Skip to content

EVIS EXERA II COLONOVIDEOSCOPE OLYMPUS CF TYPE H180AL (UDI 04953170339608)Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of EVIS EXERA II COLONOVIDEOSCOPE (28)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Olympus colonoscope is a flexible scope used to view the colon. The 70 reports that name its identifier mostly describe cleaning problems and contamination found during servicing, along with loss of image or light, with no deaths and 1 injury mentioned. There are 2 open recalls, one about missing adhesive that protected repaired scopes from vibration and temperature changes and one about repair returns with a wet channel that could become contaminated.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
72
Deaths reported
0
Injuries reported
1
Recalls naming it
2

Summary

The FDA's MAUDE database contains 72 adverse event reports that list this exact device (DI 04953170339608), received between Mar 9, 2019 and Sep 24, 2026 — classified as 1 injury, 71 malfunctions.

The most frequently reported specific device problem is “Failure to Clean Adequately” (32 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 1 reportsMar 2024: 2 reportsApr 2024: 2 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 1 reportsJan 2025: 1 reportsFeb 2025: 1 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 3 reportsAug 2025: 5 reportsSep 2025: 2 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 6 reportsFeb 2026: 3 reportsMar 2026: 1 reportsApr 2026: 2 reportsMay 2026: 5 reportsJun 2026: 3 reportsJul 2026: 2 reportsAug 2026: 1 reportsSep 2026: 2 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jan 11, 2024Class IIOpen, Classified

    Specific repaired colonovideoscopes were assembled without an adhesive which protects them from certain physical conditions such as high levels of vibration, fluctuation in temperature or severe shock loads.

    Z-1205-2024 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

  2. Sep 25, 2023Class IIOpen, Classified

    Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

    Z-0329-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Failure to Clean Adequately32
  • Microbial Contamination of Device15
  • Device Reprocessing Problem9
  • No Display/Image5
  • Contamination3
  • Corroded3
  • Communication or Transmission Problem2
  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions64
  • No Known Impact Or Consequence To Patient3
  • No Consequences Or Impact To Patient2
  • Abdominal Pain1
  • Bowel Perforation1
  • Diaphoresis1
  • Discomfort1
  • Insufficient Information1

Recent adverse event reports

The 25 most recent of 72 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 24, 2026Failure to Clean Adequately

    The customer returned the colonovideoscope to the service center for a separate issue. During the device analysis, the technical assistance indicates that a black foreign material at nozzle was found. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50876998 ↗
  • MalfunctionSep 1, 2026Corroded

    During onsite inspection of the colonovideoscope, it was found that nozzle had corrosion. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50544476 ↗
  • MalfunctionAug 25, 2026Corroded

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50435173 ↗
  • MalfunctionJul 28, 2026No Display/Image; Poor Quality Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50057828 ↗
  • MalfunctionJul 27, 2026No Display/Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50048218 ↗
  • MalfunctionJun 25, 2026Failure to Clean Adequately

    THE CUSTOMER RETURNED THE COLONOVIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS THE SERVICE TECHNICIAN, INDICATES THAT A DIRT IN THE LENSES WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25620556 ↗
  • MalfunctionJun 23, 2026Device Reprocessing Problem

    THE CUSTOMER RETURNED THE COLONOVIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE SERVICE REPAIR TECHNICIAN INDICATES THAT AN AIR WATER CYLINDER AND TUBE HAS FOREIGN MATERIAL WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25598081 ↗
  • MalfunctionJun 2, 2026Output Problem

    IT WAS REPORTED THAT THE COLONOVIDEOSCOPE EXHIBITED NO ILLUMINATION IN THE THIRD LIGHT GUIDE LENS. THE ISSUE OCCURRED DURING THE INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25365584 ↗
  • MalfunctionMay 26, 2026No Display/Image

    IT WAS REPORTED THAT THE COLONOVIDEOSCOPE HAD NO IMAGE. THIS ISSUE WAS FOUND DURING INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25283986 ↗
  • MalfunctionMay 21, 2026Failure to Clean Adequately

    THE COLONOVIDEOSCOPE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT OF LEAKY CONTROL HANDLE AND FOGGING OF OPTICS. THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER. DURING INSPECTION OF THE DEVICE, DIRT IN THE FIELD OF VIEW WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25247735 ↗
  • MalfunctionMay 21, 2026Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED FOREIGN OBJECTS INSIDE THE AIR/WATER CYLINDER AND AIR/WATER TUBE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25247490 ↗
  • MalfunctionMay 21, 2026Communication or Transmission Problem

    IT WAS REPORTED THAT THE COLONOVIDEOSCOPE HAD A COMMUNICATION PROBLEM. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25244686 ↗
  • MalfunctionMay 8, 2026Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE HAD FOREIGN OBJECTS IN UNIVERSAL CORD. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25129688 ↗
  • MalfunctionApr 17, 2026Device Reprocessing Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED FOREIGN MATERIAL ON BIOPSY CHANNEL. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24912508 ↗
  • MalfunctionApr 7, 2026No Display/Image

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED NO IMAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24808928 ↗
  • MalfunctionMar 9, 2026Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED FOREIGN OBJECT IN THE AIR WATER CYLINDER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24552057 ↗
  • MalfunctionFeb 27, 2026Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE HAD A BLACK FOREIGN MATERIAL. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24473687 ↗
  • MalfunctionFeb 3, 2026Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED LIGHT GUIDE ROD LENS WITH FOREIGN MATERIAL. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24243729 ↗
  • MalfunctionFeb 2, 2026Overheating of Device; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED FOREIGN OBJECTS COMING OUT OF THE DISTAL END, AND THE C-COVER SHOWED A BURN. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24235276 ↗
  • MalfunctionJan 28, 2026Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE HAD A FOREIGN OBJECT AT AIR/WATER TUBE AND AIR/WATER CYLINDER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24202725 ↗
  • MalfunctionJan 20, 2026Device Reprocessing Problem; Corroded

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE HAD WHITE FOREIGN MATERIAL IN AIR/WATER CYLINDER AND J-TUBE, AND THAT CONNECTING TUBE HAD CORROSION. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24114346 ↗
  • MalfunctionJan 20, 2026Obstruction of Flow; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED FOREIGN OBJECTS FROM AUXILIARY CHANNEL AND CLOGGING OF AUXILIARY CHANNEL IN CONTROL UNIT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24112945 ↗
  • MalfunctionJan 15, 2026Device Reprocessing Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED FOREIGN OBJECTS IN THE AIR/WATER CYLINDER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24083084 ↗
  • MalfunctionJan 14, 2026Communication or Transmission Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE HAD AN ERROR E217 (SCOPE ERROR). THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24069995 ↗
  • MalfunctionJan 7, 2026Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE HAD FOREIGN MATERIAL ON AIR-WATER TUBE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23988059 ↗

Other Gastroscope and accessories, flexible/rigid devices

All Gastroscope and accessories, flexible/rigid devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page