Skip to content

OES HYSTEROFIBERSCOPE OLYMPUS HYF TYPE XP (UDI 04953170058219)Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of OES HYSTEROFIBERSCOPE (5)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04953170058219), received between May 14, 2020 and May 29, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Corroded” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Corroded1
  • Mechanical Problem1
  • Microbial Contamination of Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 29, 2026Corroded

    THE CUSTOMER RETURNED THE HYSTEROFIBERSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE OLYMPUS SERVICE CENTER TECHNICIAN INDICATED THAT A CORRODED DISTAL END WAS FOUND. IT WAS DETERMINED THAT THE DISTAL END NEEDED TO BE REPLACED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25331787 ↗
  • MalfunctionMar 5, 2024Mechanical Problem; Mechanical Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE HYSTEROFIBERSCOPE EXHIBITED AN ANGLE LOCKING ROD THAT WAS TIGHT. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18843151 ↗
  • MalfunctionMay 14, 2020Microbial Contamination of Device; Microbial Contamination of Device

    OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) WAS INFORMED THAT AS A RESULT OF ROUTINE MICROBIOLOGICAL TESTING BY THE USER FACILITY, FOLLOWING MICROBES WERE DETECTED FROM THE SAMPLE COLLECTED FROM THE SUBJECT DEVICE. UNSPECIFIED MICROBES (2 CFU/100ML) THE DEVICE HAD BEEN MANUALLY REPROCESSED USING PERACETIC ACID. THERE WAS NO REPORT OF INFECTION ASSOCIATED WITH THIS REPORT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10055652 ↗