BRONCHOVIDEOSCOPE BF-P150 (UDI 04953170308178)Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Olympus BF-P150 bronchoscope is a flexible scope used to examine the airways. Reports mostly describe missing, poor or intermittent images, along with peeling or detached parts; recent reports also mention poor suction and worn openings. There are 36 reports, with no deaths or injuries. Three recalls are open: one concerns unclear instructions for use with certain treatment equipment, one addresses potentially wet channels after repair and infection risk, and one concerns the scope getting stuck in a breathing-tube connector.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
37
Deaths reported
0
Injuries reported
0
Recalls naming it
3

Summary

The FDA's MAUDE database contains 37 adverse event reports that list this exact device (DI 04953170308178), received between Oct 6, 2020 and Sep 23, 2026 — classified as 37 malfunctions.

The most frequently reported specific device problem is “No Display/Image” (17 reports).

3 FDA recall records name this device's identifier, 3 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 2 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 4 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 1 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 2 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 11, 2025Open, Classified

    Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.

    Z-0055-2026 · Root cause: Device Design · FDA record ↗

  2. Sep 25, 2023Open, Classified

    Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

    Z-0327-2024 · Root cause: Under Investigation by firm · FDA record ↗

  3. Jul 27, 2023Open, Classified

    Complaint of endoscope model becoming lodged in the endotracheal tube connector due to diameter being too large.

    Z-2490-2023 · Root cause: Labeling Change Control · FDA record ↗

Most reported problems

Device problems

  • No Display/Image17
  • Peeled/Delaminated5
  • Poor Quality Image5
  • Erratic or Intermittent Display3
  • Detachment of Device or Device Component2
  • Material Deformation2
  • Break1
  • Communication or Transmission Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions36
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 25 most recent of 37 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 23, 2026Corroded; Scratched Material

    The customer returned the Bronchovideoscope to the service center for a separate issue. During the device analysis, the Service Engineer indicated that the forceps port had scratches and corrosion was found on the port. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50872513 ↗
  • MalfunctionAug 31, 2026Insufficient Flow or Under Infusion

    It was reported that during reprocessing, the Broncho videoscope had poor aspiration with narrow scope cavity. There were no reports of patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50525074 ↗
  • MalfunctionAug 4, 2026Decrease in Suction

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50136238 ↗
  • MalfunctionMar 23, 2026Material Deformation

    IT WAS OBSERVED THAT DURING THE EVALUATION, THE BRONCHOVIDEOSCOPE EXHIBITED THAT THE FORCEPS PORT IS WORN. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24660990 ↗
  • MalfunctionJan 5, 2026Poor Quality Image

    IT WAS REPORTED THE BRONCHOVIDEOSCOPE IMAGE APPEARED NOISY DURING SETUP/INSPECTION PRIOR TO USE. THERE WAS NO REPORTED PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23956186 ↗
  • MalfunctionDec 26, 2025Erratic or Intermittent Display; Poor Quality Image; Erratic or Intermittent Display; Poor Quality Image

    IT WAS REPORTED THE BRONCHOVIDEOSCOPE IMAGE WAS ABNORMAL. THE EVENT OCCURRED DURING SET UP IN ROOM. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23897947 ↗
  • MalfunctionNov 20, 2025Degraded; Degraded

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE BRONCHOVIDEOSCOPE EXHIBITED THE INSERTION TUBE WAS PEELING. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23609518 ↗
  • MalfunctionOct 22, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE BRONCHOVIDEOSCOPE EXHIBITED THAT THE ADHESIVE ON THE BENDING SECTION COVER (A-RUBBER) IS DETACHED. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23358135 ↗
  • MalfunctionSep 14, 2025Loss of or Failure to Bond; Poor Quality Image; Component Missing; Detachment of Device or Device Component; Loss of or Failure to Bond; Poor Quality Image; Component Missing; Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE ADHESIVE ON A-RUBBER WAS DETACHED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23048209 ↗
  • MalfunctionAug 12, 2025No Display/Image; No Display/Image

    IT WAS OBSERVED DURING THE DEVICE INSPECTION THAT SUBJECT DEVICE HAD A BLACKOUT DURING ANGULATION ADJUSTMENT. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22790717 ↗
  • MalfunctionAug 11, 2025Poor Quality Image; Poor Quality Image

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE EXHIBITED A NOISY IMAGE DUE TO LOW CHARGE-COUPLED DEVICE PERFORMANCE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22775089 ↗
  • MalfunctionAug 5, 2025Erratic or Intermittent Display; No Display/Image; Erratic or Intermittent Display; No Display/Image

    IT WAS OBSERVED DURING OLYMPUS' DEVICE INSPECTION THAT THE BRONCHOVIDEOSCOPE'S IMAGE WAS INTERMITTENTLY FLICKERING AND BLACKING OUT. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22718742 ↗
  • MalfunctionAug 1, 2025No Display/Image; No Display/Image

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE BRONCHOVIDEOSCOPE EXHIBITED IMAGE ISSUES INTERMITTENTLY DUE TO DAMAGE ON THE CHARGE COUPLED DEVICE (CCD) UNIT, AND THE IMAGE IS NOT DISPLAYED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22685815 ↗
  • MalfunctionJul 23, 2025Erratic or Intermittent Display; Erratic or Intermittent Display

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE BRONCHOVIDEOSCOPE EXHIBITED IMAGE FLICKERING. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22579513 ↗
  • MalfunctionJul 6, 2025Peeled/Delaminated; Unstable; Peeled/Delaminated; Unstable

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE BRONCHOVIDEOSCOPE EXHIBITED AN UNSECURED DISTAL END DUE TO ADHESIVE PEELING OFF THE DISTAL END. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22409097 ↗
  • MalfunctionFeb 25, 2025No Display/Image; No Display/Image

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21462081 ↗
  • MalfunctionJan 17, 2025No Display/Image; No Display/Image

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21180776 ↗
  • MalfunctionAug 12, 2024No Display/Image; No Display/Image

    IT WAS REPORTED THAT THE BRONCHOVIDEOSCOPE SCREEN BECOMES BLACK, AND NO IMAGES ARE DISPLAYED. THE ISSUE OCCURRED DURING PREPARATION FOR USE FOR A DIAGNOSTIC BRONCHOSCOPY THAT WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19969003 ↗
  • MalfunctionJun 26, 2024No Display/Image; No Display/Image

    IT WAS REPORTED, THE BRONCHOVIDEOSCOPE DISPLAYED A BLACK SCREEN. THE ISSUE OCCURRED, DURING PREPARATION FOR USE OF A DIAGNOSTIC BRONCHOSCOPY PROCEDURE. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WAS NO DELAY. PATIENT WAS NOT UNDER ANESTHESIA. AND THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19614410 ↗
  • MalfunctionJun 6, 2024No Display/Image; No Display/Image

    IT WAS REPORTED, THE FLEX VIDEO SCOPE SCREEN BLACK OUT WITH NOISE DURING ANGULATION. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19479681 ↗
  • MalfunctionMay 21, 2024No Display/Image; No Display/Image

    IT WAS REPORTED THE BRONCHOVIDEOSCOPE HAD THE SCREEN BLACK OUT WITHOUT IMAGE. THE ISSUE OCCURRED DURING PREPARATION FOR USE ON NON-SPECIFIED THERAPEUTIC PROCEDURE AND WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19367183 ↗
  • MalfunctionJan 9, 2024No Display/Image; No Display/Image

    THE CUSTOMER REPORTED THAT THE PATIENT WAS NOT UNDER ANESTHESIA AND NO DELAY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18473045 ↗
  • MalfunctionDec 7, 2023No Display/Image; No Display/Image

    THE CUSTOMER REPORTED TO OLYMPUS, THE BRONCHOVIDEOSCOPE HAD NO IMAGE DURING ANGULATION. THE DEVICE WAS RETURNED FOR EVALUATION. DURING THE DEVICE EVALUATION, DAMAGE TOT HE ELECTRICAL CONNECTER AND CHARGED COUPLED DEVICE (CCD) CAUSED IMAGE TO NOT BE DISPLAYED. THERE WERE NO REPORTS OF PATIENT HARM. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18288774 ↗
  • MalfunctionOct 31, 2023Poor Quality Image; Poor Quality Image

    THE CUSTOMER REPORTED TO OLYMPUS, THE BRONCHOVIDEOSCOPE DISPLAYED A NOISY IMAGE. THE ISSUE WAS FOUND DURING REPROCESSING. THE INTENDED PROCEDURE WAS A DIAGNOSTIC BRONCHOSCOPY. THE PROCEDURE WAS COMPLETED WITH A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18042361 ↗
  • MalfunctionAug 7, 2023No Display/Image; No Display/Image

    AN OLYMPUS REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER THAT THE IMAGE WAS NOISY WHEN BRONCHOVIDEOSCOPE WAS POWERED ON AND THE OBJECT WAS NOT VISIBLE. THE PATIENT WAS NOT UNDER SEDATION AND THERE WAS NO DELAY. THERE WERE NO REPORTS OF PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. THE DEVICE WAS RETURNED, AND OLYMPUS REPAIR CENTER IDENTIFIED THE IMAGE BLACKED OUT. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTIONS FOUND DURING THE EVALUATION OF THE RETURNED DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17470931 ↗