TI-MAX nano CONTRA/HP nano95LSRecalls & reported problems
Nakanishi Inc. · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04560264566002), received between Sep 11, 2025 and Sep 12, 2025 — classified as 2 injuries.
The most frequently reported specific device problem is “Overheating of Device” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Overheating of Device2
Patient problems
- Partial thickness (Second Degree) Burn2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjurySep 12, 2025Overheating of Device; Overheating of Device
ON AUGUST 19, 2025, NAKANISHI BECAME AWARE OF A HANDPIECE OVERHEATING THROUGH A COMPLAINT INPUT INTO THE COMPLAINT DATABASE BY A DISTRIBUTOR (NSK AMERICA). DETAILS ARE AS FOLLOWS: - THE EVENT OCCURRED ON (B)(6) 2025. - THE DENTIST RAN THE NANO95LS HANDPIECE (SERIAL NO. (B)(6)) OUTSIDE OF THE PATIENT'S MOUTH TO MAKE SURE WATER AND LIGHT COMES ON. - WITHIN A FEW SECONDS OF WORKING ON THE TOOTH THE PATIENT REACTED IN PAIN. - THE DENTIST ASKED THE PATIENT IF THE TOOTH WAS NUMB AND THEY STATED IT WAS FINE, SO THE DENTIST PROCEEDED WITH THE PREPARATION OF THE TOOTH FOR THE CROWN. - ONCE THE DENTIST REMOVED THE HANDPIECE FROM THE PATIENT'S MOUTH TO CHANGE BURS, THE DENTIST NOTICED THAT THE HEAD OF THE HANDPIECE WAS EXTREMELY HOT TO THE TOUCH. - THIS IS WHEN THE DENTIST OBSERVED THAT THE PATIENT'S BUCCAL MUCOSA HAD BEEN BURNED.
Patient problems: Partial thickness (Second Degree) Burn
MAUDE report 23034849 ↗InjurySep 11, 2025Overheating of Device; Overheating of Device
THE DOCTOR REPORTED THAT THEY RAN THE HANDPIECE OUTSIDE OF THE PATIENT'S MOUTH TO MAKE SURE WATER AND LIGHT COMES ON. WITHIN A FEW SECONDS OF WORKING ON THE TOOTH THE PATIENT REACTED IN PAIN. THE DOCTOR ASKED THE PATIENT IF THE TOOTH WAS NUMB AND THEY STATED IT WAS FINE, SO THE DOCTOR PROCEEDED WITH THE PREPARATION OF THE TOOTH FOR THE CROWN. ONCE THE DOCTOR REMOVED THE HANDPIECE FROM THE PATIENT'S MOUTH TO CHANGE BURS, THE DOCTOR NOTICED THE HEAD OF THE HANDPIECE WAS EXTREMELY HOT TO THE TOUCH. THIS IS WHEN THE DOCTOR OBSERVED THAT THE PATIENT'S BUCCAL MUCOSA HAD BEEN BURNED.
Patient problems: Partial thickness (Second Degree) Burn
MAUDE report 23032289 ↗
Other Handpiece, contra- and right-angle attachment, dental devices
- Z95LNakanishi Inc. · Handpiece, contra- and right-angle attachment, dental27 reports
- Surgical Angle Handpiece SGA-E2SNakanishi Inc. · Handpiece, Contra- And Right-Angle Attachment, Dental11 reports
- Lares Research Smoothdrive 1:5 Contra angleLares Research · Handpiece, Contra- And Right-Angle Attachment, Dental8 reports
- Henry Schein 5701766Henry Schein, Inc. · Handpiece, contra- and right-angle attachment, dental4 reports
- Brasseler USA 5021285U0Peter Brasseler Holdings, LLC · Handpiece, contra- and right-angle attachment, dental3 reports
- Z99LNakanishi Inc. · Handpiece, contra- and right-angle attachment, dental3 reports
- TI-MAX Z CONTRA/HP Z95LNakanishi Inc. · Handpiece, contra- and right-angle attachment, dental3 reports
- Brasseler USA 5027608U0Peter Brasseler Holdings, LLC · Handpiece, contra- and right-angle attachment, dental2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page