OES BRONCHOFIBERSCOPE OLYMPUS BF TYPE XP60 (UDI 04953170308154)Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of OES BRONCHOFIBERSCOPE (11)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
8
Deaths reported
0
Injuries reported
4
Recalls naming it
2

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 04953170308154), received between Sep 10, 2021 and Jul 30, 2026 — classified as 4 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Contamination /Decontamination Problem” (2 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 25, 2023Class IIOpen, Classified

    Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

    Z-0327-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Jul 27, 2023Class IIOpen, Classified

    Complaint of endoscope model becoming lodged in the endotracheal tube connector due to diameter being too large.

    Z-2491-2023 · Root cause: Labeling Change Control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Contamination /Decontamination Problem2
  • Break1
  • Contamination1
  • Corroded1
  • Failure to Clean Adequately1
  • Peeled/Delaminated1

Patient problems

  • Fungal Infection4
  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 30, 2026Corroded

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50086717 ↗
  • MalfunctionSep 15, 2024Peeled/Delaminated; Peeled/Delaminated

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE BRONCHOFIBERSCOPE EXHIBITED COATING OF THE INSERTION TUBE WAS PEELED OFF. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20221104 ↗
  • MalfunctionAug 18, 2022Contamination; Failure to Clean Adequately; Contamination; Failure to Clean Adequately

    CUSTOMER RETURNED THE DEVICE FOR EVALUATION AND REPAIR OF A BLURRY IMAGE ISSUE NOTED DURING PREPARATION FOR USE. THERE IS NO PATIENT INVOLVEMENT AND NO HARM REPORTED TO ANY PATIENT. UPON EVALUATION OF THE RETURNED DEVICE, IT WAS OBSERVED THAT THE DISTAL END WAS DIRTY WITH FOREIGN MATERIAL AROUND THE BIOPSY CHANNEL OPENING. THIS MEDWATCH IS BEING SUBMITTED FOR THE ISSUE OF FOREIGN MATERIAL AROUND THE BIOPSY CHANNEL OPENING AS OBSERVED DURING DEVICE EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15250657 ↗
  • MalfunctionNov 25, 2021Break; Break

    OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) WAS INFORMED DURING THE INCOMING INSPECTION FOR REPAIR AT OLYMPUS (B)(4), THAT IT WAS FOUND THE COATING OF THE INSERTION TUBE OF THE SUBJECT DEVICE WAS PEELED OFF MORE THAN 1MM2. THE SUBJECT DEVICE HAD BEEN RETURNED TO OLYMPUS CHINA FOR REPAIR OF THE OF THE INSERTION TUBE. THERE WAS NO REPORT OF PATIENT INJURY ASSOCIATED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12876335 ↗
  • InjurySep 10, 2021Contamination /Decontamination Problem; Contamination /Decontamination Problem

    IT IS REPORTED BY THE CUSTOMER, AN UNSPECIFIED TIME AFTER A PROCEDURE USING ONE OF THEIR 10 BRONCHOSCOPES, PRELIMINARY FINDINGS OF FUNGAL CULTURE SPECIMEN COLLECTED FROM THE PATIENT VIA BRONCHOALVEOLAR LAVAGE (BAL) FROM RIGHT MIDDLE LOBE (RML) OF LUNG SHOWS: SCANT GROWTH GRAM POSITIVE BRANCHING RODS SUGGESTIVE OF AEROBIC ACTINOMYCETE (ORIGINALLY REPORTED AS MOLD). IT IS UNKNOWN WHAT TREATMENT THE PATIENT REQUIRED AS A RESULT. THE PATIENT¿S CURRENT CONDITION IS UNKNOWN. THE CUSTOMER STATES THAT IT IS NOT KNOWN WHICH OF THEIR 10 BRONCHOSCOPES WAS USED IN THE CASE, AS THE SURGERY STAFF DID NOT DOCUMENT THIS INFORMATION IN THE PROCEDURE RECORD. FOR THIS REASON, THE CUSTOMER IS REQUESTING ALL 10 BRONCHOSCOPES BE EVALUATED TO DETERMINE IF ANY OF THEM ARE POTENTIALLY A CONTRIBUTING FACTOR TO THE POSITIVE PATIENT CULTURE. ADDITIONAL DETAILS REGARDING THE PATIENT AND EVENT HAVE BEEN REQUESTED. AT THIS TIME, NO FURTHER DETAILS HAVE BEEN PROVIDED. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS A BF-XP60. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS A BF-3C40. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS A BF-3C40. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS A BF-3C40. CASE WITH PATIENT IDENTIFIER

    Patient problems: Fungal Infection

    MAUDE report 12455801 ↗
  • InjurySep 10, 2021Adverse Event Without Identified Device or Use Problem

    IT IS REPORTED BY THE CUSTOMER, AN UNSPECIFIED TIME AFTER A PROCEDURE USING ONE OF THEIR 10 BRONCHOSCOPES, PRELIMINARY FINDINGS OF FUNGAL CULTURE SPECIMEN COLLECTED FROM THE PATIENT VIA BRONCHOALVEOLAR LAVAGE (BAL) FROM RIGHT MIDDLE LOBE (RML) OF LUNG SHOWS: SCANT GROWTH GRAM POSITIVE BRANCHING RODS SUGGESTIVE OF AEROBIC ACTINOMYCETE (ORIGINALLY REPORTED AS MOLD). IT IS UNKNOWN WHAT TREATMENT THE PATIENT REQUIRED AS A RESULT. THE PATIENT¿S CURRENT CONDITION IS UNKNOWN. THE CUSTOMER STATES THAT IT IS NOT KNOWN WHICH OF THEIR 10 BRONCHOSCOPES WAS USED IN THE CASE, AS THE SURGERY STAFF DID NOT DOCUMENT THIS INFORMATION IN THE PROCEDURE RECORD. FOR THIS REASON, THE CUSTOMER IS REQUESTING ALL 10 BRONCHOSCOPES BE EVALUATED TO DETERMINE IF ANY OF THEM ARE POTENTIALLY A CONTRIBUTING FACTOR TO THE POSITIVE PATIENT CULTURE. ADDITIONAL DETAILS REGARDING THE PATIENT AND EVENT HAVE BEEN REQUESTED. AT THIS TIME, NO FURTHER DETAILS HAVE BEEN PROVIDED. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS A BF-XP60. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS A BF-3C40. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS A BF-3C40. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS A BF-3C40. CASE WITH PATIENT IDENTIFIER

    Patient problems: Fungal Infection

    MAUDE report 12455757 ↗
  • InjurySep 10, 2021Contamination /Decontamination Problem; Contamination /Decontamination Problem

    IT IS REPORTED BY THE CUSTOMER, AN UNSPECIFIED TIME AFTER A PROCEDURE USING ONE OF THEIR 10 BRONCHOSCOPES, PRELIMINARY FINDINGS OF FUNGAL CULTURE SPECIMEN COLLECTED FROM THE PATIENT VIA BRONCHOALVEOLAR LAVAGE (BAL) FROM RIGHT MIDDLE LOBE (RML) OF LUNG SHOWS: SCANT GROWTH GRAM POSITIVE BRANCHING RODS SUGGESTIVE OF AEROBIC ACTINOMYCETE (ORIGINALLY REPORTED AS MOLD). IT IS UNKNOWN WHAT TREATMENT THE PATIENT REQUIRED AS A RESULT. THE PATIENT¿S CURRENT CONDITION IS UNKNOWN. THE CUSTOMER STATES THAT IT IS NOT KNOWN WHICH OF THEIR 10 BRONCHOSCOPES WAS USED IN THE CASE, AS THE SURGERY STAFF DID NOT DOCUMENT THIS INFORMATION IN THE PROCEDURE RECORD. FOR THIS REASON, THE CUSTOMER IS REQUESTING ALL 10 BRONCHOSCOPES BE EVALUATED TO DETERMINE IF ANY OF THEM ARE POTENTIALLY A CONTRIBUTING FACTOR TO THE POSITIVE PATIENT CULTURE. ADDITIONAL DETAILS REGARDING THE PATIENT AND EVENT HAVE BEEN REQUESTED. AT THIS TIME, NO FURTHER DETAILS HAVE BEEN PROVIDED. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS A BF-XP60. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS A BF-3C40. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS A BF-3C40. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS A BF-3C40. CASE WITH PATIENT IDENTIFIER

    Patient problems: Fungal Infection

    MAUDE report 12453495 ↗
  • InjurySep 10, 2021Adverse Event Without Identified Device or Use Problem

    IT IS REPORTED BY THE CUSTOMER, AN UNSPECIFIED TIME AFTER A PROCEDURE USING ONE OF THEIR 10 BRONCHOSCOPES, PRELIMINARY FINDINGS OF FUNGAL CULTURE SPECIMEN COLLECTED FROM THE PATIENT VIA BRONCHOALVEOLAR LAVAGE (BAL) FROM RIGHT MIDDLE LOBE (RML) OF LUNG SHOWS: SCANT GROWTH GRAM POSITIVE BRANCHING RODS SUGGESTIVE OF AEROBIC ACTINOMYCETE (ORIGINALLY REPORTED AS MOLD). IT IS UNKNOWN WHAT TREATMENT THE PATIENT REQUIRED AS A RESULT. THE PATIENT¿S CURRENT CONDITION IS UNKNOWN. THE CUSTOMER STATES THAT IT IS NOT KNOWN WHICH OF THEIR 10 BRONCHOSCOPES WAS USED IN THE CASE, AS THE SURGERY STAFF DID NOT DOCUMENT THIS INFORMATION IN THE PROCEDURE RECORD. FOR THIS REASON, THE CUSTOMER IS REQUESTING ALL 10 BRONCHOSCOPES BE EVALUATED TO DETERMINE IF ANY OF THEM ARE POTENTIALLY A CONTRIBUTING FACTOR TO THE POSITIVE PATIENT CULTURE. ADDITIONAL DETAILS REGARDING THE PATIENT AND EVENT HAVE BEEN REQUESTED. AT THIS TIME, NO FURTHER DETAILS HAVE BEEN PROVIDED. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS A BF-XP60. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS A BF-3C40. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS A BF-3C40. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS A BF-3C40. CASE WITH PATIENT IDENTIFIER

    Patient problems: Fungal Infection

    MAUDE report 12453461 ↗

Other BRONCHOSCOPE (FLEXIBLE OR RIGID) devices

All BRONCHOSCOPE (FLEXIBLE OR RIGID) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page