EVIS EXERA III COLONOVIDEOSCOPE OLYMPUS PCF-HQ190LRecalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of EVIS EXERA III COLONOVIDEOSCOPE (6)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Olympus EVIS EXERA III colonoscope is a flexible scope used for colon exams. Its 2,683 reports mostly describe scopes not cleaning adequately after reprocessing, residue left after cleaning, no display, and poor image quality. The reports include no deaths and 22 injuries, and most list no clinical signs for the patient. Reports in the last 12 months (2,091) are far above the previous 12 months (206). There are 2 open recalls, covering scopes repaired without adhesive and scopes returned with wet channels.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
2,828
Deaths reported
0
Injuries reported
24
Recalls naming it
2

Summary

The FDA's MAUDE database contains 2,828 adverse event reports that list this exact device (DI 04953170416118), received between Aug 12, 2021 and Sep 30, 2026 — classified as 24 injuries, 2,804 malfunctions.

The most frequently reported specific device problem is “Failure to Clean Adequately” (1,257 reports).

2 FDA recall records name this device's identifier, 2 still open.

1 other recall cover devices cleared under the same FDA submission (K192793); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 16 reportsDec 2023: 18 reportsJan 2024: 11 reportsFeb 2024: 17 reportsMar 2024: 23 reportsApr 2024: 62 reportsMay 2024: 8 reportsJun 2024: 21 reportsJul 2024: 21 reportsAug 2024: 23 reportsSep 2024: 26 reportsOct 2024: 34 reportsNov 2024: 9 reportsDec 2024: 7 reportsJan 2025: 14 reportsFeb 2025: 10 reportsMar 2025: 10 reportsApr 2025: 8 reportsMay 2025: 22 reportsJun 2025: 10 reportsJul 2025: 21 reportsAug 2025: 29 reportsSep 2025: 23 reportsOct 2025: 43 reportsNov 2025: 101 reportsDec 2025: 433 reportsJan 2026: 369 reportsFeb 2026: 122 reportsMar 2026: 147 reportsApr 2026: 190 reportsMay 2026: 203 reportsJun 2026: 219 reportsJul 2026: 173 reportsAug 2026: 117 reportsSep 2026: 162 reportsOct 2026: 0 reports202420252026433
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jan 11, 2024Class IIOpen, Classified

    Specific repaired colonovideoscopes were assembled without an adhesive which protects them from certain physical conditions such as high levels of vibration, fluctuation in temperature or severe shock loads.

    Z-1213-2024 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

  2. Sep 25, 2023Class IIOpen, Classified

    Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

    Z-0341-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Failure to Clean Adequately1,257
  • Communication or Transmission Problem474
  • Residue After Decontamination311
  • No Display/Image278
  • Poor Quality Image245
  • Device Reprocessing Problem130
  • Microbial Contamination of Device114
  • Erratic or Intermittent Display74

Patient problems

  • No Clinical Signs, Symptoms or Conditions2,800
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available11
  • Insufficient Information9
  • Hemorrhage/Blood Loss/Bleeding4
  • Hepatitis3
  • Bowel Perforation2
  • Foreign Body In Patient2
  • Abdominal Pain1

Recent adverse event reports

The 25 most recent of 2,828 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Residue After Decontamination

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50954958 ↗
  • MalfunctionSep 30, 2026Failure to Clean Adequately

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50954708 ↗
  • MalfunctionSep 30, 2026Residue After Decontamination

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949295 ↗
  • MalfunctionSep 30, 2026Erratic or Intermittent Display

    The customer reported that the camera image was intermittently cutting in and out during a diagnostic procedure involving an Olympus Colonovideoscope. The procedure was completed using the same device. There was no patient injury reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50946817 ↗
  • MalfunctionSep 30, 2026Failure to Clean Adequately

    The customer returned the Colonovideoscope for a separate issue. During the initial evaluation of the returned device, white residue on the channel and cylinder was observed. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50946810 ↗
  • MalfunctionSep 30, 2026Failure to Clean Adequately

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50945015 ↗
  • MalfunctionSep 30, 2026Communication or Transmission Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50941468 ↗
  • MalfunctionSep 29, 2026Failure to Clean Adequately

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50940880 ↗
  • MalfunctionSep 29, 2026Erratic or Intermittent Display; No Display/Image; Communication or Transmission Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50937737 ↗
  • MalfunctionSep 29, 2026Residue After Decontamination

    The customer reported the colonovideoscope for a separate issue. During the initial evaluation of the device, the white residue in air water channel on the light guide channel side was observed. There was no patient involvement reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50935932 ↗
  • MalfunctionSep 29, 2026Contamination; Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50934409 ↗
  • MalfunctionSep 29, 2026Residue After Decontamination

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50932745 ↗
  • MalfunctionSep 29, 2026No Display/Image; Communication or Transmission Problem

    The customer returned the Colonovideoscope to the service center for a separate issue. During the device analysis, the Investigator indicated that no image and error B30 was found. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50928234 ↗
  • MalfunctionSep 28, 2026Communication or Transmission Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50925262 ↗
  • MalfunctionSep 28, 2026Residue After Decontamination

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50924601 ↗
  • MalfunctionSep 28, 2026Residue After Decontamination

    The customer returned the Colonovideoscope to the service center for a separate issue. During the initial evaluation of the device, the inside of the auxiliary water flow had white residue. There were no reports of patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50917964 ↗
  • MalfunctionSep 28, 2026Failure to Clean Adequately

    The customer returned the colonovideoscope for a separate issue. During the initial evaluation of the returned device, foreign material was observed in the air/water cylinder and channel. No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50917828 ↗
  • MalfunctionSep 28, 2026Residue After Decontamination

    The customer reported the colonovideoscope for a separate issue. During the initial evaluation of the device, the white residue in air water channel and cylinder was observed. There was no patient involvement reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50916832 ↗
  • MalfunctionSep 28, 2026Residue After Decontamination

    The customer reported the Olympus colonovideoscope had a separate issue. During the initial evaluation of the device it was observed, White residue on air water channel and cylinder and auxiliary channel were observed. No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50915990 ↗
  • MalfunctionSep 28, 2026Failure to Clean Adequately

    The customer returned the colonovideoscope for a separate issue. During the initial evaluation of the returned device, white residue was observed on both sides of the channel. No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50915043 ↗
  • MalfunctionSep 28, 2026Failure to Clean Adequately

    The customer returned the colonovideoscope for a separate issue. During the initial evaluation of the returned device, white residue was observed within the air/water supply channel. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50914970 ↗
  • MalfunctionSep 28, 2026Overheating of Device

    The customer reported the Colonovideoscope had a separate issue. During the initial evaluation of the device it was observed that, the C-cover and C-cover insulation had a burn. No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50914799 ↗
  • MalfunctionSep 28, 2026No Display/Image

    The customer reported that the image was blanking out during inspection for use involving an Olympus Colonovideoscope. There was no patient involvement reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50914764 ↗
  • MalfunctionSep 28, 2026Failure to Clean Adequately

    The customer returned the colonovideoscope for a separate issue. During the initial evaluation of the returned device, white residue was observed within the air/water channel and the air/water cylinder. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50914709 ↗
  • MalfunctionSep 28, 2026Failure to Clean Adequately

    The customer returned the colonovideoscope for a separate issue. During the initial evaluation of the returned device, white residue was observed in the channel and cylinder side. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50912529 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K192793. One submission often covers many models, so check the product and lot details in each record.

  • Z-1212-2024 · Jan 11, 2024 · Open, Classified
    Colonoscope, Model Number PCF-H190TL.

Other Endoscope, accessories, narrow band spectrum devices

All Endoscope, accessories, narrow band spectrum devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page