Colonoscope-indicated for use within the lower digestive… recallZ-0341-2024 · Aizu Olympus Co., Ltd. · initiated Sep 25, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Aizu Olympus Co., Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
FDA root cause category: Under Investigation by firm
What the company is doing
Olympus America issued Urgent Medical Device Correction letter on Sept. 25, 2023 to Endoscopy, Respiratory, Pulmonary and Urology Departments, Risk Management Department. Letter states reason for recall, health risk and action to take: Our records indicate that Olympus has serviced your flexible endoscope at one of our Olympus repair facilities. Following repair activities, Olympus uses a water flush of endoscope channels as part of our final endoscope channel inspection, and we subsequently dry the endoscope channels with compressed filtered air. Olympus discovered that our channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Olympus is in the process of implementing a validated drying process at our repair centers. In the meantime, Olympus has established interim processes to perform an additional inspection of your scope for water prior to return. 1. Ensure your users are knowledgeable on the content of this letter. 2. Complete the enclosed response form and return it to our recall partner, Sedgwick, via e-mail at [email protected] or by fax at 844-294-7184. For any questions about the acknowledgement form, please call the Sedgwick team at 888-667-1504. 3. Olympus requests that you report complaints, including infections, and endoscopes returned wet from our repair centers to Olympus. Please report complaints to Technical Assistance Center (TAC) at 1-800-848-9024, option 1. Adverse events experienced with the use of this product may also be reported to the Food and Drug Administration MedWatch Adverse Event Reporting program either online, by regular mail or by fax. If you require additional information, please do not hesitate to contact me at (647) 999-3203 or [email protected].
Recall details
- Recall number
- Z-0341-2024
- Status
- Open, Classified
- Initiated by the company
- Sep 25, 2023
- Posted by the FDA
- Nov 17, 2023
- Recalling firm
- Aizu Olympus Co., Ltd.
- Quantity
- 3101 units
- Product code
- NWB
- FDA submissions
- K131780, K192793, K980465
- Distribution
- Worldwide - US Nationwide distribution.
Product description
Colonoscope-indicated for use within the lower digestive tract (including the anus, rectum, sigmoid colon, colon, and ileocecal valve). Model Numbers: PCF-H180AI, PCF-H180AL, PCF-H190DI, PCF-H190DL, PCF-H190I, PCF-H190L, PCF-H190TI, PCF-H190TL, PCF-HQ190I, PCF-HQ190L, PCF-PH190I, PCF-PH190L, PCF-Q180AI, PCF-Q180AL, PCF-S.
Code information (models, lots, serial numbers)
Model Number (UDI-DI): PCF-H180AI (04953170339875), PCF-H180AL (04953170339899), PCF-H190DI (04953170363696), PCF-H190DL (04953170363672), PCF-H190I (04953170305214), PCF-H190L (04953170305191), PCF-H190TI (04953170420719), PCF-H190TL (04953170420702), PCF-HQ190I (04953170416132), PCF-HQ190L (04953170416118), PCF-PH190I (04953170305252), PCF-PH190L (04953170305238), PCF-Q180AI (04953170307713), PCF-Q180AL (04953170307720), PCF-S (04953170339912).
Devices named in this recall
15 devices in the FDA device registry whose identifier appears in the recall's code information.
- EVIS EXERA III COLONOVIDEOSCOPE OLYMPUS PCF-H190DLOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls2,830 reports
- EVIS EXERA III COLONOVIDEOSCOPE OLYMPUS PCF-HQ190LOlympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum2 recalls2,828 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS PCF-H190LOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls2 deaths2,513 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS PCF-H190IOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid1 recall460 reports
- EVIS EXERA III COLONOVIDEOSCOPE OLYMPUS PCF-H190TLOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid2 recalls184 reports
- EVIS EXERA III COLONOVIDEOSCOPE OLYMPUS PCF-H190DIOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid1 recall164 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS PCF-PH190LOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid1 recall139 reports
- EVIS EXERA III COLONOVIDEOSCOPE OLYMPUS PCF-H190TIOlympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum1 recall77 reports
- EVIS EXERA II COLONOVIDEOSCOPE OLYMPUS PCF TYPE H180ALOlympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum2 recalls25 reports
- EVIS EXERA II COLONOVIDEOSCOPE OLYMPUS PCF TYPE Q180ALOlympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid2 recalls6 reports
- EVIS SIGMOIDOVIDEOSCOPE OLYMPUS PCF TYPE SOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology2 recalls0 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS PCF-PH190IOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid1 recall47 reports
- EVIS EXERA III COLONOVIDEOSCOPE OLYMPUS PCF-HQ190IOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid1 recall34 reports
- EVIS EXERA II COLONOVIDEOSCOPE OLYMPUS PCF TYPE H180AIOlympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid1 recall17 reports
- EVIS EXERA II COLONOVIDEOSCOPE OLYMPUS PCF TYPE Q180AIOlympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum1 recall1 report
Product lines
Other devices under the same FDA submission
18 devices cleared under K131780, K192793, K980465 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- EVIS EXERA III VIDEO SYSTEM CENTER OLYMPUS CV-190Olympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum1 recall7,688 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS CF-HQ190LOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls2 deaths6,804 reports
- EVIS EXERA III XENON LIGHT SOURCE OLYMPUS CLV-190Olympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid6,617 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS CF-H190IOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid1 recall1,967 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS CF-HQ190IOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid2 recalls1,315 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS CF-H190LOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid1 recall1 death1,194 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS CF-HQ190LOlympus Medical Systems Corp. · Colonoscope And Accessories, Flexible/Rigid250 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS CF-H190LOlympus Medical Systems Corp. · Colonoscope And Accessories, Flexible/Rigid99 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS PCF-H190LOlympus Medical Systems Corp. · Colonoscope And Accessories, Flexible/Rigid86 reports
- EVIS EXERA III XENON LIGHT SOURCE OLYMPUS CLV-190Olympus Medical Systems Corp. · Colonoscope And Accessories, Flexible/Rigid77 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS CF-H190IOlympus Medical Systems Corp. · Colonoscope And Accessories, Flexible/Rigid66 reports
- XENON LAMP MAJ-1817Olympus Medical Systems Corp. · Colonoscope And Accessories, Flexible/Rigid61 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 93222).
- Z-0327-2024 · Open, ClassifiedBronchoscope-for endoscopy and endoscopic surgery within the airways and tracheobronchial tree. Model Numbers: BF-1T150, BF-1T60, BF-1TH190, BF-1TQ170, BF-3C160, BF-3C40, BF-H190…
- Z-0328-2024 · Open, ClassifiedBronchoscope - Ultrasonic: endoscopic real-time ultrasonic imaging, ultrasound guided needle aspiration and other endoscopic procedures within the airways, trancheobronchial tree, …
- Z-0329-2024 · Open, ClassifiedColonoscope: for endoscopy and endoscopic surgery within the lower digestive tract (including the anus, rectum, sigmoid colon, colon and ileocecal valve). Model Numbers: CF-2T160I…
- Z-0330-2024 · Open, ClassifiedCholeodochoscope. Model Numbers: CHF-BP30, CHF-CB30L, CHF-P60.
- Z-0331-2024 · Open, ClassifiedCystonephroscope-endoscopic diagnosis and treatment within the bladder, urethra, and kidney Model Numbers: CYF-5, CYF-VH, CYF-VHR.
- Z-0332-2024 · Open, ClassifiedRhinolaryngoscope-for endoscopic diagnosis and treatment within the nasal lumens and airway anatomy (including nasopharyngeal and trachea). Model Numbers: ENF-T3, ENF-VT2, ENF-V…
- Z-0333-2024 · Open, ClassifiedGastroscope-endoscopic ultrasound (EUS) guided fine needle aspiration (FNA) and for endoscopic surgery within the upper digestive tract. Model Numbers: GF-UC140P-AL5, GF-UCT180, …
- Z-0334-2024 · Open, ClassifiedGastrointestinal videoscope-indicated for use within the upper digestive tract (including the esophagus, stomach, and duodenum). Model Numbers: GIF-1TH190, GIF-1TQ160, GIF-2T160, …
- Z-0335-2024 · Open, ClassifiedHysteroscope-For endoscopic diagnosis within the uterus Model Number: HYF-1T.
- Z-0336-2024 · Open, ClassifiedHysterovideoscope-For endoscopic diagnosis within the uterus Model Number: HYF-V.
- Z-0337-2024 · Open, ClassifiedHysterofiberscope-For endoscopic diagnosis within the uterus. Model Number: HYF-XP.
- Z-0338-2024 · Open, ClassifiedTracheal fiberscope-airway management, which includes observation to access airway anatomy, endotracheal/endobronchial intubation and management. Model Numbers: LF-DP, LF-GP, LF-…
- Z-0339-2024 · Open, ClassifiedTracheal videoscope- For airway management, which includes observation to access airway anatomy, endotracheal/endobronchial intubation and management. Model Number: LF-V.
- Z-0340-2024 · Open, ClassifiedMobile airway scope-diagnosis and observation to access airway anatomy, endotracheal/endobronchial intubation and management. Model Numbers: MAF-DM2, MAF-GM, MAF-GM2, MAF-TM, MAF-…
- Z-0342-2024 · Open, ClassifiedSignmoidovideoscope-For endoscopy and endoscopic surgery within the lower digestive tract (including the anus, rectum and sigmoid colon). Model Number: PCF-S.
- Z-0343-2024 · Open, ClassifiedSmall intestinal videoscope-For endoscopy and endoscopic surgery within the upper digestive tract (including the esophagus, stomach, duodenum, and small intestine) by oral insertio…
- Z-0344-2024 · Open, ClassifiedTransnasal esophagovideoscope-For endoscopy and endoscopic surgery within the esophagus and stomach. Model Number: PEF-V.
- Z-0345-2024 · Open, ClassifiedUltrasonic gastrovideoscope-Real time ultrasound imaging,ultrasound guided needle aspiration and other endoscopic procedures within the upper gastrointestinal tract and surrounding…
- Z-0346-2024 · Open, ClassifiedDuodenovideoscope-EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery Model Numbers: TJF-145, TJF-150, TJF-Q180V…
- Z-0347-2024 · Open, ClassifiedUreteralscope- for endoscopic diagnosis and treatment within the ureter, kidney and biliary tract (common bile duct and hepatic duct). Model Numbers: URF-P6, URF-P6R.
- Z-0348-2024 · Open, ClassifiedUretero-reno fiberscope-For endoscopic diagnosis and treatment within the ureter and kidney. Model Numbers: URF-P7, URF-P7R, URF-V2, URF-V2R, URF-V3, URF-V3R.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page