Skip to content

FUJIFILM EC-760P-V/LRecalls & reported problems

Fujifilm Corporation · This device is intended for the visualization of the lower digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the rectum and large intestine. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04547410368871), received between Oct 25, 2023 and Apr 22, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Difficult to Remove” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Remove1
  • Entrapment of Device1
  • Insufficient Device Problem Information1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 22, 2025Entrapment of Device; Difficult to Remove; Entrapment of Device; Difficult to Remove

    ON (B)(6) 2025, FUJIFILM HEALTHCARE AMERICAS CORPORATION WAS INFORMED OF AN EVENT INVOLVING EC-760P-V/L. IT WAS REPORTED THAT DURING THE COLONOSCOPY DIAGNOSTIC PROCEDURE, IT APPEARED TO HAVE RETROFLEX UNINTENTIONALLY IN THE SIGMOID COLON, WHICH WAS NOT ALLOWING THE PROCEDURE TO BE COMPLETED. COLONOSCOPY WAS NOT COMPLETED. POTENTIAL SIGMOID COLON MUCOSAL BRUISE. DURING A COLONOSCOPY, THE PHYSICIAN ENCOUNTERED DIFFICULTY ADVANCING THE SCOPE DUE TO MULTIPLE DIVERTICULA AND ABORTED THE PROCEDURE; WHILE RETRACTING, THE SCOPE BECAME STUCK BUT IT WAS RELEASED AFTER PATIENT REPOSITIONING AND A DIGITAL EXAM, WHICH MIGHT POTENTIALLY RESULT IN MINOR BRUISE. HOWEVER, THERE WAS NO ADVERSE EVENTS. THE PATIENT REMAINED STABLE, WAS DISCHARGED, AND FOLLOW-UP CONFIRMED FULL RECOVERY. THE INCIDENT IS BEING REPORTED IN AN ABUNDANCE OF CAUTION AS THE MALFUNCTION COULD CAUSE CONTRIBUTE TO SERIOUS INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21885557 ↗
  • MalfunctionOct 25, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    ON SEPTEMBER 25TH 2023, FUJIFILM HEALTHCARE AMERICAS CORPORATION WAS INFORMED OF AN EVENT INVOLVING EC-760P-V/L. IT WAS REPORTED THAT DURING A DIAGNOSTIC FOR COLON PROCEDURE THERE WAS NO PERIPHERAL VISION AND DECREASE LIGHTING. THEIR PROCEDURE TOOK LONGER TO FINISH. DUE TO UNKNOWN STATUS OF PATIENT AND EQUIPMENT, THE REPORT IS SUBMITTED WITH ABUNDANCE OF CAUTION. THERE WAS NO INJURY TO THE PATIENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18003172 ↗

Other Gastroscope And Accessories, Flexible/Rigid devices

All Gastroscope And Accessories, Flexible/Rigid devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page