AUTOCLAVABLE CAMERA HEAD OTV-S7H-NA-10ERecalls & reported problems
Olympus Medical Systems Corp. · Class 2
All models of AUTOCLAVABLE CAMERA HEAD (6)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Olympus autoclavable camera head is used in general and plastic surgery laparoscopy. Reports mostly describe poor, blurry, dark or missing images, sometimes with a nicked cord or water getting inside. There are 26 reports naming its identifier, with no deaths and no injuries, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 26
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 26 adverse event reports that list this exact device (DI 04953170141805), received between May 19, 2022 and May 15, 2026 — classified as 26 malfunctions.
The most frequently reported specific device problem is “Poor Quality Image” (9 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Poor Quality Image9
- No Display/Image7
- Loose or Intermittent Connection3
- Leak/Splash2
- Connection Problem1
- Corroded1
- Crack1
- Fluid/Blood Leak1
Patient problems
- No Clinical Signs, Symptoms or Conditions26
Recent adverse event reports
The 25 most recent of 26 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 15, 2026Poor Quality Image
THE CAMERA HEAD WAS RETURNED TO THE SERVICE CENTER WITH A REPORT OF A PUNCTURE IN THE CORD. THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER. DURING INSPECTION OF THE DEVICE, LINES SHOWING ON THE IMAGE DUE TO FAULTY CHARGE COUPLED DEVICE (CCD) WERE FOUND. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25191877 ↗MalfunctionDec 10, 2025Material Separation; Material Separation
NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23768855 ↗MalfunctionMay 27, 2025Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED THAT THE SUBJECT DEVICE HAD A NICK IN THE CORD. THE ISSUE OCCURRED DURING REPROCESSING. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22087788 ↗MalfunctionNov 26, 2024Poor Quality Image; Poor Quality Image
IT WAS REPORTED THAT, THE SUBJECT DEVICE EXHIBITED POOR IMAGE. THE ISSUE WAS OBSERVED DURING SET UP BEFORE THE PATIENT WAS IN ROOM. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20778500 ↗MalfunctionOct 28, 2024Poor Quality Image; Optical Discoloration; Poor Quality Image; Optical Discoloration
IT WAS REPORTED THE AUTOCLAVABLE CAMERA HEAD HAD STARTED WORKING BAD. THE IMAGE BECAME BLURRY AND TURNED GREEN. THE ISSUE OCCURRED DURING AN UNSPECIFIED PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20551253 ↗MalfunctionOct 9, 2024Crack; Leak/Splash; Crack; Leak/Splash
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE EXHIBITED WATER INGRESS INTO THE IMAGING DEVICE, WATER INGRESS INTO THE CABLE, AND MAIN BODY CRACKING. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20411914 ↗MalfunctionSep 4, 2024Poor Quality Image; Poor Quality Image
IT WAS REPORTED THAT, THE SUBJECT DEVICE HAD A DARK IMAGE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20136479 ↗MalfunctionJul 29, 2024Poor Quality Image; Poor Quality Image
IT WAS REPORTED THAT SOMETIMES THERE WAS VIDEO IMAGE DISTORTION OF THE SUBJECT DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19850151 ↗MalfunctionJul 15, 2024Poor Quality Image; Poor Quality Image
IT WAS REPORTED, THAT THE SUBJECT DEVICE DISPLAYED POOR IMAGE QUALITY. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19742739 ↗MalfunctionJul 5, 2024Loose or Intermittent Connection; Loose or Intermittent Connection
IT WAS REPORTED THE SUBJECT DEVICE HAD AN AXIS MISALIGNMENT WITH THE RING ROTATED AROUND. THE ISSUE OCCURRED DURING PRE-USED INSPECTION. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19680348 ↗MalfunctionJun 6, 2024Corroded; Leak/Splash; Corroded; Leak/Splash
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE CAMERA HEAD EXHIBITED CABLE AND IMAGE FLOODING AND SCOPE MOUNT CORROSION. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19480434 ↗MalfunctionMay 13, 2024Poor Quality Image; Poor Quality Image
IT WAS REPORTED, THE CAMERA HEAD HAD A NOISE IMAGE DEFECT. THE ISSUE OCCURRED AT INSPECTION BEFORE USE OF A THERAPEUTIC PROCEDURE. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19296658 ↗MalfunctionApr 2, 2024Poor Quality Image; Poor Quality Image
IT WAS REPORTED, THERE WAS POOR IMAGE QUALITY WITH THE CAMERA HEAD. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19027801 ↗MalfunctionMar 12, 2024Loose or Intermittent Connection; Loose or Intermittent Connection
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE'S COUPLER WAS FOUND TO BE LOOSE. THERE WAS NO REPORT OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18888088 ↗MalfunctionMar 11, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS OBSERVED DURING DEVICE EVALUATION THAT THE SUBJECT DEVICE HAD WATER INVASION FROM FOCUS RING. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18874246 ↗MalfunctionJan 2, 2024Loose or Intermittent Connection; Poor Quality Image; Loose or Intermittent Connection; Poor Quality Image
THE CUSTOMER REPORTED TO OLYMPUS, THE AUTOCLAVABLE CAMERA HEAD HAD POOR CONTACT AND IMAGE WAS BLURRY. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18431474 ↗MalfunctionOct 24, 2023No Display/Image; No Display/Image
THE CUSTOMER REPORTED TO OLYMPUS THAT BEFORE AN UNSPECIFIED PROCEDURE DURING PREPARATION FOR USE, THE AUTOCLAVABLE CAMERA HEAD HAD NO IMAGE. THE DEVICE WAS INSPECTED BEFORE USE. THERE WAS NO PATIENT HARM REPORTED TO OLYMPUS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18000332 ↗MalfunctionJun 9, 2023Output Problem; Output Problem
THE CUSTOMER REPORTED TO OLYMPUS, THE AUTOCLAVABLE CAMERA HEAD WAS UNABLE TO WHITE BALANCE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17099744 ↗MalfunctionJun 7, 2023Image Display Error/Artifact; Image Display Error/Artifact
THE CUSTOMER REPORTED THE OLYMPUS AUTOCLAVABLE CAMERA HEAD CAUSED THE DISPLAYED IMAGE TO PARTIALLY DISAPPEAR. THE ISSUE WAS IDENTIFIED DURING REPROCESSING. NO ADVERSE EFFECTS TO PATIENT WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17083939 ↗MalfunctionFeb 23, 2023No Display/Image; No Display/Image
THE CUSTOMER REPORTED TO OLYMPUS THAT WHILE USING THE AUTOCLAVABLE CAMERA HEAD, THE CUSTOMER OBSERVED COMPLETELY DEAD / DARK IMAGE. THERE WERE NO REPORTS OF PATIENT HARM OR IMPACT ASSOCIATED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16429528 ↗MalfunctionDec 27, 2022No Display/Image; No Display/Image
THE CUSTOMER REPORTED TO OLYMPUS THAT THERE WAS AN IMAGE PROBLEM WHILE USING THE AUTOCLAVABLE CAMERA HEAD. THE ISSUE OCCURRED DURING THE PROCEDURE AND THERE WAS NO PATIENT HARM ASSOCIATED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16050906 ↗MalfunctionNov 9, 2022No Display/Image; No Display/Image
THE ISSUE OCCURRED DURING PREPARATION FOR USE (FUNCTION CHECK). THERE WAS NO DELAY AND THE PROCEDURE WAS COMPLETED WITH A SIMILAR DEVICE. THERE WERE NO OTHER DEVICES REPLACED OR INVOLVED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15760479 ↗MalfunctionAug 16, 2022No Display/Image; No Display/Image
THE CUSTOMER REPORTED THERE WAS NO IMAGE FROM THE SUBJECT DEVICE. THERE WAS NO EFFECT ON A PATIENT DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15237428 ↗MalfunctionJun 27, 2022Connection Problem; Connection Problem
OLYMPUS (OMSC) WAS INFORMED THAT THE CUSTOMER AUTOCLAVABLE CAMERA HEAD WAS RETURNED TO OLYMPUS FOR REPAIR FOR A REPORTED ¿DISCONNECTION OF THE CONNECTOR CONNECTION¿. THE REPORTED PROBLEM WAS FOUND DURING AN UNSPECIFIED EVENT. NO DEATH, INJURY OR HARM WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14836949 ↗MalfunctionMay 24, 2022No Display/Image; No Display/Image
OLYMPUS WAS INFORMED VIA SERVICE INQUIRY THAT THE CUSTOMER AUTOCLAVABLE CAMERA HEAD HAD A NO IMAGE MALFUNCTION, THE IMAGE IS BLACK. THE EVENT WAS FOUND DURING AN UNSPECIFIED EVENT. NO DEATH, INJURY OR HARM WAS REPORTED TO OLYMPUS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14483667 ↗
Other Laparoscope, General & Plastic Surgery devices
- HIGH DEFINITION LCD MONITOR OEV262HOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,937 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-12EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,886 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-10EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,086 reports
- Auto Suture OMS-T10BTCovidien LP · Laparoscope, general & plastic surgery1 recall955 reports
- PDB OMS-PDB1000Covidien LP · Laparoscope, general & plastic surgery859 reports
- VISERA 4K UHD CAMERA CONTROL UNIT OLYMPUS OTV-S400Olympus Medical Systems Corp. · Laparoscope, general & plastic surgery777 reports
- Auto Suture OMS-T10SBCovidien LP · Laparoscope, general & plastic surgery1 recall697 reports
- Spacemaker SMBTTRNDXCovidien LP · Laparoscope, general & plastic surgery590 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page