RELATED TO A SEPARATE ISSUE. DURING DEVICE EVALUATION, THE REPAIR CENTER TECHNICIAN IDENTIFIED THE DEVICE HAD FOREIGN MATERIALS IN AUXILIARY WATER CHANNEL (J-TUBE). THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25703446 ↗
  • MalfunctionJul 1, 2026No Device Output

    THE CUSTOMER RETURNED THE COLONOVIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE SERVICE REPAIR TECHNICIAN INDICATES THAT NO ILLUMINATION OBTAINED WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25683036 ↗
  • MalfunctionJun 26, 2026Poor Quality Image

    THE CUSTOMER RETURNED THE COLONOVIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE SERVICE REPORT TECHNICIAN INDICATES THAT NOISY IMAGE WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25641118 ↗
  • MalfunctionJun 11, 2026Poor Quality Image

    THE CUSTOMER REPORTED LIMITED FIELD OF VIEW ON THE RIGHT SIDE, AND A LONGITUDINAL SMEAR ON THE MONITOR DURING AN UNSPECIFIED PROCEDURE INVOLVING AN OLYMPUS COLONOVIDEOSCOPE. THERE WAS NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25473085 ↗
  • MalfunctionJun 6, 2026No Display/Image

    THE CUSTOMER REPORTED THE SCREEN GOES BLANK DURING UNSPECIFIED PROCEDURE INVOLVING AN OLYMPUS GASTROINTESTINAL VIDEOSCOPE. THERE WAS NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25418800 ↗
  • MalfunctionJun 4, 2026Display or Visual Feedback Problem; Failure to Clean Adequately

    THE CUSTOMER RETURNED THE COLONOVIDEOSCOPE TO THE SERVICE CENTER FOR EVALUATION RELATED TO A SEPARATE ISSUE. DURING TESTING, THE SERVICE REPAIR TECHNICIAN IDENTIFIED THE DEVICE HAD DIRT ON THE OBJECTIVE LENS AND THE IMAGE DID NOT APPEAR COMPLETELY ON THE MONITOR. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25395071 ↗
  • MalfunctionJun 4, 2026Failure to Clean Adequately

    THE CUSTOMER RETURNED THE COLONOVIDEOSCOPE WHICH IS AN OLYMPUS ASSET WITH NO REPORTED COMPLAINT. DURING THE EVALUATION OF THE DEVICE, FOREIGN OBJECTS IN NOZZLE WAS OBSERVED. THE SERVICE REPAIR TECHNICIAN INDICATED THAT THE MOST LIKELY CAUSE OF THE FOREIGN OBJECTS IN NOZZLE COULD NOT BE ESTABLISHED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25392041 ↗
  • MalfunctionMay 31, 2026Corroded

    THE CUSTOMER RETURNED THE COLONOVIDEOSCOPE WHICH IS AN OLYMPUS ASSET WITH NO REPORTED COMPLAINT. DURING THE DEVICE ANALYSIS, CORRODED NOZZLE WAS FOUND. IT WAS DETERMINED THAT THE NOZZLE NEEDED TO BE REPLACED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25342465 ↗
  • MalfunctionMay 22, 2026Corroded

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED CORROSION ON THE ELECTRICAL CONNECTOR AND THE SCOPE COVER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25254370 ↗
  • MalfunctionMay 19, 2026No Display/Image

    IT WAS REPORTED THAT THE COLONOVIDEOSCOPE EXHIBITED NO IMAGE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25219843 ↗
  • MalfunctionMay 12, 2026Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE HAD FOREIGN OBJECT IN AUXILIARY TUBE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25153732 ↗
  • MalfunctionMay 8, 2026Device Reprocessing Problem

    IT WAS OBSERVED THAT DURING DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED FOREIGN MATERIAL IN THE LIGHT GUIDE ROD LENS AT THE DISTAL END OF THE INSERTION PART. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25127127 ↗
  • MalfunctionMay 1, 2026Poor Quality Image

    IT WAS REPORTED THAT THE COLONOVIDEOSCOPE HAD A LINE VISIBLE IN THE PICTURE. THE EVENT OCCURRED DURING INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25057008 ↗
  • MalfunctionApr 21, 2026No Display/Image

    IT WAS REPORTED THAT THE COLONOVIDEOSCOPE HAD NO IMAGE. THE EVENT HAD OCCURRED DURING MAINTENANCE. THERE WERE NO REPORTS OF PATIENT HARM. NO IMAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24945820 ↗
  • MalfunctionApr 21, 2026Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE PRESENTED FOREIGN OBJECTS IN THE CONNECTING TUBE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24943254 ↗
  • MalfunctionApr 16, 2026Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE HAD A STICKY AT THE SCOPE CONNECTOR COVER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24900956 ↗

Other Gastroscope and accessories, flexible/rigid devices

All Gastroscope and accessories, flexible/rigid devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page