TI-MAX Z CONTRA/HP Z95LRecalls & reported problems
Nakanishi Inc. · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04560264565951), received between Jul 10, 2025 and Jul 24, 2025 — classified as 3 injuries.
The most frequently reported specific device problem is “Overheating of Device” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Overheating of Device3
Patient problems
- Partial thickness (Second Degree) Burn2
- Superficial (First Degree) Burn1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryJul 24, 2025Overheating of Device; Overheating of Device
ON JULY 3, 2025, NAKANISHI BECAME AWARE OF A HANDPIECE OVERHEATING THROUGH A COMPLAINT INPUT INTO THE COMPLAINT DATABASE BY A DISTRIBUTOR (NSK AMERICA). DETAILS ARE AS FOLLOWS: - THE DENTIST WAS PERFORMING A DENTAL PROCEDURE ON A PATIENT USING THE Z95L HANDPIECE (SERIAL NO. (B)(6)). - DURING THE PROCEDURE, THE HEAD OF THE HANDPIECE MADE CONTACT WITH THE PATIENT'S LIP AS THE DENTIST WAS REMOVING IT FROM THEIR MOUTH, AND THE PATIENT RECEIVED A MINOR THERMAL BURN. - THE PATIENT CHECKED THE HANDPIECE AND FOUND THE HEAD HAD BECOME EXTREMELY HOT BUT WAS NOT AWARE THAT IT HAD HEATED UP. - THERE WAS NO VISIBLE BLISTERING TO THE PATIENT'S LIP, AND VASELINE WAS APPLIED TO THE AFFECTED AREA AT THE TIME OF THE INCIDENT. - THE PATIENT WAS REPORTED TO BE FINE THE NEXT DAY WITH NO DISCOMFORT.
Patient problems: Superficial (First Degree) Burn
MAUDE report 22584551 ↗InjuryJul 10, 2025Overheating of Device; Overheating of Device
ON JUNE 17, 2025, NAKANISHI BECAME AWARE OF A HANDPIECE OVERHEATING THROUGH A COMPLAINT INPUT INTO THE COMPLAINT DATABASE BY A DISTRIBUTOR (NSK AMERICA). DETAILS ARE AS FOLLOWS: - THE EVENT OCCURRED ON (B)(6) 2025. - THE DENTIST WAS PERFORMING A CROWN PREPARATION PROCEDURE ON A PATIENT USING THE Z95L HANDPIECE (SERIAL NO. (B)(6)). - DURING THE PROCEDURE, THE HANDPIECE OVERHEATED, AND THE PATIENT RECEIVED A MINOR THERMAL BURN INJURY TO THEIR LOWER LIP. - NO MEDICAL TREATMENT WAS NECESSARY AT THE TIME OF THE INJURY. - THE DENTIST HAD A FOLLOW-UP VISIT WITH THE PATIENT AND THE PATIENT IS REPORTED TO BE HEALING AS EXPECTED ALTHOUGH THE PATIENT DID MENTION FEELING SIGNIFICANT PAIN AROUND THE AFFECTED AREA.
Patient problems: Partial thickness (Second Degree) Burn
MAUDE report 22465736 ↗InjuryJul 10, 2025Overheating of Device; Overheating of Device
ON JUNE 17, 2025, NAKANISHI BECAME AWARE OF A HANDPIECE OVERHEATING THROUGH A COMPLAINT INPUT INTO THE COMPLAINT DATABASE BY A DISTRIBUTOR (B)(4). DETAILS ARE AS FOLLOWS: - THE EVENT OCCURRED ON (B)(6) 2025. - THE DENTIST WAS PERFORMING A RESTORATIVE FILLING PROCEDURE ON TEETH NUMBERS 2, 3, 30, 31 OF A PATIENT USING THE Z95L HANDPIECE (SERIAL NO. (B)(6)) - DURING THE PROCEDURE, THE HANDPIECE OVERHEATED, AND THE PATIENT RECEIVED A SEVERE THERMAL BURN INJURY TO THEIR INNER LOWER LIP. - NO MEDICAL TREATMENT WAS NECESSARY AT THE TIME OF THE INJURY. - THE DENTIST HAD A FOLLOW-UP VISIT WITH THE PATIENT AND THE PATIENT IS REPORTED TO BE HEALING AS EXPECTED ALTHOUGH THE PATIENT DID MENTION FEELING NUMBNESS AROUND THE AFFECTED AREA.
Patient problems: Partial thickness (Second Degree) Burn
MAUDE report 22465725 ↗
Other Handpiece, contra- and right-angle attachment, dental devices
- Z95LNakanishi Inc. · Handpiece, contra- and right-angle attachment, dental27 reports
- Surgical Angle Handpiece SGA-E2SNakanishi Inc. · Handpiece, Contra- And Right-Angle Attachment, Dental11 reports
- Lares Research Smoothdrive 1:5 Contra angleLares Research · Handpiece, Contra- And Right-Angle Attachment, Dental8 reports
- Henry Schein 5701766Henry Schein, Inc. · Handpiece, contra- and right-angle attachment, dental4 reports
- Brasseler USA 5021285U0Peter Brasseler Holdings, LLC · Handpiece, contra- and right-angle attachment, dental3 reports
- Z99LNakanishi Inc. · Handpiece, contra- and right-angle attachment, dental3 reports
- Brasseler USA 5027608U0Peter Brasseler Holdings, LLC · Handpiece, contra- and right-angle attachment, dental2 reports
- TI-MAX nano CONTRA/HP nano95LSNakanishi Inc. · Handpiece, contra- and right-angle attachment, dental2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page