Skip to content

Z45LRecalls & reported problems

Nakanishi Inc. · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04560264574083), received between Mar 25, 2025 and Mar 25, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Overheating of Device” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Overheating of Device1

Patient problems

  • Partial thickness (Second Degree) Burn1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 25, 2025Overheating of Device; Overheating of Device

    THE HANDPIECE HEAD OVERHEATED DURING A TOOTH EXTRACTION PROCEDURE ON #3 AND #31 TEETH AND THE PATIENT RECEIVED A THERMAL BURN INJURY TO THE INSIDE OF THEIR CHEEK. THERE HAS BEEN A FOLLOW-UP VISIT WITH THE PATIENT AT THE OFFICE AND THE INJURY IS REPORTED TO HAVE HEALED NORMALLY WITHOUT NEED FOR ADDITIONAL MEDICAL TREATMENT.

    Patient problems: Partial thickness (Second Degree) Burn

    MAUDE report 21692113 ↗

Other Handpiece, contra- and right-angle attachment, dental devices

All Handpiece, contra- and right-angle attachment, dental devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page