Skip to content

MAINTENANCE UNIT MU-1 (REF 5685852)Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

Quick summary

This Olympus MU-1 maintenance unit is used to maintain endoscopes. Its 148 reports describe breaking or cracked parts, power inlets and switches failing, and fluid getting into the unit, and some mention a dirty pump tube. The reports include 3 injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
147
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 147 adverse event reports that list this exact device (DI 04953170060434), received between Oct 26, 2020 and Jun 24, 2026 — classified as 3 injuries, 144 malfunctions.

The most frequently reported specific device problem is “Break” (78 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 3 reportsDec 2023: 2 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 4 reportsApr 2024: 4 reportsMay 2024: 3 reportsJun 2024: 1 reportsJul 2024: 3 reportsAug 2024: 2 reportsSep 2024: 3 reportsOct 2024: 4 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 1 reportsMar 2025: 4 reportsApr 2025: 1 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 2 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 2 reportsMar 2026: 7 reportsApr 2026: 3 reportsMay 2026: 2 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520267
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break78
  • Crack14
  • Electrical /Electronic Property Problem9
  • Failure to Power Up8
  • Power Problem8
  • Defective Component5
  • Failure to Clean Adequately5
  • Contamination3

Patient problems

  • No Clinical Signs, Symptoms or Conditions127
  • No Patient Involvement10
  • No Consequences Or Impact To Patient8
  • Electric Shock3
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1

Recent adverse event reports

The 25 most recent of 147 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 24, 2026Material Discolored; Failure to Clean Adequately

    THE CUSTOMER RETURNED THE MAINTENANCE UNIT TO THE SERVICE CENTER FOR AN SEPARATE ISSUE. DURING DEVICE ANALYSIS, THE SERVICE TECHNICIAN NOTED YELLOWING OF THE PUMP TUBE, STAINS ON THE PUMP TUBE, AND THAT THE INTERNAL TUBE WAS DIRTY AND YELLOW. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25609259 ↗
  • MalfunctionMay 11, 2026Break

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE MAINTENANCE UNIT HAD DAMAGE TO THE POWER SWITCH. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25144130 ↗
  • MalfunctionMay 8, 2026Break

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE MAINTENANCE UNIT EXHIBITED A DAMAGED POWER INLET. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25131258 ↗
  • MalfunctionApr 23, 2026Leak/Splash

    IT WAS REPORTED THAT MAINTENANCE UNIT HAD FLUID INVASION. THE EVENT OCCURRED DURING INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24975184 ↗
  • MalfunctionApr 13, 2026No Device Output; Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE MAINTENANCE UNIT EXHIBITED MISSING CAP AND FAULTY POWER SWITCH LIGHT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24861943 ↗
  • MalfunctionApr 9, 2026Connection Problem

    IT WAS REPORTED THAT THE MAINTENANCE UNIT POWER INLET IS BROKEN. THE ISSUE WAS FOUND DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24836058 ↗
  • MalfunctionMar 23, 2026Crack

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE MAINTENANCE UNIT HAD DAMAGE ON THE INLET PLUG. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24661532 ↗
  • MalfunctionMar 18, 2026Break

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE MAINTENANCE UNIT HAD CRUSHED POWER SWITCH, AND IT IS MECHANICALLY DAMAGED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24625566 ↗
  • MalfunctionMar 18, 2026Degraded

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE MAINTENANCE UNIT HAD A DETERIORATED INLET PLUG. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24625532 ↗
  • MalfunctionMar 17, 2026Break

    IT WAS REPORTED THAT THE MAINTENANCE UNIT HAD A DAMAGED PLUG DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24622188 ↗
  • MalfunctionMar 17, 2026Break

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE MAINTENANCE UNIT EXIBITED BROKEN POWER SWITCH. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24621028 ↗
  • MalfunctionMar 9, 2026No Device Output

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE MAINTENANCE UNIT HAD EXHIBITED A POWER SOCKET FAILURE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24548340 ↗
  • MalfunctionMar 6, 2026Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE MAINTENANCE UNIT HAD A DETACHED INLET PLUG. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24529994 ↗
  • MalfunctionFeb 27, 2026Break

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE MAINTENANCE UNIT PRESENTED DEFECTIVE MAINS SWITCH WITH COVER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24475278 ↗
  • MalfunctionFeb 2, 2026Break

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE MAINTENANCE UNIT EXHIBITED A BROKEN INLET PLUG. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24229199 ↗
  • MalfunctionOct 27, 2025Crack; Crack

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE MAINTENANCE UNIT EXHIBITED THE INLET PLUG IS CRACKED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23397796 ↗
  • MalfunctionOct 22, 2025Break; Break

    NO NEW ADDITIONAL INFORMAL RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23361981 ↗
  • MalfunctionAug 22, 2025Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    IT WAS REPORTED THE SUBJECT DEVICE HAD TRIPPED THE FI (FAULT INTERRUPTER) AND EXHIBITED A HIGHER PROTECTIVE CONDUCTOR LEAKAGE CURRENT DURING SETUP AND INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22873374 ↗
  • MalfunctionAug 21, 2025Break; Break

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22867399 ↗
  • MalfunctionMay 15, 2025Break; Break

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22032445 ↗
  • MalfunctionApr 2, 2025Break; Break

    IT WAS REPORTED THAT THE SUBJECT DEVICE EXHIBITED A POWER SWITCH BROKEN, THE ISSUE OCCURRED DURING SERVICE MAINTENANCE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21751777 ↗
  • MalfunctionMar 28, 2025Break; Break

    IT WAS REPORTED THAT THE SUBJECT DEVICE HAD A BROKEN LIGHT BULB IN THE SWITCH. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21722129 ↗
  • MalfunctionMar 27, 2025Break; Break

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE EXHIBITED DETERIORATION OF TUBE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21713870 ↗
  • MalfunctionMar 17, 2025Corroded; Corroded

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE EXHIBITED DIRT ON CONNECTOR. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21627822 ↗
  • MalfunctionMar 5, 2025Break; Break

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GI IMAGING EXHIBITED A TUBE BROKEN. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21525705 ↗

Other Colonoscope And Accessories, Flexible/Rigid devices

All Colonoscope And Accessories, Flexible/Rigid devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.