iCare iCare C2 Type (120V)Recalls & reported problems

Nakanishi Inc. · Class 1

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04560264538986), received between May 14, 2026 and May 27, 2026 — classified as 2 injuries.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 2 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component2

Patient problems

  • Needle Stick/Puncture2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 27, 2026Detachment of Device or Device Component

    ON (B)(6) 2026, NAKANISHI BECAME AWARE OF DENTAL ASSISTANT INJURY CAUSED BY A HANDPIECE MAINTENANCE UNIT THROUGH A COMPLAINT INPUT INTO THE COMPLAINT DATABASE BY A DISTRIBUTOR (NSK AMERICA). THE DETAILS NAKANISHI OBTAINED ARE AS FOLLOWS: THE EVENT OCCURRED ON (B)(6) 2026. A DENTAL ASSISTANT WAS PERFORMING THE CHUCK MAINTENANCE ON THEIR HANDPIECE USING THE ICARE HANDPIECE MAINTENANCE UNIT (SERIAL NO. (B)(6)) BEFORE REPROCESSING. THE ASSISTANT WAS LUBRICATING THE HANDPIECE CHUCK USING THE CHUCK LUBRICATION NOZZLE INCLUDED WITH THE ICARE WHEN THE NOZZLE SHOT OFF THE CONNECTOR UNEXPECTEDLY AND THE TIP OF THE NOZZLE PUNCTURED THE ASSISTANT'S FINGER. THE DENTAL ASSISTANT IS REPORTED TO HAVE HEALED FROM THE INJURY WITHOUT NEED FOR ANY ADDITIONAL MEDICAL TREATMENT.

    Patient problems: Needle Stick/Puncture

    MAUDE report 25311162 ↗
  • InjuryMay 14, 2026Detachment of Device or Device Component

    A DENTAL ASSISTANT SUSTAINED AN INJURY TO THEIR FINGER FROM AN ICARE MAINTENANCE UNIT WHILE PERFORMING THE CHUCK MAINTENANCE ON THEIR HANDPIECE BEFORE REPROCESSING. THE ASSISTANT WAS LUBRICATING THE HANDPIECE CHUCK USING THE CHUCK LUBRICATION NOZZLE INCLUDED WITH THE ICARE WHEN THE NOZZLE SHOT OFF THE CONNECTOR UNEXPECTEDLY AND THE TIP OF THE NOZZLE PUNCTURED THE ASSISTANT'S FINGER. THE DENTAL ASSISTANT IS REPORTED TO HAVE HEALED FROM THE INJURY WITHOUT NEED FOR ANY ADDITIONAL MEDICAL TREATMENT.

    Patient problems: Needle Stick/Puncture

    MAUDE report 25175281 ↗

Other Handpiece, air-powered, dental devices