EVIS EXERA II COLONOVIDEOSCOPE OLYMPUS CF TYPE H180AL (UDI 04953170202445)Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of EVIS EXERA II COLONOVIDEOSCOPE (28)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Olympus EVIS Exera II colonoscope is a flexible camera used to look inside the bowel during procedures. Most of the 37 reports describe cleaning and reprocessing trouble — leftover material found inside the suction cylinder, channels, tubing, and knobs — plus a few cases of a flickering image, no lamp light, and melted or burnt plastic at the instrument channel port. No deaths and no injuries were reported, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
37
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 37 adverse event reports that list this exact device (DI 04953170202445), received between Aug 4, 2020 and Mar 4, 2026 — classified as 37 malfunctions.

The most frequently reported specific device problem is “Device Reprocessing Problem” (17 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 14 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 2 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202614
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Reprocessing Problem17
  • Failure to Clean Adequately13
  • Mechanical Problem2
  • Crack1
  • Detachment of Device or Device Component1
  • Erratic or Intermittent Display1
  • Material Twisted/Bent1
  • Mechanical Jam1

Patient problems

  • No Clinical Signs, Symptoms or Conditions31
  • No Consequences Or Impact To Patient3
  • No Patient Involvement3

Recent adverse event reports

The 25 most recent of 37 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 4, 2026Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED FOREIGN MATERIAL IN SUCTION CYLINDER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24510729 ↗
  • MalfunctionFeb 18, 2026Erratic or Intermittent Display; No Device Output

    IT WAS REPORTED THAT THE COLONVIDEOSCOPE HAD AN ISSUE WHERE THE IMAGE WAS FLICKERING ON AND OFF AND THE LAMP WAS NOT ILLUMINATING. THE ISSUE OCCURRED DURING A DIAGNOSTIC UPPER ENDOSCOPY PROCEDURE, WHICH WAS COMPLETED USING THE SAME DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24377299 ↗
  • MalfunctionDec 29, 2025Microbial Contamination of Device; Microbial Contamination of Device

    NO NEW INFORMATION RECEIVED BY THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23905668 ↗
  • MalfunctionDec 5, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS REPORTED THAT THERE WAS FOREIGN MATERIAL EXITING FROM THE CHANNEL (INCLUDING AUXILIARY WATER CHANNEL). THE ISSUE OCCURRED DURING AN UNSPECIFIED PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20852179 ↗
  • MalfunctionNov 5, 2024Thermal Decomposition of Device; Melted; Thermal Decomposition of Device; Melted

    IT WAS REPORTED THE COLONOVIDEOSCOPE HAD BURNING AND MELTING OF THE PLASTIC DISTAL END COVER AT THE FORCEPS PORT. IT IS UNSPECIFIED WHEN THIS ISSUE OCCURRED. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20609393 ↗
  • MalfunctionSep 23, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE CONNECTING TUBE EXHIBITED FOREIGN MATERIAL. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20279265 ↗
  • MalfunctionSep 18, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE EVIS EXERA II COLONOVIDEOSCOPE EXHIBITED FOREIGN MATERIAL ON THE SWITCH AND SWITCH BOX AS WELL AS ON THE UP/DOWN AND RIGHT/LEFT KNOBS. FOREIGN MATERIAL COME OUT FROM THE JET TUBE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20250059 ↗
  • MalfunctionAug 22, 2024Device Reprocessing Problem; Device Reprocessing Problem

    IT WAS REPORTED THAT WHILE CONDUCTING THE FACILITY REVIEW SUMMARY ON THE EVIS EXERA II COLONOVIDEOSCOPE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20039210 ↗
  • MalfunctionMay 6, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THE ADHESIVE ON THE BENDING SECTION COVER AND THE BENDING SECTION COVER OF THE COLONOVIDEOSCOPE HAD FOREIGN OBJECTS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19247848 ↗
  • MalfunctionMar 7, 2024Device Reprocessing Problem; Device Reprocessing Problem

    THE CUSTOMER REPORTED THE ENDOSCOPE WAS INCORRECTLY REPROCESSED. THE SCOPE WAS REPROCESSED USING AN ENDOSCOPE REPROCESSOR, WHERE THE INTERNAL AIR AND WATER FILTERS HAD NOT BEEN REPLACED SINCE THE DATE OF INSTALL. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18855335 ↗
  • MalfunctionMar 7, 2024Device Reprocessing Problem; Device Reprocessing Problem

    THE CUSTOMER REPORTED THE ENDOSCOPE WAS INCORRECTLY REPROCESSED. THE SCOPE WAS REPROCESSED USING AN ENDOSCOPE REPROCESSOR, WHERE THE INTERNAL AIR AND WATER FILTERS HAD NOT BEEN REPLACED SINCE THE DATE OF INSTALL. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18854947 ↗
  • MalfunctionMar 6, 2024Device Reprocessing Problem; Device Reprocessing Problem

    THE CUSTOMER REPORTED THE ENDOSCOPE WAS INCORRECTLY REPROCESSED. THE SCOPE WAS REPROCESSED USING AN ENDOSCOPE REPROCESSOR, WHERE THE INTERNAL AIR AND WATER FILTERS HAD NOT BEEN REPLACED SINCE THE DATE OF INSTALL. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18852746 ↗
  • MalfunctionMar 6, 2024Device Reprocessing Problem; Device Reprocessing Problem

    THE CUSTOMER REPORTED THE ENDOSCOPE WAS INCORRECTLY REPROCESSED. THE SCOPE WAS REPROCESSED USING AN ENDOSCOPE REPROCESSOR, WHERE THE INTERNAL AIR AND WATER FILTERS HAD NOT BEEN REPLACED SINCE THE DATE OF INSTALL. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18852037 ↗
  • MalfunctionMar 6, 2024Device Reprocessing Problem; Device Reprocessing Problem

    THE CUSTOMER REPORTED THE ENDOSCOPE WAS INCORRECTLY REPROCESSED. THE SCOPE WAS REPROCESSED USING AN ENDOSCOPE REPROCESSOR, WHERE THE INTERNAL AIR AND WATER FILTERS HAD NOT BEEN REPLACED SINCE THE DATE OF INSTALL. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18851875 ↗
  • MalfunctionMar 6, 2024Device Reprocessing Problem; Device Reprocessing Problem

    THE CUSTOMER REPORTED THE ENDOSCOPE WAS INCORRECTLY REPROCESSED. THE SCOPE WAS REPROCESSED USING AN ENDOSCOPE REPROCESSOR, WHERE THE INTERNAL AIR AND WATER FILTERS HAD NOT BEEN REPLACED SINCE THE DATE OF INSTALL. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18851755 ↗
  • MalfunctionMar 6, 2024Device Reprocessing Problem; Device Reprocessing Problem

    THE CUSTOMER REPORTED THE ENDOSCOPE WAS INCORRECTLY REPROCESSED. THE SCOPE WAS REPROCESSED USING AN ENDOSCOPE REPROCESSOR, WHERE THE INTERNAL AIR AND WATER FILTERS HAD NOT BEEN REPLACED SINCE THE DATE OF INSTALL. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18849950 ↗
  • MalfunctionMar 6, 2024Device Reprocessing Problem; Device Reprocessing Problem

    THE CUSTOMER REPORTED THE ENDOSCOPE WAS INCORRECTLY REPROCESSED. THE SCOPE WAS REPROCESSED USING AN ENDOSCOPE REPROCESSOR, WHERE THE INTERNAL AIR AND WATER FILTERS HAD NOT BEEN REPLACED SINCE THE DATE OF INSTALL. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18849099 ↗
  • MalfunctionMar 6, 2024Device Reprocessing Problem; Device Reprocessing Problem

    THE CUSTOMER REPORTED THE ENDOSCOPE WAS INCORRECTLY REPROCESSED. THE SCOPE WAS REPROCESSED USING AN ENDOSCOPE REPROCESSOR, WHERE THE INTERNAL AIR AND WATER FILTERS HAD NOT BEEN REPLACED SINCE THE DATE OF INSTALL. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18848774 ↗
  • MalfunctionMar 6, 2024Device Reprocessing Problem; Device Reprocessing Problem

    THE CUSTOMER REPORTED THE ENDOSCOPE WAS INCORRECTLY REPROCESSED. THE SCOPE WAS REPROCESSED USING AN ENDOSCOPE REPROCESSOR, WHERE THE INTERNAL AIR AND WATER FILTERS HAD NOT BEEN REPLACED SINCE THE DATE OF INSTALL. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18848772 ↗
  • MalfunctionMar 6, 2024Device Reprocessing Problem; Device Reprocessing Problem

    THE CUSTOMER REPORTED THE ENDOSCOPE WAS INCORRECTLY REPROCESSED. THE SCOPE WAS REPROCESSED USING AN ENDOSCOPE REPROCESSOR, WHERE THE INTERNAL AIR AND WATER FILTERS HAD NOT BEEN REPLACED SINCE THE DATE OF INSTALL. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18848279 ↗
  • MalfunctionMar 6, 2024Device Reprocessing Problem; Device Reprocessing Problem

    THE CUSTOMER REPORTED THE ENDOSCOPE WAS INCORRECTLY REPROCESSED. THE SCOPE WAS REPROCESSED USING AN ENDOSCOPE REPROCESSOR, WHERE THE INTERNAL AIR AND WATER FILTERS HAD NOT BEEN REPLACED SINCE THE DATE OF INSTALL. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18847592 ↗
  • MalfunctionMar 6, 2024Device Reprocessing Problem; Device Reprocessing Problem

    THE CUSTOMER REPORTED THE ENDOSCOPE WAS INCORRECTLY REPROCESSED. THE SCOPE WAS REPROCESSED USING AN ENDOSCOPE REPROCESSOR, WHERE THE INTERNAL AIR AND WATER FILTERS HAD NOT BEEN REPLACED SINCE THE DATE OF INSTALL. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18846774 ↗
  • MalfunctionMar 6, 2024Device Reprocessing Problem; Device Reprocessing Problem

    THE CUSTOMER REPORTED THE ENDOSCOPE WAS INCORRECTLY REPROCESSED. THE SCOPE WAS REPROCESSED USING AN ENDOSCOPE REPROCESSOR, WHERE THE INTERNAL AIR AND WATER FILTERS HAD NOT BEEN REPLACED SINCE THE DATE OF INSTALL. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18846126 ↗
  • MalfunctionDec 22, 2023Failure to Clean Adequately; Failure to Clean Adequately

    THE CUSTOMER REPORTED TO OLYMPUS, THE EVIS EXERA II COLONOVIDEOSCOPE HAD AN ANGULATION WIRE AND UP/DOWN KNOB ISSUE. THE ISSUE WAS FOUND DURING REPROCESSING. THE DEVICE WAS RETURNED FOR EVALUATION. DURING THE DEVICE EVALUATION, IT WAS FOUND THAT THE BENDING SECTION COVER HAD FOREIGN OBJECTS. THERE WERE NO REPORTS OF PATIENT HARM. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18382822 ↗
  • MalfunctionNov 24, 2023Failure to Clean Adequately; Failure to Clean Adequately

    THE OLYMPUS FIELD SERVICE ENGINEER (ON BEHALF OF THE CUSTOMER) REPORTED TO OLYMPUS THAT THE COLONOVIDEOSCOPE HAD ANGULATION PLAY AND NOT ENOUGH AND INSERTION TUBE DISCOLORED/STAIN/WRINKLED. CUSTOMER FOUND THE PROBLEM IN THEIR REPROCESSING TIME. NEEDED DETAILED INSPECTION OF THE SUBJECT DEVICE. THERE WAS NO REPORT OF PATIENT HARM ASSOCIATED WITH THE EVENT. THE DEVICE WAS RETURNED FOR EVALUATION. DURING THE DEVICE EVALUATION, FOUND THE SCOPE HAD FOREIGN OBJECTS IN THE FOLLOWING PARTS: THE PLASTIC DISTAL END COVER, BENDING SECTION COVER AND IN THE CONNECTING TUBE. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18206370 ↗

Other Gastroscope and accessories, flexible/rigid devices

All Gastroscope and accessories, flexible/rigid devices →