Skip to content

PENTAX EG-3270UKRecalls & reported problems

Hoya Corporation · EG-3270UK ULTRASOUND VIDEOSCOPE · Class 2

All models of PENTAX (72)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 04961333233793), received between Aug 6, 2021 and Aug 5, 2026 — classified as 10 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak5
  • Optical Obstruction2
  • Material Puncture/Hole1
  • Material Split, Cut or Torn1
  • Obstruction of Flow1
  • Unintended Movement1

Patient problems

  • No Clinical Signs, Symptoms or Conditions10
  • Insufficient Information1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 5, 2026Fluid/Blood Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50137434 ↗
  • MalfunctionJun 18, 2026Optical Obstruction

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE (CLOUDY).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25542254 ↗
  • MalfunctionMay 12, 2026Fluid/Blood Leak

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(FLUID DAMAGE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25146011 ↗
  • MalfunctionJun 25, 2025Optical Obstruction; Optical Obstruction

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE (CLOUDY).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22325453 ↗
  • MalfunctionOct 31, 2024Fluid/Blood Leak; Fluid/Blood Leak

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE (FLUID DAMAGE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20575833 ↗
  • MalfunctionSep 24, 2024Fluid/Blood Leak; Fluid/Blood Leak

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(FLUID DAMAGE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20289348 ↗
  • MalfunctionMar 8, 2022Fluid/Blood Leak; Fluid/Blood Leak

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(FLUID DAMAGE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13691881 ↗
  • MalfunctionDec 8, 2021Unintended Movement; Unintended Movement

    THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12958451 ↗
  • MalfunctionAug 26, 2021Obstruction of Flow; Obstruction of Flow

    PENTAX MEDICAL WAS MADE AWARE OF AN EVENT THAT OCCURRED IN THE UNITED STATES. THE INFORMATION PROVIDED INDICATED THAT THE SUCTION CHANNEL WAS BLOCKED INVOLVING A PENTAX MEDICAL ULTRASOUND VIDEO GASTROSCOPE MODEL EG-3270UK, SERIAL NUMBER (B)(4). THE CUSTOMER STATED THERE WAS AN INTERNAL LEAK INSIDE THE CHANNEL. GOOD FAITH EFFORTS(GFE) EMAILS HAVE BEEN SENT. NO ADDITIONAL INFORMATION HAS BEEN PROVIDED. THE CUSTOMER OWNED ENDOSCOPE WAS RECEIVED BY PENTAX MEDICAL FOR EVALUATION ON 02-AUG-2021. THE ENDOSCOPE WAS INSPECTED BY PENTAX MEDICAL SERVICE UNDER SERVICE ORDER (B)(4) AND DURING QUALITY INSPECTION THE TECHNICIAN CONFIRMED THE CUSTOMERS EXPERIENCE AS THE "OPERATION CHANNEL- PRIMARY SLICE BY ACCESSORY" WAS IDENTIFIED AND ALSO DOCUMENTED THE FOLLOWING INSPECTION FINDINGS: LEAK AT BIOPSY CHANNEL (LARGE/ PRIMARY) DISTAL SIDE, FAILED WET LEAK TEST, FAILED DRY LEAK TEST, SUCTION TUBE MILD RESISTANCE, ULTRASOUND WARNING LABEL CUT, SUCTION FUNCTION NOT PERFORMED UNIT COMPROMISED, OPERATION CHANNEL- PRIMARY SLICE BY ACCESSORY, FLUID INVASION NOT OBSERVED IN PVE CONNECTOR, FLUID INVASION NOT OBSERVED IN CONTROL BODY. THE DEVICE UNDERWENT REPAIRS INCLUDING THE FOLLOWING COMPONENTS: O-RINGS AN

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 12373950 ↗
  • MalfunctionAug 6, 2021Material Puncture/Hole; Material Split, Cut or Torn; Material Puncture/Hole; Material Split, Cut or Torn

    THIS TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. LIGHT GUIDE PLUG CORRODED. THIS MODEL SHOULD BE COVERED BY (B)(4) ALSO.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12285954 ↗

Other Endoscopic Ultrasound System, Gastroenterology-Urology devices

All Endoscopic Ultrasound System, Gastroenterology-Urology devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page