BALLOON3 MAJ-233 (UDI 04953170058783)Recalls & reported problems
Olympus Medical Systems Corp. · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04953170058783), received between Sep 21, 2023 and Sep 26, 2023 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Separation Failure” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Separation Failure2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 26, 2023Separation Failure; Separation Failure
A CUSTOMER SUBMITTED A MEDSUN REPORT TO FDA (UNITED STATES FOOD AND DRUG ADMINISTRATION) STATING, THE REMNANT OF A BALLOON WAS LEFT BEHIND ON A SCOPE AFTER AN ENDOSCOPIC ULTRASOUND AND WENT THROUGH THE HIGH-LEVEL DISINFECTION PROCESS. IT IS NOT CLEAR IF THE BALLOON WAS NOT FULLY REMOVED PRIOR TO REPROCESSING OR IT BROKE, LEAVING THE ELASTIC BAND BEHIND. THIS WENT UNNOTICED DUE TO THE COLOR OF THE BALLOON BEING A SIMILAR COLOR TO THAT OF THE SCOPE. A NEW BALLOON WAS THEN APPLIED OVER THE REMNANT AND UTILIZED ON A SECOND PATIENT. THIS PLACED THE PATIENT AT RISK OF POTENTIALLY BEING EXPOSED TO A BLOODBORNE PATHOGEN. THERE WAS NO REPORT OF PATIENT HARM ASSOCIATED WITH THIS EVENT. THERE IS A REQUEST TO CONSIDER MANUFACTURING BALLOONS IN A DIFFERENT COLOR OTHER THAN OFF-WHITE, TO ENSURE THE BALLOON IS EASILY IDENTIFIABLE. THERE WERE NO ALERTS WHEN THE SUBJECT DEVICE WAS GOING THROUGH THE HIGH-LEVEL DISINFECTION PROCESS. THE REMNANT WAS NOT IDENTIFIED UNTIL AFTER THE PROCEDURE WAS COMPLETED. THE INTENDED PROCEDURE WAS A DIAGNOSTIC ENDOSCOPIC ULTRASOUND WITH BIOPSY. THE PATIENT WAS OFFERED LAB MONITORING FOR COMMUNICABLE AGENTS SUCH AS HUMAN IMMUNODEFICIENCY VIRUS (HIV) AND HEPATITIS (HEP
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17822222 ↗MalfunctionSep 21, 2023Separation Failure; Separation Failure
A CUSTOMER SUBMITTED A MEDSUN REPORT TO FDA (UNITED STATES FOOD AND DRUG ADMINISTRATION) STATING, THE REMNANT OF A BALLOON WAS LEFT BEHIND ON A SCOPE AFTER AN ENDOSCOPIC ULTRASOUND AND WENT THROUGH THE HIGH-LEVEL DISINFECTION PROCESS. IT IS NOT CLEAR IF THE BALLOON WAS NOT FULLY REMOVED PRIOR TO REPROCESSING OR IT BROKE, LEAVING THE ELASTIC BAND BEHIND. THIS WENT UNNOTICED DUE TO THE COLOR OF THE BALLOON BEING A SIMILAR COLOR TO THAT OF THE SCOPE. A NEW BALLOON WAS THEN APPLIED OVER THE REMNANT AND UTILIZED ON A SECOND PATIENT. THIS PLACED THE PATIENT AT RISK OF POTENTIALLY BEING EXPOSED TO A BLOODBORNE PATHOGEN. THERE WAS NO REPORT OF PATIENT HARM ASSOCIATED WITH THIS EVENT. THERE IS A REQUEST TO CONSIDER MANUFACTURING BALLOONS IN A DIFFERENT COLOR OTHER THAN OFF-WHITE, TO ENSURE THE BALLOON IS EASILY IDENTIFIABLE. RELATED PATIENT IDENTIFIERS: (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17795689 ↗
Other Colonoscope and accessories, flexible/rigid devices
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS CF-HQ190LOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls2 deaths6,804 reports
- EVIS EXERA III XENON LIGHT SOURCE OLYMPUS CLV-190Olympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid6,617 reports
- EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-HQ190Olympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls5 deaths6,244 reports
- EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-H190Olympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid2 recalls7 deaths5,572 reports
- EVIS EXERA III COLONOVIDEOSCOPE OLYMPUS PCF-H190DLOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls2,830 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS PCF-H190LOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls2 deaths2,513 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS CF-H190IOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid1 recall1,967 reports
- PENTAX Medical EC38-i10cLHoya Corporation · Colonoscope and accessories, flexible/rigid1,618 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page