TRACHEAL INTUBATION FIBERSCOPE OLYMPUS LF-TPRecalls & reported problems

Olympus Medical Systems Corp. · LF-TP TRACHEAL INTUBATION FIBE RSCOPE-PO · Class 2

All models of TRACHEAL INTUBATION FIBERSCOPE (7)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Olympus LF-TP is a flexible scope used to guide breathing tubes into the airway. Its 141 reports often describe a peeling inner lining, and some describe bacterial contamination found during routine checks. There are no deaths and no injuries. 1 recall, still open, began in July 2023 over scopes that may get stuck in the breathing tube connector because they are too wide. Reports fell to 2 in the last 12 months from 8 before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
141
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 141 adverse event reports that list this exact device (DI 04953170136825), received between Jun 13, 2018 and Jan 8, 2026 — classified as 141 malfunctions.

The most frequently reported specific device problem is “Peeled/Delaminated” (100 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 5 reportsDec 2023: 1 reportsJan 2024: 3 reportsFeb 2024: 4 reportsMar 2024: 8 reportsApr 2024: 5 reportsMay 2024: 4 reportsJun 2024: 1 reportsJul 2024: 3 reportsAug 2024: 2 reportsSep 2024: 3 reportsOct 2024: 5 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 4 reportsJun 2025: 2 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520268
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 27, 2023Class IIOpen, Classified

    Complaint of endoscope model becoming lodged in the endotracheal tube connector due to diameter being too large.

    Z-2504-2023 · Root cause: Labeling Change Control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Peeled/Delaminated100
  • Microbial Contamination of Device20
  • Failure to Clean Adequately7
  • Degraded5
  • Detachment of Device or Device Component4
  • Break3
  • Device Markings/Labelling Problem3
  • No Display/Image2

Patient problems

  • No Clinical Signs, Symptoms or Conditions138
  • No Consequences Or Impact To Patient2
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 25 most recent of 141 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 8, 2026Loss of or Failure to Bond

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE TRACHEAL INTUBATION FIBERSCOPE EXHIBITED THE CEMENT GLUE ON THE LIGHT GUIDE (LG) GLASS/LENS COVER IS MISSING. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24003057 ↗
  • MalfunctionDec 18, 2025Corroded; Corroded

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE FIBERSCOPE EXHIBITED SUCTION CYLINDER DEFECTIVE (CORRODED). THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23840861 ↗
  • MalfunctionAug 21, 2025Physical Resistance/Sticking; Physical Resistance/Sticking

    IT WAS REPORTED THE TRACHEAL INTUBATION FIBERSCOPE HAD LOCKED ANGULATION. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22860052 ↗
  • MalfunctionJun 26, 2025Microbial Contamination of Device; Microbial Contamination of Device

    IT WAS REPORTED THAT DURING REPROCESSING, THE SUBJECT FLEXIBLE SCOPE TESTED POSITIVE FOR > 80 COLONY FORMING UNITS (CFUS) OF STENOTROPHOMONAS (XANTHOMONAS) MALTOPHILIA AND PSEUDOMONAS AERUGINOSA. THE BACTERIAL WAS FOUND IN WORKING CHANNEL. THE USER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THE SCOPE COULD HAVE BEEN A CONTRIBUTORY CAUSE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22336258 ↗
  • MalfunctionJun 25, 2025Microbial Contamination of Device; Microbial Contamination of Device

    IT WAS REPORTED THAT THE TRACHEAL INTUBATION FIBERSCOPE TESTED POSITIVE FOR AN UNEXPECTED CONTAMINATION. THE ISSUE WAS FOUND DURING A ROUTINE CULTURE OF THE SCOPE. THE USER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THE SCOPE COULD HAVE BEEN A CONTRIBUTORY CAUSE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22325451 ↗
  • MalfunctionMay 8, 2025Microbial Contamination of Device; Microbial Contamination of Device

    IT WAS REPORTED THAT THE TRACHEAL INTUBATION FIBERSCOPE TESTED POSITIVE FOR A STENOTROPHOMONAS CONTAMINATION. THE ISSUE WAS FOUND DURING A ROUTINE CULTURE OF THE SCOPE. THE USER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THE SCOPE COULD HAVE BEEN A CONTRIBUTORY CAUSE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21988666 ↗
  • MalfunctionMay 8, 2025Microbial Contamination of Device; Microbial Contamination of Device

    IT WAS REPORTED THAT THE SCOPE TESTED POSITIVE FOR AN UNEXPECTED CONTAMINATION. THE ISSUE WAS FOUND DURING A ROUTINE CULTURE OF THE SCOPE. THE USER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THE SCOPE COULD HAVE BEEN A CONTRIBUTORY CAUSE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21984590 ↗
  • MalfunctionMay 8, 2025Microbial Contamination of Device; Microbial Contamination of Device

    IT WAS REPORTED THAT THE SCOPE TESTED POSITIVE FOR AN UNEXPECTED CONTAMINATION. THE ISSUE WAS FOUND DURING A ROUTINE CULTURE OF THE SCOPE. THE USER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THE SCOPE COULD HAVE BEEN A CONTRIBUTORY CAUSE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21984585 ↗
  • MalfunctionMay 5, 2025Microbial Contamination of Device; Detachment of Device or Device Component; Microbial Contamination of Device; Detachment of Device or Device Component

    IT WAS REPORTED THAT, DURING REPROCESSING, THE SUBJECT DEVICE TESTED POSITIVE FOR STENOTROPHOMONAS DESPITE STERILIZATION. THE USER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THE SCOPE COULD HAVE BEEN A CONTRIBUTORY CAUSE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21963812 ↗
  • MalfunctionNov 1, 2024Peeled/Delaminated; Peeled/Delaminated

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE COATING OF THE FIBERSCOPE IG HOSE HAD PEELED OFF. (1MM2 OR MORE). THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20586618 ↗
  • MalfunctionOct 29, 2024Peeled/Delaminated; Peeled/Delaminated

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE TRACHEAL INTUBATION FIBERSCOPE EXHIBITED PEELING OF THE CONNECTING TUBE COATING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20557929 ↗
  • MalfunctionOct 28, 2024Material Disintegration; Peeled/Delaminated; Material Disintegration; Peeled/Delaminated

    IT WAS OBSERVED DURING THE DEVICE INSPECTION THAT THE INTUBATION FIBERSCOPE EXHIBITED A MISSING AREA ON THE CONNECTING TUBE AND PEELING OF THE INSTRUMENT GUIDE TUBE COATING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20553542 ↗
  • MalfunctionOct 25, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING DEVICE EVALUATION THAT FOREIGN MATERIAL CAME OUT OF THE FORCEPS CHANNEL OF THE FIBERSCOPE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20532879 ↗
  • MalfunctionOct 10, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE BENDING SECTION COVER OF THE TRACHEAL INTUBATION FIBERSCOPE HAD FALLEN OFF. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20422861 ↗
  • MalfunctionOct 8, 2024Peeled/Delaminated; Peeled/Delaminated

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE TRACHEAL INTUBATION FIBERSCOPE INSERTION SECTION FLEXIBLE TUBE COATING PEELED OFF. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20397471 ↗
  • MalfunctionSep 17, 2024Peeled/Delaminated; Peeled/Delaminated

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE INTUBATION FIBERSCOPE EXHIBITED PEELING ON THE COATING OF THE IMAGE GUIDE HOSE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20241233 ↗
  • MalfunctionSep 13, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE INTUBATION FIBERSCOPE EXHIBITED RESIDUAL LIQUID AND FOREIGN OBJECTS CAME OUT OF THE FORCEPS CHANNEL. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20211777 ↗
  • MalfunctionSep 5, 2024Peeled/Delaminated; Peeled/Delaminated

    IT WAS OBSERVED, DURING THE DEVICE INSPECTION. THAT THE TRACHEAL INTUBATION FIBERSCOPE EXHIBITED COATING PEELING OFF THE IMAGE GUIDE HOSE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20153252 ↗
  • MalfunctionAug 13, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE TRACHEAL INTUBATION FIBERSCOPE EXHIBITED A FOREIGN MATERIAL INSIDE THE FORCEPS CHANNEL. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19975449 ↗
  • MalfunctionAug 7, 2024Peeled/Delaminated; Peeled/Delaminated

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE TRACHAEL INTUBATION FIBERSCOPE EXHIBITED A PEELED IMAGE GUIDE TUBE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19928006 ↗
  • MalfunctionJul 23, 2024Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THE TRACHEAL INTUBATION FIBERSCOPE COATING WAS PEELING GREATER THAN OR EQUAL TO 1 X 1 MM2. THE ISSUE OCCURRED DURING OTHER EVENTS. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19807559 ↗
  • MalfunctionJul 17, 2024Device Markings/Labelling Problem; Device Markings/Labelling Problem

    IT WAS OBSERVED THAT DURING THE DEVICE INSPECTION, THE FIBERSCOPE EXHIBITED A SOFT TUBE COATING FLOATING (1MM2 OR MORE). THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19762462 ↗
  • MalfunctionJul 3, 2024Device Markings/Labelling Problem; Device Markings/Labelling Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE INTUBATION FIBERSCOPE'S INSERTION SECTION, SOFT TUBE MARKING DISAPPEARED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19668132 ↗
  • MalfunctionJun 11, 2024Peeled/Delaminated; Peeled/Delaminated

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE ENDOSCOPE REPROCESSOR DISTAL END WAS PEELING OFF THE COATING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19512226 ↗
  • MalfunctionMay 23, 2024Material Erosion; Peeled/Delaminated; Material Erosion; Peeled/Delaminated

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE TRACHEAL INTUBATION FIBERSCOPE EXHIBITED FLEXIBLE TUBE COATING PEELING AND DISPLAY MISSING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19384600 ↗