EVIS EXERA II COLONOVIDEOSCOPE OLYMPUS PCF TYPE H180AI (UDI 04953170237300)Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of EVIS EXERA II COLONOVIDEOSCOPE (28)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Olympus colonoscope is a flexible colonoscope. Reports mostly describe cleaning and reprocessing problems, microbial contamination, corrosion, and image display errors, with no deaths or injuries among 41 malfunction reports. There are no recalls naming this device. Reports in the last 12 months rose to 14 from 6.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
42
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 42 adverse event reports that list this exact device (DI 04953170237300), received between Sep 20, 2018 and Sep 16, 2026 — classified as 42 malfunctions.

The most frequently reported specific device problem is “Failure to Clean Adequately” (21 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 2 reportsJul 2024: 3 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 2 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 1 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 1 reportsMar 2026: 2 reportsApr 2026: 5 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 3 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Clean Adequately21
  • Device Reprocessing Problem10
  • Microbial Contamination of Device7
  • Corroded2
  • Image Display Error/Artifact2
  • Poor Quality Image2
  • Contamination1
  • Device Contamination with Chemical or Other Material1

Patient problems

  • No Clinical Signs, Symptoms or Conditions37
  • No Known Impact Or Consequence To Patient5

Recent adverse event reports

The 25 most recent of 42 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 16, 2026Poor Quality Image

    The customer reported the image had distortions during inspection for use involving an Olympus colonovideoscope. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50791665 ↗
  • MalfunctionAug 31, 2026Device Reprocessing Problem; Image Display Error/Artifact

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50529812 ↗
  • MalfunctionAug 17, 2026Corroded

    The customer returned the colonovideoscope for a separate issue. During the initial evaluation of the returned device, corrosion was observed on the light guide lens. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50276191 ↗
  • MalfunctionAug 6, 2026Device Reprocessing Problem

    The customer returned the Colonovideoscope to the service center for a separate issue. During the device analysis the service repair technician, indicates that there's foreign material in the air water cylinder and in the air water tube was found. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50150528 ↗
  • MalfunctionMay 25, 2026Image Display Error/Artifact; Poor Quality Image

    THE CUSTOMER REPORTED PIXEL ARTIFACTS THAT CONCENTRATED IN THE CENTRAL IMAGE AREA, AND INTERFERENCE OCCURRED DURING INSPECTION FOR USE INVOLVING AN OLYMPUS COLONOVIDEOSCOPE. THE CUSTOMER SUSPECTED THAT THIS ISSUE WAS DUE TO A DEFECT IN THE CHARGED COUPLED DEVICE (CCD) UNIT THERE WAS NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25273383 ↗
  • MalfunctionApr 27, 2026Corroded

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE HAD CORRODED DISTAL END. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24995347 ↗
  • MalfunctionApr 17, 2026Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE HAD FOREIGN OBJECTS IN THE AIR/WATER CYLINDER, TUBE AND JET TUBE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24907810 ↗
  • MalfunctionApr 13, 2026Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED FOREIGN OBJECTS IN AIR/WATER CYLINDER AND AIR/WATER TUBE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24860859 ↗
  • MalfunctionApr 8, 2026Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE HAD FOREIGN OBJECTS IN AIR/WATER CHANNEL AND AIR/WATER CYLINDER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24816228 ↗
  • MalfunctionApr 2, 2026Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED FOREIGN OBJECTS IN AIR/WATER CYLINDER AND AIR/WATER TUBE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24763376 ↗
  • MalfunctionMar 30, 2026Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE HAD FOREIGN OBJECTS IN AIR/WATER TUBE AND AIR/WATER CYLINDER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24733488 ↗
  • MalfunctionMar 16, 2026Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED WHITE FOREIGN MATERIAL IN THE AIR/WATER CYLINDER AND AIR/WATER TUBE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24611884 ↗
  • MalfunctionFeb 17, 2026Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE WAS FOUND TO CONTAIN A FOREIGN OBJECT IN THE AIR-WATER CHANNEL AND IN THE AIR-WATER CYLINDER. THERE WAS NO PATIENTS WERE INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24368851 ↗
  • MalfunctionJan 27, 2026Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED FOREIGN OBJECTS IN C-COVER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24190430 ↗
  • MalfunctionNov 14, 2025No Display/Image; No Display/Image

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23555951 ↗
  • MalfunctionSep 2, 2025Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE'S LIGHT GUIDE LENS HAD FOREIGN OBJECTS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22948569 ↗
  • MalfunctionAug 26, 2025Device Reprocessing Problem; Device Reprocessing Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED FOREIGN MATTER IN THE AUXILIARY WATER CHANNEL/TUBE (J-TUBE), AND THE AIR/WATER TUBE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22898983 ↗
  • MalfunctionJul 17, 2025Device Reprocessing Problem; Device Reprocessing Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED FOREIGN MATERIAL IN THE SCOPE AIR/WATER CYLINDER (TUBE)/TUBE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22530275 ↗
  • MalfunctionJun 23, 2025Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE AIR/WATER CYLINDER AND THE AIR/WATER TUBE OF THE COLONOVIDEOSCOPE HAD FOREIGN OBJECTS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22309865 ↗
  • MalfunctionApr 30, 2025Device Reprocessing Problem; Device Reprocessing Problem

    DURING DEVICE EVALUATION, A FOREIGN OBJECT WAS FOUND IN THE COLONOVIDEOSCOPE'S JET TUBE. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21936025 ↗
  • MalfunctionApr 3, 2025Device Reprocessing Problem; Device Reprocessing Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED AIR/WATER TUBE AND CYLINDER HAD FOREIGN OBJECTS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21769952 ↗
  • MalfunctionJul 24, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE COLONO VIDEO SCOPE EXHIBITED FOREIGN OBJECTS IN AIR/WATER TUBE AND CYLINDER. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19822667 ↗
  • MalfunctionJul 5, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE FLEX VIDEO SCOPE EXHIBITED AIR/WATER CYLINDER, AIR/WATER TUBE, AND THE JET TUBE HAD FOREIGN OBJECTS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19682160 ↗
  • MalfunctionJul 1, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION THAT THE FLEX VIDEO SCOPE HAD WHITISH FOREIGN MATERIAL WAS FOUND INSIDE THE AIR/WATER TUBE, AIR/WATER CYLINDER AND JET TUBE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19650562 ↗
  • MalfunctionJun 18, 2024Device Reprocessing Problem; Failure to Clean Adequately; Device Reprocessing Problem; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE COLONOVIDEOSCOPE EXHIBITED FOREIGN MATERIAL BLOCKING THE NOZZLE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19557963 ↗

Other Colonoscope and accessories, flexible/rigid devices

All Colonoscope and accessories, flexible/rigid devices →