COLONOVIDEOSCOPE CF-Q150I (UDI 04953170229473)Recalls & reported problems
Olympus Medical Systems Corp. · Class 2
All models of COLONOVIDEOSCOPE (12)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04953170229473), received between Dec 22, 2019 and Sep 25, 2026 — classified as 1 injury, 5 malfunctions.
The most frequently reported specific device problem is “Failure to Clean Adequately” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Clean Adequately2
- Adverse Event Without Identified Device or Use Problem1
- Break1
- Device Contamination with Chemical or Other Material1
- Device Reprocessing Problem1
- No Display/Image1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- No Consequences Or Impact To Patient1
- Perforation1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 25, 2026No Display/Image
The customer reported that no image was displayed with the colonovideoscope during reprocessing. During the initial evaluation of the device, image blackout was observed. There were no reports of patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50898172 ↗MalfunctionDec 12, 2025Failure to Clean Adequately; Failure to Clean Adequately
IT WAS REPORTED THAT THE COLOVIDEOSCOPE HAD DIRTY OBJECTIVE LENS AND LIGHT GUIDE LENS. THE ISSUE HAD OCCURRED DURING SERVICE/MAINTENANCE. THERE WAS NO REPORT OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23787386 ↗MalfunctionDec 8, 2025Device Reprocessing Problem; Device Reprocessing Problem
NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23752535 ↗MalfunctionJul 17, 2025Failure to Clean Adequately; Failure to Clean Adequately
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED ELECTRONICAL CONNECTOR PART WAS CRYSTALIZED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22537958 ↗MalfunctionJul 18, 2020Break; Device Contamination with Chemical or Other Material; Break; Device Contamination with Chemical or Other Material
OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) WAS INFORMED THAT DURING THE INCOMING INSPECTION AT THE SERVICE DEPARTMENT OF OLYMPUS (B)(6), IT WAS FOUND THAT THERE WAS FOREIGN MATERIAL ON THE CHANNEL OF THE SUBJECT DEVICE DUE TO THE DAMAGED CHANNEL. THERE WAS NO REPORT OF PATIENT INJURY ASSOCIATED WITH THIS EVENT. THE USER FACILITY DID NOT PROVIDE OTHER DETAILED INFORMATION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10293157 ↗InjuryDec 22, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) WAS INFORMED THAT THE USER FACILITY NOTICED THE INTESTINAL PERFORATION OF THE PATIENT DURING THE COLONOSCOPY. THE USER FACILITY OPERATED THE LAPAROSCOPIC PROCEDURE AND TREATED THE INTESTINAL PERFORATION FOR THE PATIENT. AFTER THE TREATMENT, THE PATIENT RECOVERED WELL AND LEFT THE USER FACILITY WITHOUT ANY ADVERSE REACTIONS.
Patient problems: Perforation
MAUDE report 9508730 ↗
Other Colonoscope and accessories, flexible/rigid devices
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS CF-HQ190LOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls2 deaths6,804 reports
- EVIS EXERA III XENON LIGHT SOURCE OLYMPUS CLV-190Olympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid6,617 reports
- EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-HQ190Olympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls5 deaths6,244 reports
- EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-H190Olympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid2 recalls7 deaths5,572 reports
- EVIS EXERA III COLONOVIDEOSCOPE OLYMPUS PCF-H190DLOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls2,830 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS PCF-H190LOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls2 deaths2,513 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS CF-H190IOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid1 recall1,967 reports
- PENTAX Medical EC38-i10cLHoya Corporation · Colonoscope and accessories, flexible/rigid1,618 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.