Colonoscope: for endoscopy and endoscopic surgery within… recallZ-0329-2024 · Aizu Olympus Co., Ltd. · initiated Sep 25, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Aizu Olympus Co., Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
FDA root cause category: Under Investigation by firm
What the company is doing
Olympus America issued Urgent Medical Device Correction letter on Sept. 25, 2023 to Endoscopy, Respiratory, Pulmonary and Urology Departments, Risk Management Department. Letter states reason for recall, health risk and action to take: Our records indicate that Olympus has serviced your flexible endoscope at one of our Olympus repair facilities. Following repair activities, Olympus uses a water flush of endoscope channels as part of our final endoscope channel inspection, and we subsequently dry the endoscope channels with compressed filtered air. Olympus discovered that our channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Olympus is in the process of implementing a validated drying process at our repair centers. In the meantime, Olympus has established interim processes to perform an additional inspection of your scope for water prior to return. 1. Ensure your users are knowledgeable on the content of this letter. 2. Complete the enclosed response form and return it to our recall partner, Sedgwick, via e-mail at [email protected] or by fax at 844-294-7184. For any questions about the acknowledgement form, please call the Sedgwick team at 888-667-1504. 3. Olympus requests that you report complaints, including infections, and endoscopes returned wet from our repair centers to Olympus. Please report complaints to Technical Assistance Center (TAC) at 1-800-848-9024, option 1. Adverse events experienced with the use of this product may also be reported to the Food and Drug Administration MedWatch Adverse Event Reporting program either online, by regular mail or by fax. If you require additional information, please do not hesitate to contact me at (647) 999-3203 or [email protected].
Recall details
- Recall number
- Z-0329-2024
- Status
- Open, Classified
- Initiated by the company
- Sep 25, 2023
- Posted by the FDA
- Nov 17, 2023
- Recalling firm
- Aizu Olympus Co., Ltd.
- Quantity
- 3344 units
- Product code
- FET
- FDA submissions
- K100584, K954451, K051645, K111756, K112680, K001241
- Distribution
- Worldwide - US Nationwide distribution.
Product description
Colonoscope: for endoscopy and endoscopic surgery within the lower digestive tract (including the anus, rectum, sigmoid colon, colon and ileocecal valve). Model Numbers: CF-2T160I, CF-2T160L, CF-H170I, CF-H170L, CF-H180AI, CF-H180AL, CF-H180DI, CF-H180DL, CF-H190I, CF-H190L, CF-HQ190I, CF-HQ190L, CF-LV1I, CF-LV1L, CF-Q150I, CF-Q150L, CF-Q160S, CF-Q180AI, CF-Q180AL. 510(k) Numbers: K954451, K112680, K100584, K051645, K111756, K001241, K911278.
Code information (models, lots, serial numbers)
Model Number (UDI-DI): CF-2T160I (04953170307645), CF-2T160L (04953170307652), CF-H170I (04953170334184), CF-H170L (04953170334160), CF-H180AI (04953170339585), CF-H180AL (04953170339608), CF-H180DI (04953170339639), CF-H180DL (04953170339653), CF-H190I (04953170305177), CF-H190L (04953170305153), CF-HQ190I (04953170305139), CF-HQ190L (04953170305115), CF-LV1I (04953170312410), CF-LV1L (04953170312397), CF-Q150I (04953170307621), CF-Q150L (04953170307638), CF-Q160S (04953170307669), CF-Q180AI (04953170307690), CF-Q180AL (04953170307706).
Devices named in this recall
19 devices in the FDA device registry whose identifier appears in the recall's code information.
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS CF-HQ190LOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls2 deaths6,804 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS CF-H190IOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid1 recall1,967 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS CF-HQ190IOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid2 recalls1,315 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS CF-H190LOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid1 recall1 death1,194 reports
- COLONOVIDEOSCOPE CF-H170LOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid1 recall377 reports
- COLONOVIDEOSCOPE CF-H170IOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid1 recall272 reports
- COLONOVIDEOSCOPE CF-Q150LOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid1 recall162 reports
- EVIS EXERA II COLONOVIDEOSCOPE OLYMPUS CF TYPE H180ALOlympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid2 recalls72 reports
- EVIS EXERA II COLONOVIDEOSCOPE OLYMPUS CF TYPE H180AIOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid1 recall117 reports
- EVIS EXERA II COLONOVIDEOSCOPE OLYMPUS CF TYPE Q180ALOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid2 recalls5 reports
- EVIS EXERA II COLONOVIDEOSCOPE OLYMPUS CF TYPE Q180AIOlympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum1 recall16 reports
- COLONOVIDEOSCOPE OLYMPUS CF-LV1LOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid1 recall10 reports
- EVIS EXERA II COLONOVIDEOSCOPE OLYMPUS CF TYPE H180DLOlympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid1 recall9 reports
- COLONOVIDEOSCOPE CF-Q150IOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid1 recall6 reports
- COLONOVIDEOSCOPE OLYMPUS CF-LV1IOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid1 recall6 reports
- EVIS EXERA II COLONOVIDEOSCOPE OLYMPUS CF TYPE H180DIOlympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid1 recall4 reports
- EVIS EXERA COLONOVIDEOSCOPE OLYMPUS CF TYPE 2T160IOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology1 recall1 report
- EVIS EXERA SIGMOIDOVIDEOSCOPE OLYMPUS CF TYPE Q160SOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology1 recall1 report
- EVIS EXERA COLONOVIDEOSCOPE OLYMPUS CF TYPE 2T160LOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology1 recall0 reports
Product lines
- COLONOVIDEOSCOPEOlympus Medical Systems Corp. · 12 models · 2,094 reports · 1 recall
- EVIS EXERA II COLONOVIDEOSCOPEOlympus Medical Systems Corp. · 28 models · 2,142 reports · 8 recalls
- EVIS EXERA lll COLONOVIDEOSCOPEOlympus Medical Systems Corp. · 16 models · 15,013 reports · 7 recalls
- EVIS EXERA COLONOVIDEOSCOPEOlympus Medical Systems Corp. · 5 models · 5 reports · 1 recall
- EVIS EXERA SIGMOIDOVIDEOSCOPEOlympus Medical Systems Corp. · 3 models · 33 reports · 1 recall
Other products in the same recall event
The FDA groups related products into one recall event (event 93222).
- Z-0327-2024 · Open, ClassifiedBronchoscope-for endoscopy and endoscopic surgery within the airways and tracheobronchial tree. Model Numbers: BF-1T150, BF-1T60, BF-1TH190, BF-1TQ170, BF-3C160, BF-3C40, BF-H190…
- Z-0328-2024 · Open, ClassifiedBronchoscope - Ultrasonic: endoscopic real-time ultrasonic imaging, ultrasound guided needle aspiration and other endoscopic procedures within the airways, trancheobronchial tree, …
- Z-0330-2024 · Open, ClassifiedCholeodochoscope. Model Numbers: CHF-BP30, CHF-CB30L, CHF-P60.
- Z-0331-2024 · Open, ClassifiedCystonephroscope-endoscopic diagnosis and treatment within the bladder, urethra, and kidney Model Numbers: CYF-5, CYF-VH, CYF-VHR.
- Z-0332-2024 · Open, ClassifiedRhinolaryngoscope-for endoscopic diagnosis and treatment within the nasal lumens and airway anatomy (including nasopharyngeal and trachea). Model Numbers: ENF-T3, ENF-VT2, ENF-V…
- Z-0333-2024 · Open, ClassifiedGastroscope-endoscopic ultrasound (EUS) guided fine needle aspiration (FNA) and for endoscopic surgery within the upper digestive tract. Model Numbers: GF-UC140P-AL5, GF-UCT180, …
- Z-0334-2024 · Open, ClassifiedGastrointestinal videoscope-indicated for use within the upper digestive tract (including the esophagus, stomach, and duodenum). Model Numbers: GIF-1TH190, GIF-1TQ160, GIF-2T160, …
- Z-0335-2024 · Open, ClassifiedHysteroscope-For endoscopic diagnosis within the uterus Model Number: HYF-1T.
- Z-0336-2024 · Open, ClassifiedHysterovideoscope-For endoscopic diagnosis within the uterus Model Number: HYF-V.
- Z-0337-2024 · Open, ClassifiedHysterofiberscope-For endoscopic diagnosis within the uterus. Model Number: HYF-XP.
- Z-0338-2024 · Open, ClassifiedTracheal fiberscope-airway management, which includes observation to access airway anatomy, endotracheal/endobronchial intubation and management. Model Numbers: LF-DP, LF-GP, LF-…
- Z-0339-2024 · Open, ClassifiedTracheal videoscope- For airway management, which includes observation to access airway anatomy, endotracheal/endobronchial intubation and management. Model Number: LF-V.
- Z-0340-2024 · Open, ClassifiedMobile airway scope-diagnosis and observation to access airway anatomy, endotracheal/endobronchial intubation and management. Model Numbers: MAF-DM2, MAF-GM, MAF-GM2, MAF-TM, MAF-…
- Z-0341-2024 · Open, ClassifiedColonoscope-indicated for use within the lower digestive tract (including the anus, rectum, sigmoid colon, colon, and ileocecal valve). Model Numbers: PCF-H180AI, PCF-H180AL, PCF-…
- Z-0342-2024 · Open, ClassifiedSignmoidovideoscope-For endoscopy and endoscopic surgery within the lower digestive tract (including the anus, rectum and sigmoid colon). Model Number: PCF-S.
- Z-0343-2024 · Open, ClassifiedSmall intestinal videoscope-For endoscopy and endoscopic surgery within the upper digestive tract (including the esophagus, stomach, duodenum, and small intestine) by oral insertio…
- Z-0344-2024 · Open, ClassifiedTransnasal esophagovideoscope-For endoscopy and endoscopic surgery within the esophagus and stomach. Model Number: PEF-V.
- Z-0345-2024 · Open, ClassifiedUltrasonic gastrovideoscope-Real time ultrasound imaging,ultrasound guided needle aspiration and other endoscopic procedures within the upper gastrointestinal tract and surrounding…
- Z-0346-2024 · Open, ClassifiedDuodenovideoscope-EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery Model Numbers: TJF-145, TJF-150, TJF-Q180V…
- Z-0347-2024 · Open, ClassifiedUreteralscope- for endoscopic diagnosis and treatment within the ureter, kidney and biliary tract (common bile duct and hepatic duct). Model Numbers: URF-P6, URF-P6R.
- Z-0348-2024 · Open, ClassifiedUretero-reno fiberscope-For endoscopic diagnosis and treatment within the ureter and kidney. Model Numbers: URF-P7, URF-P7R, URF-V2, URF-V2R, URF-V3, URF-V3R.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page