Skip to content

TRACHEAL INTUBATION FIBERSCOPERecalls & reported problems across all models

Olympus Medical Systems Corp. · 7 models in the FDA device registry

Data updated Oct 18, 2026 from official FDA records · Editorial policy

5 open recalls name at least one TRACHEAL INTUBATION FIBERSCOPE model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
616
Deaths reported
0
Injuries reported
3
Recalls
5

We found 7 models of TRACHEAL INTUBATION FIBERSCOPE by Olympus Medical Systems Corp. in the FDA's device registry (GUDID). Together they are named in 616 adverse event reports in the FDA's MAUDE database, received between Jun 13, 2018 and Aug 31, 2026 — classified as 3 injuries, 613 malfunctions.

The most frequently reported specific device problem is “Peeled/Delaminated” (495 reports).

5 FDA recall records name at least one of these models by its device identifier, 5 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 23 reportsDec 2023: 13 reportsJan 2024: 13 reportsFeb 2024: 21 reportsMar 2024: 29 reportsApr 2024: 16 reportsMay 2024: 14 reportsJun 2024: 16 reportsJul 2024: 12 reportsAug 2024: 13 reportsSep 2024: 11 reportsOct 2024: 12 reportsNov 2024: 3 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 3 reportsApr 2025: 0 reportsMay 2025: 5 reportsJun 2025: 3 reportsJul 2025: 2 reportsAug 2025: 3 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 3 reportsJan 2026: 2 reportsFeb 2026: 4 reportsMar 2026: 4 reportsApr 2026: 3 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 2 reportsAug 2026: 3 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202629
Last 36 months. The most recent month may be incomplete.

TRACHEAL INTUBATION FIBERSCOPE recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Sep 28, 2023Class IIOpen, Classified

    Fiberscopes do not have adequate data to support that the suction flow rate meets the device specification.

    Z-0557-2024 · Root cause: Under Investigation by firm · names 2 of these models · Recall details · FDA record ↗

  2. Sep 25, 2023Class IIOpen, Classified

    Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

    Z-0338-2024 · Root cause: Under Investigation by firm · names 2 of these models · Recall details · FDA record ↗

  3. Jul 27, 2023Class IIOpen, Classified

    Complaint of endoscope model becoming lodged in the endotracheal tube connector due to diameter being too large.

    Z-2504-2023 · Root cause: Labeling Change Control · names 5 of these models · Recall details · FDA record ↗

  4. Jul 27, 2023Class IIOpen, Classified

    Complaint of endoscope model becoming lodged in the endotracheal tube connector due to diameter being too large.

    Z-2501-2023 · Root cause: Labeling Change Control · names 2 of these models · Recall details · FDA record ↗

  5. May 26, 2023Class IIOpen, Classified

    Revised OER-Pro, OER-Elite, OER-Mini endoscope reprocessor labeling- the LF-V and LF-P are no longer compatible endoscopes for reprocessing in the OERs. The addendum highlights the following changes: 1. Removal of LF-V and LF-P as compatible with the OER 2. Change conditions for ETO gas sterilization from 12% EtO to 100% EtO 3. Removing the reusable cleaning brush for the manual cleaning. Single Use brushes remain compatible

    Z-2341-2023 · Root cause: Labeling Change Control · Recall details · FDA record ↗

Most reported TRACHEAL INTUBATION FIBERSCOPE problems

Device problems

  • Peeled/Delaminated495
  • Microbial Contamination of Device30
  • Failure to Clean Adequately25
  • Degraded17
  • Detachment of Device or Device Component16
  • Device Reprocessing Problem14
  • Break12
  • Material Split, Cut or Torn7

Patient problems

  • No Clinical Signs, Symptoms or Conditions607
  • No Consequences Or Impact To Patient3
  • No Known Impact Or Consequence To Patient3
  • Foreign Body In Patient2
  • Device Embedded In Tissue or Plaque1

TRACHEAL INTUBATION FIBERSCOPE models

7 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent TRACHEAL INTUBATION FIBERSCOPE adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Olympus Medical Systems Corp. products

All Olympus Medical Systems Corp. devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page