EVIS EXERA II DUODENOVIDEOSCOPE TJF-Q180V (UDI 04953170229503)Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of EVIS EXERA II DUODENOVIDEOSCOPE (3)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Olympus EVIS EXERA II duodenoscope appears in 638 adverse event reports, including 37 injuries and 3 deaths. Many reports describe scopes that did not clean adequately, tested positive for bacteria, or had cracks. Reports rose to 231 in the last 12 months from 118 in the 12 months before. There is 1 open recall, started in September 2023, over scopes that went back to customers with water left in their channels.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
653
Deaths reported
3
Injuries reported
37
Recalls naming it
1

Summary

The FDA's MAUDE database contains 653 adverse event reports that list this exact device (DI 04953170229503), received between May 24, 2018 and Sep 30, 2026 — classified as 3 deaths, 37 injuries, 613 malfunctions.

The most frequently reported specific device problem is “Failure to Clean Adequately” (166 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 8 reportsJun 2024: 12 reportsJul 2024: 7 reportsAug 2024: 13 reportsSep 2024: 10 reportsOct 2024: 14 reportsNov 2024: 3 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 5 reportsMar 2025: 13 reportsApr 2025: 9 reportsMay 2025: 11 reportsJun 2025: 22 reportsJul 2025: 23 reportsAug 2025: 21 reportsSep 2025: 6 reportsOct 2025: 5 reportsNov 2025: 17 reportsDec 2025: 11 reportsJan 2026: 34 reportsFeb 2026: 30 reportsMar 2026: 18 reportsApr 2026: 40 reportsMay 2026: 28 reportsJun 2026: 20 reportsJul 2026: 17 reportsAug 2026: 15 reportsSep 2026: 16 reportsOct 2026: 0 reports20242025202640
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 25, 2023Class IIOpen, Classified

    Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

    Z-0346-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Failure to Clean Adequately166
  • Microbial Contamination of Device137
  • Device Reprocessing Problem117
  • Crack79
  • Degraded36
  • Break34
  • Corroded32
  • Material Discolored32

Patient problems

  • No Clinical Signs, Symptoms or Conditions449
  • No Known Impact Or Consequence To Patient91
  • No Patient Involvement36
  • No Consequences Or Impact To Patient34
  • Unspecified Infection21
  • Bacterial Infection8
  • Blood Loss3
  • Drug Resistant Bacterial Infection3

Recent adverse event reports

The 25 most recent of 653 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Poor Quality Image

    The customer returned the Olympus Duodenovideoscope to the service center for a separate issue. During the initial evaluation of the device, striping and flacking of the image was observed. No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949257 ↗
  • MalfunctionSep 28, 2026Corroded; Residue After Decontamination

    The customer returned the Duodenovideoscope to the service center for a separate issue. During the initial evaluation of the returned device a corroded cylinder and Forceps elevator has foreign material was found. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50919665 ↗
  • MalfunctionSep 28, 2026Device Reprocessing Problem

    The customer returned the Duodenovideoscope to the service center for a separate issue. During the device analysis, the service repair technician indicates forceps elevator had foreign material. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50917362 ↗
  • MalfunctionSep 24, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50889865 ↗
  • MalfunctionSep 22, 2026Crack; Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50862212 ↗
  • MalfunctionSep 22, 2026Failure to Clean Adequately

    The customer returned the Duodenovideoscope to the service center for a separate issue. During device analysis, the service center identified foreign material on the forceps elevator. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50858807 ↗
  • MalfunctionSep 22, 2026Device Reprocessing Problem; Crack

    During onsite inspection of duodenovideoscope, it was found chip in adhesive around light guide lens and foreign material in the forceps elevator. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50858322 ↗
  • MalfunctionSep 15, 2026Device Reprocessing Problem; Corroded

    The customer returned the duodenovideoscope to the service center for a separate issue. During the device analysis, foreign objects were found at the light guide lens and distal end had corrosion. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50776860 ↗
  • MalfunctionSep 10, 2026Material Split, Cut or Torn; Failure to Clean Adequately

    The customer returned the duodenovideoscope to the service center for a separate issue. During the device analysis, the service team indicates that a Forceps elevator had foreign material and chipping at the adhesives around light guide lens was found. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50716202 ↗
  • MalfunctionSep 9, 2026Material Discolored; Detachment of Device or Device Component; Failure to Clean Adequately

    The customer reported that the Duodenovideoscope had a separate issue. During the initial evaluation of the device, discoloration inside the light guide lens, a chip in the adhesive around the light guide lens, and foreign material at the forceps elevator were observed. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50681854 ↗
  • MalfunctionSep 8, 2026Crack

    The customer reported that the duodenovideoscope had a separate issue. During the initial evaluation of the device, it was observed that there was chipping of the adhesives around the LG lens and the adhesives around the objective lens were chipped. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50664764 ↗
  • MalfunctionSep 2, 2026Degraded

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50568815 ↗
  • MalfunctionSep 2, 2026Device Reprocessing Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50566083 ↗
  • MalfunctionSep 1, 2026Material Discolored; Residue After Decontamination

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50553264 ↗
  • MalfunctionSep 1, 2026Corroded

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50551547 ↗
  • MalfunctionSep 1, 2026Residue After Decontamination

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50548011 ↗
  • MalfunctionAug 31, 2026Peeled/Delaminated; Failure to Clean Adequately

    The customer returned the duodenovideoscope to the service center for a separate issue. During the device analysis, the investigation observed that the forceps elevator had foreign material and the adhesive around the light guide lens had a chip. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50530606 ↗
  • MalfunctionAug 28, 2026Detachment of Device or Device Component; Failure to Clean Adequately

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50507672 ↗
  • MalfunctionAug 28, 2026Residue After Decontamination

    The customer returned the Duodenovideoscope to the service center for evaluation related to a separate issue. During testing the Repair Center identified the device had foreign material in forceps elevator. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50502278 ↗
  • MalfunctionAug 25, 2026Degraded

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50437701 ↗
  • MalfunctionAug 18, 2026Device Reprocessing Problem; Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50311622 ↗
  • MalfunctionAug 17, 2026Device Reprocessing Problem; Crack; Degraded

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50283533 ↗
  • MalfunctionAug 13, 2026Corroded; Failure to Clean Adequately

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50241571 ↗
  • MalfunctionAug 11, 2026Device Reprocessing Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50211953 ↗
  • MalfunctionAug 10, 2026Device Reprocessing Problem; Degraded

    The customer returned the duodeno videoscope to the service center for a separate issue. During the device analysis, the investigation indicates that worn adhesive around objective lens, foreign objects inside of light guide lens and foreign material in forceps elevator was found. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50196657 ↗

Other Endoscope, accessories, narrow band spectrum devices

All Endoscope, accessories, narrow band spectrum devices →