Skip to content

EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF TYPE H180 (UDI 04953170202285)Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE (20)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Olympus EVIS EXERA II is a gastrointestinal videoscope. Many of the 135 reports describe the scope not being cleaned adequately or being contaminated with microbes, and some describe blurry or noisy images. There are no deaths and 11 injuries, and reports rose from 9 in the previous 12 months to 26 in the latest 12 months. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
134
Deaths reported
0
Injuries reported
11
Recalls naming it
0

Summary

The FDA's MAUDE database contains 134 adverse event reports that list this exact device (DI 04953170202285), received between May 15, 2018 and Aug 25, 2026 — classified as 11 injuries, 123 malfunctions.

The most frequently reported specific device problem is “Failure to Clean Adequately” (45 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 1 reportsJan 2024: 2 reportsFeb 2024: 5 reportsMar 2024: 3 reportsApr 2024: 2 reportsMay 2024: 2 reportsJun 2024: 5 reportsJul 2024: 6 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 1 reportsMar 2025: 2 reportsApr 2025: 1 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 2 reportsJan 2026: 3 reportsFeb 2026: 3 reportsMar 2026: 6 reportsApr 2026: 3 reportsMay 2026: 6 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Clean Adequately45
  • Microbial Contamination of Device44
  • Poor Quality Image12
  • Adverse Event Without Identified Device or Use Problem11
  • Device Reprocessing Problem8
  • Corroded5
  • Erratic or Intermittent Display4
  • No Display/Image4

Patient problems

  • No Clinical Signs, Symptoms or Conditions92
  • No Known Impact Or Consequence To Patient22
  • Injury9
  • No Consequences Or Impact To Patient9
  • Laceration(s)1
  • Perforation1
  • Pulmonary Embolism1

Recent adverse event reports

The 25 most recent of 134 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 25, 2026Failure to Clean Adequately

    No additional information received from customer.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50422530 ↗
  • MalfunctionJun 23, 2026Corroded

    THE CUSTOMER RETURNED THE GASTROINTESTINAL VIDEOSCOPE TO THE SERVICE CENTER FOR EVALUATION RELATED TO A SEPARATE ISSUE. DURING THE EVALUATION OF THE DEVICE, CORROSION ON THE LIGHT GUIDE LENS WAS IDENTIFIED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25593252 ↗
  • MalfunctionMay 27, 2026Poor Quality Image

    THE CUSTOMER RETURNED THE GASTROINTESTINAL VIDEOSCOPE, WHICH IS AN OLYMPUS ASSET WITH SEPARATE CASE. DURING THE EVALUATION OF THE DEVICE, BLURRY IMAGE WAS OBSERVED. THE SERVICE REPAIR TECHNICIAN INDICATED THAT THE MOST LIKELY CAUSE OF THE BLURRY IMAGE WAS DUE TO POLISHED OBJECTIVE LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25300723 ↗
  • MalfunctionMay 20, 2026Erratic or Intermittent Display

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE HAD A NOISY IMAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25233841 ↗
  • MalfunctionMay 20, 2026Poor Quality Image

    IT WAS REPORTED THAT THE GASTROINTESTINAL VIDEOSCOPE PRODUCED A BLURRY IMAGE. THE ISSUE WAS FOUND DURING PRE-USE INSPECTION AND THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25231989 ↗
  • MalfunctionMay 15, 2026Corroded

    THE GASTROINTESTINAL VIDEOSCOPE WAS RETURNED TO OLYMPUS WITH A REPORTED COMPLAINT OF MISALIGNMENT OF ANGLES AND PLAY IN THE KNOBS, DURING PREVENTIVE MAINTENANCE; THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, AN MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER, THE NOZZLE WAS CORRODED. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25184136 ↗
  • MalfunctionMay 13, 2026No Display/Image

    THE GASTROINTESTINAL VIDEOSCOPE WAS RETURNED TO OLYMPUS WITH A REPORTED COMPLAINT OF PROBLEM WITH THE IMAGE; THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, AN MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER, THE GASTROINTESTINAL VIDEOSCOPE HAD NO IMAGE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25159273 ↗
  • MalfunctionMay 12, 2026Erratic or Intermittent Display; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE VIDEOSCOPE HAD THE UNIVERSAL CORD BE STICKY AND THE IMAGE HAD LINEAR SCRATCHES CAUSING A NOISY IMAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25150497 ↗
  • MalfunctionApr 8, 2026Corroded; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED CORROSION ON NOZZLE AND DIRTY ELECTRONIC CONNECTOR. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24815679 ↗
  • MalfunctionApr 6, 2026Poor Quality Image

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED BLURRY IMAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24792802 ↗
  • MalfunctionApr 2, 2026No Display/Image

    IT WAS OBSERVED DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE HAD NO IMAGE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24766702 ↗
  • MalfunctionMar 30, 2026Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE PRESENTED STICKY UNIVERSAL CORD AND SCOPE COVER UNIT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24717702 ↗
  • MalfunctionMar 24, 2026Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE C-COVER OF THE GASTROINTESTINAL VIDEOSCOPE HAD FOREIGN MATERIAL. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24675422 ↗
  • MalfunctionMar 20, 2026Poor Quality Image

    IT WAS REPORTED THE GASTROINTESTINAL VIDEOSCOPE HAD IMAGE INTERFERENCE. THERE WAS NOISE IN THE IMAGE AND THE IMAGE RECORDED ON ITS OWN. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24649272 ↗
  • MalfunctionMar 16, 2026Corroded

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE HAD CORROSION IN THE AIR/WATER COMPARTMENT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24603216 ↗
  • MalfunctionMar 13, 2026Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED A STICKY CONNECTING TUBE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24592354 ↗
  • MalfunctionMar 2, 2026Corroded; No Display/Image

    IT WAS OBSERVED DURING DEVICE EVALUATION THAT THE GASTROINTESTINAL VIDEOSCOPE HAD A CORRODED PLASTIC DISTAL END COVER AND DID NOT PRODUCE AN IMAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24485598 ↗
  • MalfunctionFeb 26, 2026Poor Quality Image

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE VIDEOSCOPE HAD IMAGE INTERFERENCE OR NOISINESS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24456149 ↗
  • MalfunctionFeb 13, 2026Erratic or Intermittent Display

    IT WAS OBSERVED DURING DEVICE EVALUATION, THAT THE GASTROINTESTINAL VIDEOSCOPE HAD IMAGE INTERFERENCE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24335698 ↗
  • MalfunctionFeb 4, 2026Device Reprocessing Problem; Erratic or Intermittent Display

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24258824 ↗
  • MalfunctionJan 21, 2026Poor Quality Image

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24134383 ↗
  • MalfunctionJan 8, 2026Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE NOZZLE OF THE GASTROINTESTINAL VIDEOSCOPE HAD FIBROUS FOREIGN MATERIAL. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24004481 ↗
  • MalfunctionJan 2, 2026No Display/Image

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE HAD NO IMAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23943981 ↗
  • MalfunctionDec 30, 2025Poor Quality Image; Poor Quality Image

    NO NEW INFORMATION RECEIVED BY THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23926860 ↗
  • MalfunctionDec 18, 2025Failure to Clean Adequately; Failure to Clean Adequately

    NO NEW INFORMATION RECEIVED BY THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23838130 ↗

Other Colonoscope and accessories, flexible/rigid devices

All Colonoscope and accessories, flexible/rigid devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page