Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04589551391585), received between May 7, 2026 and Sep 30, 2026 — classified as 3 injuries.
The most frequently reported specific device problem is “Overheating of Device” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Overheating of Device3
Patient problems
- Partial thickness (Second Degree) Burn3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjurySep 30, 2026Overheating of Device
Patient problems: Partial thickness (Second Degree) Burn
MAUDE report 50952771 ↗InjuryMay 13, 2026Overheating of Device
ON APRIL 25, 2026, NAKANISHI BECAME AWARE OF HANDPIECE OVERHEATING THROUGH A COMPLAINT INPUT INTO THE COMPLAINT DATABASE BY A DISTRIBUTOR (NSK AMERICA). DETAILS ARE AS FOLLOWS: - THE EVENT OCCURRED ON APRIL 8, 2026. - THE DENTIST WAS PERFORMING A CROWN PREPARATION PROCEDURE ON A PATIENT USING THE Z99L HANDPIECE (SERIAL NO. (B)(6)). - DURING THE PROCEDURE, THE HANDPIECE BACK-CAP OVERHEATED ABNORMALLY, AND THE PATIENT RECEIVED A THERMAL BURN INJURY TO THE INSIDE OF THEIR CHEEK. - THE DENTIST REPORTED THE FOLLOWING "I WAS FINALIZING MY PREPARATION OF BOTH #29 AND #30, SO AT THAT TIME WE TOOK OUT THE CHEEK RETRACTOR I HAD BEEN USING THROUGHOUT THE PROCEDURE SO THE PATIENT COULD BITE, AND I COULD VERIFY MY OCCLUSAL REDUCTION WAS ADEQUATE. AFTER REALIZING I NEEDED A MINOR AMOUNT OF EXTRA REDUCTION, I TILTED MY HANDPIECE TO BE PARALLEL TO THE OCCLUSAL TABLE TO DO THE FINAL ADJUSTMENT, LEADING TO THE BACK OF THE HANDPIECE (WHERE I'D PUSH TO EJECT A BUR) RESTING ON THE PATIENT'S CHEEK. DURING THIS LAST STEP, THE PATIENT ABRUPTLY RAISED HER HAND AND INDICATED IT WAS HOT. I TOUCHED THE BACK OF THE HANDPIECE, AND IT WAS EXTREMELY HOT AND CHARRED LOOKING (BLACKENED). AT THIS POINT I SAW A SIGNI
Patient problems: Partial thickness (Second Degree) Burn
MAUDE report 25156905 ↗InjuryMay 7, 2026Overheating of Device
DURING A CROWN PREPARATION PROCEDURE, THE HANDPIECE BACK-CAP OVERHEATED ABNORMALLY, AND THE PATIENT RECEIVED A THERMAL BURN INJURY TO THE INSIDE OF THEIR CHEEK. THE DOCTOR REPORTED THE FOLLOWING "I WAS FINALIZING MY PREPARATION OF BOTH #29 AND #30, SO AT THAT TIME WE TOOK OUT THE CHEEK RETRACTOR I HAD BEEN USING THROUGHOUT THE PROCEDURE SO THE PATIENT COULD BITE, AND I COULD VERIFY MY OCCLUSAL REDUCTION WAS ADEQUATE. AFTER REALIZING I NEEDED A MINOR AMOUNT OF EXTRA REDUCTION, I TILTED MY HANDPIECE TO BE PARALLEL TO THE OCCLUSAL TABLE TO DO THE FINAL ADJUSTMENT, LEADING TO THE BACK OF THE HANDPIECE (WHERE I'D PUSH TO EJECT A BUR) RESTING ON THE PATIENT'S CHEEK. DURING THIS LAST STEP, THE PATIENT ABRUPTLY ROSE HER HAND AND INDICATED IT WAS HOT. I TOUCHED THE BACK OF THE HANDPIECE, AND IT WAS EXTREMELY HOT AND CHARRED LOOKING (BLACKENED). AT THIS POINT I SAW A SIGNIFICANT BURN ON THE PATIENT'S CHEEK. I RE-ANESTHETIZED FOR COMFORT SO WE COULD SCAN THE TEETH AND FABRICATED TEMPS. I INFORMED THE PATIENT OF WHAT HAPPENED." THE PATIENT WAS PRESCRIBED PAIN MEDICATION AND AN ORAL RINSE AT THE TIME OF THE INCIDENT. THE PATIENT REPORTED TO THE DOCTOR THAT THEY WERE IN SO MUCH PAIN AFTER THE LO
Patient problems: Partial thickness (Second Degree) Burn
MAUDE report 25117505 ↗
Other Handpiece, contra- and right-angle attachment, dental devices
- Z95LNakanishi Inc. · Handpiece, contra- and right-angle attachment, dental27 reports
- Surgical Angle Handpiece SGA-E2SNakanishi Inc. · Handpiece, Contra- And Right-Angle Attachment, Dental11 reports
- Lares Research Smoothdrive 1:5 Contra angleLares Research · Handpiece, Contra- And Right-Angle Attachment, Dental8 reports
- Henry Schein 5701766Henry Schein, Inc. · Handpiece, contra- and right-angle attachment, dental4 reports
- Brasseler USA 5021285U0Peter Brasseler Holdings, LLC · Handpiece, contra- and right-angle attachment, dental3 reports
- TI-MAX Z CONTRA/HP Z95LNakanishi Inc. · Handpiece, contra- and right-angle attachment, dental3 reports
- TI-MAX nano CONTRA/HP nano95LSNakanishi Inc. · Handpiece, contra- and right-angle attachment, dental2 reports
- Brasseler USA 5027608U0Peter Brasseler Holdings, LLC · Handpiece, contra- and right-angle attachment, dental2 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.