This Olympus EVIS EXERA II duodenoscope is a video endoscope used to examine the upper digestive tract. Its 2,335 reports often describe the scope not being cleaned adequately between uses, along with cracks and contamination, and mention 86 injuries and 6 deaths. Some mention infections. Reports rose from 221 in the prior 12 months to 504 in the latest 12 months. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
2,362
Deaths reported
6
Injuries reported
86
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2,362 adverse event reports that list this exact device (DI 04953170367311), received between May 11, 2018 and Sep 30, 2026 — classified as 6 deaths, 86 injuries, 2,270 malfunctions.

The most frequently reported specific device problem is “Failure to Clean Adequately” (992 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 42 reportsDec 2023: 51 reportsJan 2024: 54 reportsFeb 2024: 59 reportsMar 2024: 47 reportsApr 2024: 36 reportsMay 2024: 36 reportsJun 2024: 40 reportsJul 2024: 25 reportsAug 2024: 22 reportsSep 2024: 15 reportsOct 2024: 47 reportsNov 2024: 19 reportsDec 2024: 2 reportsJan 2025: 6 reportsFeb 2025: 5 reportsMar 2025: 23 reportsApr 2025: 11 reportsMay 2025: 27 reportsJun 2025: 26 reportsJul 2025: 31 reportsAug 2025: 30 reportsSep 2025: 22 reportsOct 2025: 19 reportsNov 2025: 34 reportsDec 2025: 22 reportsJan 2026: 62 reportsFeb 2026: 51 reportsMar 2026: 41 reportsApr 2026: 76 reportsMay 2026: 72 reportsJun 2026: 35 reportsJul 2026: 42 reportsAug 2026: 59 reportsSep 2026: 37 reportsOct 2026: 0 reports20242025202676
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Clean Adequately992
  • Device Reprocessing Problem291
  • Crack242
  • Contamination216
  • Microbial Contamination of Device198
  • Device Contamination with Chemical or Other Material134
  • Degraded116
  • Break93

Patient problems

  • No Clinical Signs, Symptoms or Conditions2,111
  • No Consequences Or Impact To Patient97
  • No Known Impact Or Consequence To Patient42
  • Bacterial Infection35
  • Unspecified Infection34
  • No Patient Involvement31
  • Hemorrhage/Blood Loss/Bleeding7
  • Insufficient Information7

Recent adverse event reports

The 25 most recent of 2,362 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Device Reprocessing Problem; Crack

    The customer returned the duodenovideoscope for a separate issue. During the initial evaluation of the returned device it was observed foreign objects inside of light guide lens, chipping of adhesive around objective lens and chipping of adhesive around light guide lens. No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50951172 ↗
  • MalfunctionSep 30, 2026Failure to Clean Adequately

    No additional information provided by the customer.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50950773 ↗
  • MalfunctionSep 30, 2026Material Discolored

    The customer returned the Duodenovideoscope for a separate issue. During the initial evaluation of the device, a stain was observed inside the light guide lens. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50943405 ↗
  • MalfunctionSep 29, 2026Failure to Clean Adequately

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50935636 ↗
  • MalfunctionSep 29, 2026Failure to Clean Adequately

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50934554 ↗
  • MalfunctionSep 28, 2026Degraded

    The customer reported the Duodenovideoscope for a separate issue. During the initial evaluation of the device, adhesive wear around the lenses was observed. No patient involvement was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50918790 ↗
  • MalfunctionSep 25, 2026Failure to Clean Adequately

    The customer returned the Duodenovideoscope for a separate issue. During the evaluation of the device, Forceps elevator foreign material was observed. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50895711 ↗
  • MalfunctionSep 24, 2026Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50892402 ↗
  • MalfunctionSep 24, 2026Material Discolored; Failure to Clean Adequately

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50892002 ↗
  • MalfunctionSep 24, 2026Failure to Clean Adequately

    The customer returned the duodenovideoscope for a separate issue. During the initial evaluation of the returned device, the control knob was found to require cleaning, and foreign material was observed on the forceps elevator. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50889748 ↗
  • MalfunctionSep 23, 2026Residue After Decontamination

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50876469 ↗
  • MalfunctionSep 23, 2026Failure to Clean Adequately

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50872749 ↗
  • MalfunctionSep 23, 2026Corroded; Material Discolored; Detachment of Device or Device Component

    The customer reported that the Duodenovideoscope experienced distal tip loss during reprocessing. In addition, during the initial evaluation of the returned device, chipping of the adhesive around the Light Guide (LG) lens and Objective lens, as well as discoloration inside the Light Guide (LG) lens was observed. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50868785 ↗
  • MalfunctionSep 22, 2026Failure to Clean Adequately

    The customer returned the Duodenovideoscope, which is an Olympus asset, with no reported complaint. During the evaluation of the device, foreign material was observed on the forceps elevator. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50856523 ↗
  • MalfunctionSep 18, 2026Naturally Worn

    The customer returned the duodenovideoscope for a separate issue. During the initial evaluation of the returned device it was observed worn adhesive at the lens. No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50827465 ↗
  • MalfunctionSep 17, 2026Failure to Clean Adequately

    The customer returned the duodenovideoscope to the service center for a separate issue. During the device analysis, the service team indicates that a forceps elevator had foreign material or a stain. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50808833 ↗
  • MalfunctionSep 16, 2026Failure to Clean Adequately

    The customer returned the Duodenovideoscope to the service center for a separate issue. During the device analysis the service repair technician, indicates foreign material in the Forceps elevator was found. There was no patient involvement

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50794947 ↗
  • MalfunctionSep 16, 2026Failure to Clean Adequately

    During onsite inspection of duodenovideoscope, it was found that the forceps elevator had foreign material. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50794350 ↗
  • MalfunctionSep 15, 2026Failure to Clean Adequately

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50780850 ↗
  • MalfunctionSep 15, 2026Corroded

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50777187 ↗
  • MalfunctionSep 11, 2026Microbial Contamination of Device

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50744393 ↗
  • MalfunctionSep 11, 2026Material Split, Cut or Torn; Failure to Clean Adequately

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50742348 ↗
  • MalfunctionSep 11, 2026Failure to Clean Adequately

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50740090 ↗
  • MalfunctionSep 11, 2026Degraded

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50739229 ↗
  • MalfunctionSep 11, 2026Material Discolored; Failure to Clean Adequately

    The customer reported that the duodenovideoscope had a separate issue. During the initial evaluation of the returned device, a stain inside the light guide (LG) lens and foreign material in the forceps elevator were observed. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50725584 ↗

Other Endoscope, accessories, narrow band spectrum devices

All Endoscope, accessories, narrow band spectrum devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page