OES BRONCHOFIBERSCOPE OLYMPUS BF TYPE XP60 (UDI 04953170308154)Recalls & reported problems
Olympus Medical Systems Corp. · Class 2
All models of OES BRONCHOFIBERSCOPE (11)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 04953170308154), received between Sep 10, 2021 and Jul 30, 2026 — classified as 4 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Contamination /Decontamination Problem” (2 reports).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Sep 25, 2023Class IIOpen, Classified
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Z-0327-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Jul 27, 2023Class IIOpen, Classified
Complaint of endoscope model becoming lodged in the endotracheal tube connector due to diameter being too large.
Z-2491-2023 · Root cause: Labeling Change Control · Recall details · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Contamination /Decontamination Problem2
- Break1
- Contamination1
- Corroded1
- Failure to Clean Adequately1
- Peeled/Delaminated1
Patient problems
- Fungal Infection4
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 30, 2026Corroded
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50086717 ↗MalfunctionSep 15, 2024Peeled/Delaminated; Peeled/Delaminated
IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE BRONCHOFIBERSCOPE EXHIBITED COATING OF THE INSERTION TUBE WAS PEELED OFF. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20221104 ↗MalfunctionAug 18, 2022Contamination; Failure to Clean Adequately; Contamination; Failure to Clean Adequately
CUSTOMER RETURNED THE DEVICE FOR EVALUATION AND REPAIR OF A BLURRY IMAGE ISSUE NOTED DURING PREPARATION FOR USE. THERE IS NO PATIENT INVOLVEMENT AND NO HARM REPORTED TO ANY PATIENT. UPON EVALUATION OF THE RETURNED DEVICE, IT WAS OBSERVED THAT THE DISTAL END WAS DIRTY WITH FOREIGN MATERIAL AROUND THE BIOPSY CHANNEL OPENING. THIS MEDWATCH IS BEING SUBMITTED FOR THE ISSUE OF FOREIGN MATERIAL AROUND THE BIOPSY CHANNEL OPENING AS OBSERVED DURING DEVICE EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15250657 ↗MalfunctionNov 25, 2021Break; Break
OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) WAS INFORMED DURING THE INCOMING INSPECTION FOR REPAIR AT OLYMPUS (B)(4), THAT IT WAS FOUND THE COATING OF THE INSERTION TUBE OF THE SUBJECT DEVICE WAS PEELED OFF MORE THAN 1MM2. THE SUBJECT DEVICE HAD BEEN RETURNED TO OLYMPUS CHINA FOR REPAIR OF THE OF THE INSERTION TUBE. THERE WAS NO REPORT OF PATIENT INJURY ASSOCIATED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12876335 ↗InjurySep 10, 2021Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT IS REPORTED BY THE CUSTOMER, AN UNSPECIFIED TIME AFTER A PROCEDURE USING ONE OF THEIR 10 BRONCHOSCOPES, PRELIMINARY FINDINGS OF FUNGAL CULTURE SPECIMEN COLLECTED FROM THE PATIENT VIA BRONCHOALVEOLAR LAVAGE (BAL) FROM RIGHT MIDDLE LOBE (RML) OF LUNG SHOWS: SCANT GROWTH GRAM POSITIVE BRANCHING RODS SUGGESTIVE OF AEROBIC ACTINOMYCETE (ORIGINALLY REPORTED AS MOLD). IT IS UNKNOWN WHAT TREATMENT THE PATIENT REQUIRED AS A RESULT. THE PATIENT¿S CURRENT CONDITION IS UNKNOWN. THE CUSTOMER STATES THAT IT IS NOT KNOWN WHICH OF THEIR 10 BRONCHOSCOPES WAS USED IN THE CASE, AS THE SURGERY STAFF DID NOT DOCUMENT THIS INFORMATION IN THE PROCEDURE RECORD. FOR THIS REASON, THE CUSTOMER IS REQUESTING ALL 10 BRONCHOSCOPES BE EVALUATED TO DETERMINE IF ANY OF THEM ARE POTENTIALLY A CONTRIBUTING FACTOR TO THE POSITIVE PATIENT CULTURE. ADDITIONAL DETAILS REGARDING THE PATIENT AND EVENT HAVE BEEN REQUESTED. AT THIS TIME, NO FURTHER DETAILS HAVE BEEN PROVIDED. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS A BF-XP60. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS A BF-3C40. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS A BF-3C40. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS A BF-3C40. CASE WITH PATIENT IDENTIFIER
Patient problems: Fungal Infection
MAUDE report 12455801 ↗InjurySep 10, 2021Adverse Event Without Identified Device or Use Problem
IT IS REPORTED BY THE CUSTOMER, AN UNSPECIFIED TIME AFTER A PROCEDURE USING ONE OF THEIR 10 BRONCHOSCOPES, PRELIMINARY FINDINGS OF FUNGAL CULTURE SPECIMEN COLLECTED FROM THE PATIENT VIA BRONCHOALVEOLAR LAVAGE (BAL) FROM RIGHT MIDDLE LOBE (RML) OF LUNG SHOWS: SCANT GROWTH GRAM POSITIVE BRANCHING RODS SUGGESTIVE OF AEROBIC ACTINOMYCETE (ORIGINALLY REPORTED AS MOLD). IT IS UNKNOWN WHAT TREATMENT THE PATIENT REQUIRED AS A RESULT. THE PATIENT¿S CURRENT CONDITION IS UNKNOWN. THE CUSTOMER STATES THAT IT IS NOT KNOWN WHICH OF THEIR 10 BRONCHOSCOPES WAS USED IN THE CASE, AS THE SURGERY STAFF DID NOT DOCUMENT THIS INFORMATION IN THE PROCEDURE RECORD. FOR THIS REASON, THE CUSTOMER IS REQUESTING ALL 10 BRONCHOSCOPES BE EVALUATED TO DETERMINE IF ANY OF THEM ARE POTENTIALLY A CONTRIBUTING FACTOR TO THE POSITIVE PATIENT CULTURE. ADDITIONAL DETAILS REGARDING THE PATIENT AND EVENT HAVE BEEN REQUESTED. AT THIS TIME, NO FURTHER DETAILS HAVE BEEN PROVIDED. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS A BF-XP60. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS A BF-3C40. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS A BF-3C40. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS A BF-3C40. CASE WITH PATIENT IDENTIFIER
Patient problems: Fungal Infection
MAUDE report 12455757 ↗InjurySep 10, 2021Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT IS REPORTED BY THE CUSTOMER, AN UNSPECIFIED TIME AFTER A PROCEDURE USING ONE OF THEIR 10 BRONCHOSCOPES, PRELIMINARY FINDINGS OF FUNGAL CULTURE SPECIMEN COLLECTED FROM THE PATIENT VIA BRONCHOALVEOLAR LAVAGE (BAL) FROM RIGHT MIDDLE LOBE (RML) OF LUNG SHOWS: SCANT GROWTH GRAM POSITIVE BRANCHING RODS SUGGESTIVE OF AEROBIC ACTINOMYCETE (ORIGINALLY REPORTED AS MOLD). IT IS UNKNOWN WHAT TREATMENT THE PATIENT REQUIRED AS A RESULT. THE PATIENT¿S CURRENT CONDITION IS UNKNOWN. THE CUSTOMER STATES THAT IT IS NOT KNOWN WHICH OF THEIR 10 BRONCHOSCOPES WAS USED IN THE CASE, AS THE SURGERY STAFF DID NOT DOCUMENT THIS INFORMATION IN THE PROCEDURE RECORD. FOR THIS REASON, THE CUSTOMER IS REQUESTING ALL 10 BRONCHOSCOPES BE EVALUATED TO DETERMINE IF ANY OF THEM ARE POTENTIALLY A CONTRIBUTING FACTOR TO THE POSITIVE PATIENT CULTURE. ADDITIONAL DETAILS REGARDING THE PATIENT AND EVENT HAVE BEEN REQUESTED. AT THIS TIME, NO FURTHER DETAILS HAVE BEEN PROVIDED. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS A BF-XP60. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS A BF-3C40. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS A BF-3C40. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS A BF-3C40. CASE WITH PATIENT IDENTIFIER
Patient problems: Fungal Infection
MAUDE report 12453495 ↗InjurySep 10, 2021Adverse Event Without Identified Device or Use Problem
IT IS REPORTED BY THE CUSTOMER, AN UNSPECIFIED TIME AFTER A PROCEDURE USING ONE OF THEIR 10 BRONCHOSCOPES, PRELIMINARY FINDINGS OF FUNGAL CULTURE SPECIMEN COLLECTED FROM THE PATIENT VIA BRONCHOALVEOLAR LAVAGE (BAL) FROM RIGHT MIDDLE LOBE (RML) OF LUNG SHOWS: SCANT GROWTH GRAM POSITIVE BRANCHING RODS SUGGESTIVE OF AEROBIC ACTINOMYCETE (ORIGINALLY REPORTED AS MOLD). IT IS UNKNOWN WHAT TREATMENT THE PATIENT REQUIRED AS A RESULT. THE PATIENT¿S CURRENT CONDITION IS UNKNOWN. THE CUSTOMER STATES THAT IT IS NOT KNOWN WHICH OF THEIR 10 BRONCHOSCOPES WAS USED IN THE CASE, AS THE SURGERY STAFF DID NOT DOCUMENT THIS INFORMATION IN THE PROCEDURE RECORD. FOR THIS REASON, THE CUSTOMER IS REQUESTING ALL 10 BRONCHOSCOPES BE EVALUATED TO DETERMINE IF ANY OF THEM ARE POTENTIALLY A CONTRIBUTING FACTOR TO THE POSITIVE PATIENT CULTURE. ADDITIONAL DETAILS REGARDING THE PATIENT AND EVENT HAVE BEEN REQUESTED. AT THIS TIME, NO FURTHER DETAILS HAVE BEEN PROVIDED. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS A BF-XP60. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS A BF-3C40. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS A BF-3C40. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS A BF-3C40. CASE WITH PATIENT IDENTIFIER
Patient problems: Fungal Infection
MAUDE report 12453461 ↗
Other BRONCHOSCOPE (FLEXIBLE OR RIGID) devices
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)3 recalls5 deaths1,557 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-H190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls1 death1,214 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-XP190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls1,146 reports
- Ion 380748Intuitive Surgical, Inc. · Bronchoscope (flexible or rigid)1 recall97 deaths1,278 reports
- VISERA ELITE XENON LIGHT SOURCE OLYMPUS CLV-S190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)1,155 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-P190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)5 recalls3 deaths621 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Olympus Medical Systems Corp. · Bronchoscope (Flexible Or Rigid)3 recalls333 reports
- BRONCHOVIDEOSCOPE BF-1T150Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls212 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page