Bronchoscope-for endoscopy and endoscopic surgery within… recallZ-0327-2024 · Aizu Olympus Co., Ltd. · initiated Sep 25, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Aizu Olympus Co., Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
FDA root cause category: Under Investigation by firm
What the company is doing
Olympus America issued Urgent Medical Device Correction letter on Sept. 25, 2023 to Endoscopy, Respiratory, Pulmonary and Urology Departments, Risk Management Department. Letter states reason for recall, health risk and action to take: Our records indicate that Olympus has serviced your flexible endoscope at one of our Olympus repair facilities. Following repair activities, Olympus uses a water flush of endoscope channels as part of our final endoscope channel inspection, and we subsequently dry the endoscope channels with compressed filtered air. Olympus discovered that our channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Olympus is in the process of implementing a validated drying process at our repair centers. In the meantime, Olympus has established interim processes to perform an additional inspection of your scope for water prior to return. 1. Ensure your users are knowledgeable on the content of this letter. 2. Complete the enclosed response form and return it to our recall partner, Sedgwick, via e-mail at [email protected] or by fax at 844-294-7184. For any questions about the acknowledgement form, please call the Sedgwick team at 888-667-1504. 3. Olympus requests that you report complaints, including infections, and endoscopes returned wet from our repair centers to Olympus. Please report complaints to Technical Assistance Center (TAC) at 1-800-848-9024, option 1. Adverse events experienced with the use of this product may also be reported to the Food and Drug Administration MedWatch Adverse Event Reporting program either online, by regular mail or by fax. If you require additional information, please do not hesitate to contact me at (647) 999-3203 or [email protected].
Recall details
- Recall number
- Z-0327-2024
- Status
- Open, Classified
- Initiated by the company
- Sep 25, 2023
- Posted by the FDA
- Nov 17, 2023
- Recalling firm
- Aizu Olympus Co., Ltd.
- Quantity
- 1123 units
- Product code
- EOQ
- FDA submissions
- K033225, K201758, K963033, K023984, K121959, K172726
- Distribution
- Worldwide - US Nationwide distribution.
Product description
Bronchoscope-for endoscopy and endoscopic surgery within the airways and tracheobronchial tree. Model Numbers: BF-1T150, BF-1T60, BF-1TH190, BF-1TQ170, BF-3C160, BF-3C40, BF-H190, BF-MP190F, BF-MP60, BF-P150, BF-P190, BF-P60, BF-Q170, BF-Q190, BF-XP190, BF-XP60, BF-XT160, BF-XT190. 510(k)s: K023984, K121959, K963033, K172726, K201758, K033225, K183419.
Code information (models, lots, serial numbers)
Model Number (UDI-DI): BF-1T150 (04953170308185), BF-1T60 (04953170308161), BF-1TH190 (04953170335181), BF-1TQ170 (04953170342943), BF-3C160 (04953170340031), BF-3C40 (04953170308130), BF-H190 (04953170335174), BF-MP190F (04953170395581), BF-MP60 (04953170308277), BF-P150 (04953170308178), BF-P190 (04953170342110), BF-P60 (04953170308123), BF-Q170 (04953170342912), BF-Q190 (04953170335198), BF-XP190 (04953170342134), BF-XP60 (04953170308154), BF-XT160 (04953170340147), BF-XT190 (04953170402470).
Devices named in this recall
18 devices in the FDA device registry whose identifier appears in the recall's code information.
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)3 recalls5 deaths1,557 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-H190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls1 death1,214 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-XP190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls1,146 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-P190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)5 recalls3 deaths621 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-Q190Olympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum5 recalls280 reports
- BRONCHOVIDEOSCOPE BF-1T150Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls212 reports
- BRONCHOVIDEOSCOPE BF-Q170Olympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum5 recalls151 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-XT190Olympus Medical Systems Corp. · Bronchoscope (flexible or rigid)5 recalls2 deaths127 reports
- BRONCHOVIDEOSCOPE BF-1TQ170Olympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum2 recalls160 reports
- EVIS EXERA BRONCHOVIDEOSCOPE OLYMPUS BF TYPE XT160Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls1 death41 reports
- EVIS EXERA BRONCHOVIDEOSCOPE OLYMPUS BF TYPE 3C160Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)3 recalls45 reports
- BRONCHOVIDEOSCOPE BF-P150Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)3 recalls37 reports
- EVIS EXERA III BRONCHOFIBERVIDEOSCOPE OLYMPUS BF-MP190FOlympus Medical Systems Corp. · Bronchoscope (flexible or rigid)2 recalls82 reports
- OES BRONCHOFIBERSCOPE OLYMPUS BF TYPE P60Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)2 recalls11 reports
- OES BRONCHOFIBERSCOPE OLYMPUS BF TYPE XP60Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)2 recalls8 reports
- OES BRONCHOFIBERSCOPE OLYMPUS BF TYPE 1T60Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)2 recalls7 reports
- OES BRONCHOFIBERSCOPE OLYMPUS BF TYPE 3C40Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)2 recalls1 report
- OES BRONCHOFIBERSCOPE OLYMPUS BF TYPE MP60Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)2 recalls1 report
Product lines
- EVIS EXERA III BRONCHOVIDEOSCOPEOlympus Medical Systems Corp. · 11 models · 6,060 reports · 16 recalls
- OES BRONCHOFIBERSCOPEOlympus Medical Systems Corp. · 11 models · 211 reports · 9 recalls
- BRONCHOVIDEOSCOPEOlympus Medical Systems Corp. · 8 models · 880 reports · 13 recalls
- EVIS EXERA BRONCHOVIDEOSCOPEOlympus Medical Systems Corp. · 3 models · 91 reports · 6 recalls
Other devices under the same FDA submission
6 devices cleared under K033225, K201758, K963033, K023984, K121959, K172726 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- EVIS EXERA III VIDEO SYSTEM CENTER OLYMPUS CV-190Olympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum1 recall7,688 reports
- EVIS EXERA III XENON LIGHT SOURCE OLYMPUS CLV-190Olympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid6,617 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Olympus Medical Systems Corp. · Bronchoscope (Flexible Or Rigid)3 recalls333 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-H190Olympus Medical Systems Corp. · Bronchoscope (Flexible Or Rigid)2 recalls240 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-Q190Olympus Medical Systems Corp. · Bronchoscope (Flexible Or Rigid)2 recalls69 reports
- EVIS EXERA III XENON LIGHT SOURCE OLYMPUS CLV-190Olympus Medical Systems Corp. · Colonoscope And Accessories, Flexible/Rigid77 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 93222).
- Z-0328-2024 · Open, ClassifiedBronchoscope - Ultrasonic: endoscopic real-time ultrasonic imaging, ultrasound guided needle aspiration and other endoscopic procedures within the airways, trancheobronchial tree, …
- Z-0329-2024 · Open, ClassifiedColonoscope: for endoscopy and endoscopic surgery within the lower digestive tract (including the anus, rectum, sigmoid colon, colon and ileocecal valve). Model Numbers: CF-2T160I…
- Z-0330-2024 · Open, ClassifiedCholeodochoscope. Model Numbers: CHF-BP30, CHF-CB30L, CHF-P60.
- Z-0331-2024 · Open, ClassifiedCystonephroscope-endoscopic diagnosis and treatment within the bladder, urethra, and kidney Model Numbers: CYF-5, CYF-VH, CYF-VHR.
- Z-0332-2024 · Open, ClassifiedRhinolaryngoscope-for endoscopic diagnosis and treatment within the nasal lumens and airway anatomy (including nasopharyngeal and trachea). Model Numbers: ENF-T3, ENF-VT2, ENF-V…
- Z-0333-2024 · Open, ClassifiedGastroscope-endoscopic ultrasound (EUS) guided fine needle aspiration (FNA) and for endoscopic surgery within the upper digestive tract. Model Numbers: GF-UC140P-AL5, GF-UCT180, …
- Z-0334-2024 · Open, ClassifiedGastrointestinal videoscope-indicated for use within the upper digestive tract (including the esophagus, stomach, and duodenum). Model Numbers: GIF-1TH190, GIF-1TQ160, GIF-2T160, …
- Z-0335-2024 · Open, ClassifiedHysteroscope-For endoscopic diagnosis within the uterus Model Number: HYF-1T.
- Z-0336-2024 · Open, ClassifiedHysterovideoscope-For endoscopic diagnosis within the uterus Model Number: HYF-V.
- Z-0337-2024 · Open, ClassifiedHysterofiberscope-For endoscopic diagnosis within the uterus. Model Number: HYF-XP.
- Z-0338-2024 · Open, ClassifiedTracheal fiberscope-airway management, which includes observation to access airway anatomy, endotracheal/endobronchial intubation and management. Model Numbers: LF-DP, LF-GP, LF-…
- Z-0339-2024 · Open, ClassifiedTracheal videoscope- For airway management, which includes observation to access airway anatomy, endotracheal/endobronchial intubation and management. Model Number: LF-V.
- Z-0340-2024 · Open, ClassifiedMobile airway scope-diagnosis and observation to access airway anatomy, endotracheal/endobronchial intubation and management. Model Numbers: MAF-DM2, MAF-GM, MAF-GM2, MAF-TM, MAF-…
- Z-0341-2024 · Open, ClassifiedColonoscope-indicated for use within the lower digestive tract (including the anus, rectum, sigmoid colon, colon, and ileocecal valve). Model Numbers: PCF-H180AI, PCF-H180AL, PCF-…
- Z-0342-2024 · Open, ClassifiedSignmoidovideoscope-For endoscopy and endoscopic surgery within the lower digestive tract (including the anus, rectum and sigmoid colon). Model Number: PCF-S.
- Z-0343-2024 · Open, ClassifiedSmall intestinal videoscope-For endoscopy and endoscopic surgery within the upper digestive tract (including the esophagus, stomach, duodenum, and small intestine) by oral insertio…
- Z-0344-2024 · Open, ClassifiedTransnasal esophagovideoscope-For endoscopy and endoscopic surgery within the esophagus and stomach. Model Number: PEF-V.
- Z-0345-2024 · Open, ClassifiedUltrasonic gastrovideoscope-Real time ultrasound imaging,ultrasound guided needle aspiration and other endoscopic procedures within the upper gastrointestinal tract and surrounding…
- Z-0346-2024 · Open, ClassifiedDuodenovideoscope-EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery Model Numbers: TJF-145, TJF-150, TJF-Q180V…
- Z-0347-2024 · Open, ClassifiedUreteralscope- for endoscopic diagnosis and treatment within the ureter, kidney and biliary tract (common bile duct and hepatic duct). Model Numbers: URF-P6, URF-P6R.
- Z-0348-2024 · Open, ClassifiedUretero-reno fiberscope-For endoscopic diagnosis and treatment within the ureter and kidney. Model Numbers: URF-P7, URF-P7R, URF-V2, URF-V2R, URF-V3, URF-V3R.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page