EVIS EXERA II COLONOVIDEOSCOPE OLYMPUS CF TYPE H180AL (UDI 04953170202445)Recalls & reported problems
Olympus Medical Systems Corp. · Class 2
All models of EVIS EXERA II COLONOVIDEOSCOPE (28)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Olympus EVIS Exera II colonoscope is a flexible camera used to look inside the bowel during procedures. Most of the 37 reports describe cleaning and reprocessing trouble — leftover material found inside the suction cylinder, channels, tubing, and knobs — plus a few cases of a flickering image, no lamp light, and melted or burnt plastic at the instrument channel port. No deaths and no injuries were reported, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 37
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 37 adverse event reports that list this exact device (DI 04953170202445), received between Aug 4, 2020 and Mar 4, 2026 — classified as 37 malfunctions.
The most frequently reported specific device problem is “Device Reprocessing Problem” (17 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device Reprocessing Problem17
- Failure to Clean Adequately13
- Mechanical Problem2
- Crack1
- Detachment of Device or Device Component1
- Erratic or Intermittent Display1
- Material Twisted/Bent1
- Mechanical Jam1
Patient problems
- No Clinical Signs, Symptoms or Conditions31
- No Consequences Or Impact To Patient3
- No Patient Involvement3
Recent adverse event reports
The 25 most recent of 37 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 4, 2026Failure to Clean Adequately
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED FOREIGN MATERIAL IN SUCTION CYLINDER. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24510729 ↗MalfunctionFeb 18, 2026Erratic or Intermittent Display; No Device Output
IT WAS REPORTED THAT THE COLONVIDEOSCOPE HAD AN ISSUE WHERE THE IMAGE WAS FLICKERING ON AND OFF AND THE LAMP WAS NOT ILLUMINATING. THE ISSUE OCCURRED DURING A DIAGNOSTIC UPPER ENDOSCOPY PROCEDURE, WHICH WAS COMPLETED USING THE SAME DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24377299 ↗MalfunctionDec 29, 2025Microbial Contamination of Device; Microbial Contamination of Device
NO NEW INFORMATION RECEIVED BY THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23905668 ↗MalfunctionDec 5, 2024Failure to Clean Adequately; Failure to Clean Adequately
IT WAS REPORTED THAT THERE WAS FOREIGN MATERIAL EXITING FROM THE CHANNEL (INCLUDING AUXILIARY WATER CHANNEL). THE ISSUE OCCURRED DURING AN UNSPECIFIED PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20852179 ↗MalfunctionNov 5, 2024Thermal Decomposition of Device; Melted; Thermal Decomposition of Device; Melted
IT WAS REPORTED THE COLONOVIDEOSCOPE HAD BURNING AND MELTING OF THE PLASTIC DISTAL END COVER AT THE FORCEPS PORT. IT IS UNSPECIFIED WHEN THIS ISSUE OCCURRED. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20609393 ↗MalfunctionSep 23, 2024Failure to Clean Adequately; Failure to Clean Adequately
IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE CONNECTING TUBE EXHIBITED FOREIGN MATERIAL. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20279265 ↗MalfunctionSep 18, 2024Failure to Clean Adequately; Failure to Clean Adequately
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE EVIS EXERA II COLONOVIDEOSCOPE EXHIBITED FOREIGN MATERIAL ON THE SWITCH AND SWITCH BOX AS WELL AS ON THE UP/DOWN AND RIGHT/LEFT KNOBS. FOREIGN MATERIAL COME OUT FROM THE JET TUBE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20250059 ↗MalfunctionAug 22, 2024Device Reprocessing Problem; Device Reprocessing Problem
IT WAS REPORTED THAT WHILE CONDUCTING THE FACILITY REVIEW SUMMARY ON THE EVIS EXERA II COLONOVIDEOSCOPE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20039210 ↗MalfunctionMay 6, 2024Failure to Clean Adequately; Failure to Clean Adequately
IT WAS OBSERVED DURING THE DEVICE INSPECTION, THE ADHESIVE ON THE BENDING SECTION COVER AND THE BENDING SECTION COVER OF THE COLONOVIDEOSCOPE HAD FOREIGN OBJECTS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19247848 ↗MalfunctionMar 7, 2024Device Reprocessing Problem; Device Reprocessing Problem
THE CUSTOMER REPORTED THE ENDOSCOPE WAS INCORRECTLY REPROCESSED. THE SCOPE WAS REPROCESSED USING AN ENDOSCOPE REPROCESSOR, WHERE THE INTERNAL AIR AND WATER FILTERS HAD NOT BEEN REPLACED SINCE THE DATE OF INSTALL. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18855335 ↗MalfunctionMar 7, 2024Device Reprocessing Problem; Device Reprocessing Problem
THE CUSTOMER REPORTED THE ENDOSCOPE WAS INCORRECTLY REPROCESSED. THE SCOPE WAS REPROCESSED USING AN ENDOSCOPE REPROCESSOR, WHERE THE INTERNAL AIR AND WATER FILTERS HAD NOT BEEN REPLACED SINCE THE DATE OF INSTALL. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18854947 ↗MalfunctionMar 6, 2024Device Reprocessing Problem; Device Reprocessing Problem
THE CUSTOMER REPORTED THE ENDOSCOPE WAS INCORRECTLY REPROCESSED. THE SCOPE WAS REPROCESSED USING AN ENDOSCOPE REPROCESSOR, WHERE THE INTERNAL AIR AND WATER FILTERS HAD NOT BEEN REPLACED SINCE THE DATE OF INSTALL. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18852746 ↗MalfunctionMar 6, 2024Device Reprocessing Problem; Device Reprocessing Problem
THE CUSTOMER REPORTED THE ENDOSCOPE WAS INCORRECTLY REPROCESSED. THE SCOPE WAS REPROCESSED USING AN ENDOSCOPE REPROCESSOR, WHERE THE INTERNAL AIR AND WATER FILTERS HAD NOT BEEN REPLACED SINCE THE DATE OF INSTALL. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18852037 ↗MalfunctionMar 6, 2024Device Reprocessing Problem; Device Reprocessing Problem
THE CUSTOMER REPORTED THE ENDOSCOPE WAS INCORRECTLY REPROCESSED. THE SCOPE WAS REPROCESSED USING AN ENDOSCOPE REPROCESSOR, WHERE THE INTERNAL AIR AND WATER FILTERS HAD NOT BEEN REPLACED SINCE THE DATE OF INSTALL. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18851875 ↗MalfunctionMar 6, 2024Device Reprocessing Problem; Device Reprocessing Problem
THE CUSTOMER REPORTED THE ENDOSCOPE WAS INCORRECTLY REPROCESSED. THE SCOPE WAS REPROCESSED USING AN ENDOSCOPE REPROCESSOR, WHERE THE INTERNAL AIR AND WATER FILTERS HAD NOT BEEN REPLACED SINCE THE DATE OF INSTALL. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18851755 ↗MalfunctionMar 6, 2024Device Reprocessing Problem; Device Reprocessing Problem
THE CUSTOMER REPORTED THE ENDOSCOPE WAS INCORRECTLY REPROCESSED. THE SCOPE WAS REPROCESSED USING AN ENDOSCOPE REPROCESSOR, WHERE THE INTERNAL AIR AND WATER FILTERS HAD NOT BEEN REPLACED SINCE THE DATE OF INSTALL. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18849950 ↗MalfunctionMar 6, 2024Device Reprocessing Problem; Device Reprocessing Problem
THE CUSTOMER REPORTED THE ENDOSCOPE WAS INCORRECTLY REPROCESSED. THE SCOPE WAS REPROCESSED USING AN ENDOSCOPE REPROCESSOR, WHERE THE INTERNAL AIR AND WATER FILTERS HAD NOT BEEN REPLACED SINCE THE DATE OF INSTALL. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18849099 ↗MalfunctionMar 6, 2024Device Reprocessing Problem; Device Reprocessing Problem
THE CUSTOMER REPORTED THE ENDOSCOPE WAS INCORRECTLY REPROCESSED. THE SCOPE WAS REPROCESSED USING AN ENDOSCOPE REPROCESSOR, WHERE THE INTERNAL AIR AND WATER FILTERS HAD NOT BEEN REPLACED SINCE THE DATE OF INSTALL. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18848774 ↗MalfunctionMar 6, 2024Device Reprocessing Problem; Device Reprocessing Problem
THE CUSTOMER REPORTED THE ENDOSCOPE WAS INCORRECTLY REPROCESSED. THE SCOPE WAS REPROCESSED USING AN ENDOSCOPE REPROCESSOR, WHERE THE INTERNAL AIR AND WATER FILTERS HAD NOT BEEN REPLACED SINCE THE DATE OF INSTALL. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18848772 ↗MalfunctionMar 6, 2024Device Reprocessing Problem; Device Reprocessing Problem
THE CUSTOMER REPORTED THE ENDOSCOPE WAS INCORRECTLY REPROCESSED. THE SCOPE WAS REPROCESSED USING AN ENDOSCOPE REPROCESSOR, WHERE THE INTERNAL AIR AND WATER FILTERS HAD NOT BEEN REPLACED SINCE THE DATE OF INSTALL. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18848279 ↗MalfunctionMar 6, 2024Device Reprocessing Problem; Device Reprocessing Problem
THE CUSTOMER REPORTED THE ENDOSCOPE WAS INCORRECTLY REPROCESSED. THE SCOPE WAS REPROCESSED USING AN ENDOSCOPE REPROCESSOR, WHERE THE INTERNAL AIR AND WATER FILTERS HAD NOT BEEN REPLACED SINCE THE DATE OF INSTALL. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18847592 ↗MalfunctionMar 6, 2024Device Reprocessing Problem; Device Reprocessing Problem
THE CUSTOMER REPORTED THE ENDOSCOPE WAS INCORRECTLY REPROCESSED. THE SCOPE WAS REPROCESSED USING AN ENDOSCOPE REPROCESSOR, WHERE THE INTERNAL AIR AND WATER FILTERS HAD NOT BEEN REPLACED SINCE THE DATE OF INSTALL. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18846774 ↗MalfunctionMar 6, 2024Device Reprocessing Problem; Device Reprocessing Problem
THE CUSTOMER REPORTED THE ENDOSCOPE WAS INCORRECTLY REPROCESSED. THE SCOPE WAS REPROCESSED USING AN ENDOSCOPE REPROCESSOR, WHERE THE INTERNAL AIR AND WATER FILTERS HAD NOT BEEN REPLACED SINCE THE DATE OF INSTALL. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18846126 ↗MalfunctionDec 22, 2023Failure to Clean Adequately; Failure to Clean Adequately
THE CUSTOMER REPORTED TO OLYMPUS, THE EVIS EXERA II COLONOVIDEOSCOPE HAD AN ANGULATION WIRE AND UP/DOWN KNOB ISSUE. THE ISSUE WAS FOUND DURING REPROCESSING. THE DEVICE WAS RETURNED FOR EVALUATION. DURING THE DEVICE EVALUATION, IT WAS FOUND THAT THE BENDING SECTION COVER HAD FOREIGN OBJECTS. THERE WERE NO REPORTS OF PATIENT HARM. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18382822 ↗MalfunctionNov 24, 2023Failure to Clean Adequately; Failure to Clean Adequately
THE OLYMPUS FIELD SERVICE ENGINEER (ON BEHALF OF THE CUSTOMER) REPORTED TO OLYMPUS THAT THE COLONOVIDEOSCOPE HAD ANGULATION PLAY AND NOT ENOUGH AND INSERTION TUBE DISCOLORED/STAIN/WRINKLED. CUSTOMER FOUND THE PROBLEM IN THEIR REPROCESSING TIME. NEEDED DETAILED INSPECTION OF THE SUBJECT DEVICE. THERE WAS NO REPORT OF PATIENT HARM ASSOCIATED WITH THE EVENT. THE DEVICE WAS RETURNED FOR EVALUATION. DURING THE DEVICE EVALUATION, FOUND THE SCOPE HAD FOREIGN OBJECTS IN THE FOLLOWING PARTS: THE PLASTIC DISTAL END COVER, BENDING SECTION COVER AND IN THE CONNECTING TUBE. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18206370 ↗
Other Gastroscope and accessories, flexible/rigid devices
- PENTAX Medical EG29-i10cHoya Corporation · Gastroscope And Accessories, Flexible/Rigid1,623 reports
- EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-1TH190Olympus Medical Systems Corp. · Gastroscope And Accessories, Flexible/Rigid4 recalls6 deaths1,025 reports
- EVIS EXERA lll GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-XP190NOlympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid1 recall559 reports
- EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE GIF-Q165Olympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid349 reports
- EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF TYPE H180Olympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid1 death302 reports
- PENTAX EG27-i10Hoya Corporation · Gastroscope and accessories, flexible/rigid261 reports
- EVIS EXERA II COLONOVIDEOSCOPE OLYMPUS CF TYPE H180ALOlympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid2 deaths252 reports
- BALLOON MAJ-1351Olympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid174 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page