Quick summary
This Olympus EVIS EXERA II bronchoscope is a flexible scope used to look at the airways. Most of its 48 reports describe the scope failing to produce an image or a noisy, blacked-out picture, plus a peeling outer layer, corroded parts, a leaking biopsy channel and burn marks; no deaths or injuries were reported. One open recall, started in July 2023, says the scope can get stuck in an endotracheal tube connector because it is too wide.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 48
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 48 adverse event reports that list this exact device (DI 04953170308192), received between Jul 16, 2021 and Aug 26, 2026 — classified as 48 malfunctions.
The most frequently reported specific device problem is “No Display/Image” (30 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Jul 27, 2023Class IIOpen, Classified
Complaint of endoscope model becoming lodged in the endotracheal tube connector due to diameter being too large.
Z-2493-2023 · Root cause: Labeling Change Control · Recall details · FDA record ↗
Most reported problems
Device problems
- No Display/Image30
- Peeled/Delaminated6
- Corroded3
- Detachment of Device or Device Component2
- Erratic or Intermittent Display2
- Failure to Clean Adequately2
- Material Split, Cut or Torn2
- Break1
Patient problems
- No Clinical Signs, Symptoms or Conditions48
Recent adverse event reports
The 25 most recent of 48 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 26, 2026Corroded
The customer returned the BRONCHOVIDEOSCOPE to the service center for a separate issue. During the device analysis, the investigator indicated the connecting tube had corrosion. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50456827 ↗MalfunctionAug 25, 2026Device Reprocessing Problem; Corroded; Improper Flow or Infusion
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50436483 ↗MalfunctionAug 5, 2026Corroded
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50142805 ↗MalfunctionJul 29, 2026Detachment of Device or Device Component
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50067628 ↗MalfunctionMar 31, 2026Suction Failure
THE BRONCHOVIDEOSCOPE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT OF A LEAKING BIOPSY CHANNEL. THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER. DURING INSPECTION OF THE DEVICE, IT WAS FOUND THAT THE SUCTION VOLUME DID NOT MEET THE STANDARD VALUE DUE TO DAMAGE ON THE CHANNEL TUBE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24741660 ↗MalfunctionJan 6, 2026Erratic or Intermittent Display
IT WAS OBSERVED DURING THE DEVICE EVALUATION, THAT THE BRONCHOVIDEOSCOPE DISPLAYED A NOISY IMAGE DUE TO DAMAGE ON CCD UNIT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23986466 ↗MalfunctionDec 8, 2025Melted; Melted
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SCOPE EXHIBITED BURN MARKS ON DISTAL END PROBE COVER (C-COVER). THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23737046 ↗MalfunctionAug 21, 2025No Display/Image; No Display/Image
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE BRONCHOVIDEOSCOPE EXHIBITED A BLACKOUT WHEN THE POWER WAS TURNED ON. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22865954 ↗MalfunctionJul 24, 2025Material Discolored; Material Deformation; Sharp Edges; Material Discolored; Material Deformation; Sharp Edges
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE BRONCHOVIDEOSCOPE EXHIBITED DAMAGE ON BENDING TUBE, METAL PART IS STICKING OUT FROM BREAKAGE ON A-RUBBER) AND CORROSION ON THE CONNECTING TUBE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22601991 ↗MalfunctionMay 7, 2025No Display/Image; Peeled/Delaminated; No Display/Image; Peeled/Delaminated
IT WAS REPORTED, THE BRONCHOVIDEOSCOPE EXHIBITED NO IMAGE. THE ISSUE OCCURRED DURING SERVICE/MAINTENANCE (NOT IN A CLINICAL SETTING). THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21975736 ↗MalfunctionMar 28, 2025No Display/Image; No Display/Image
IT WAS REPORTED THAT THERE WAS NO IMAGE ON THE SUBJECT DEVICE. THIS OCCURRED DURING MAINTENANCE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21724077 ↗MalfunctionOct 28, 2024No Display/Image; No Display/Image
IT WAS REPORTED THE BRONCHOVIDEOSCOPE HAD NO IMAGE. THE ISSUE OCCURRED DURING PREPARATION FOR USE IN A DIAGNOSTIC BRONCHOSCOPY PROCEDURE. THERE WAS NO DELAY, AND THE PROCEDURE WAS COMPLETED WITH A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20548651 ↗MalfunctionAug 26, 2024Thermal Decomposition of Device; Thermal Decomposition of Device
IT WAS OBSERVED THAT DURING THE DEVICE INSPECTION, THE BRONCHOVIDEOSCOPE EXHIBITED BURNED PLASTIC DISTAL END COVER AROUND INSTRUMENT CHANNEL. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20065213 ↗MalfunctionJul 25, 2024Peeled/Delaminated; Peeled/Delaminated
IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE BRONCHOVIDEOSCOPE EXHIBITED PEELING OF THE COATING ON THE INSERTION TUBE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19827079 ↗MalfunctionJul 3, 2024Peeled/Delaminated; Peeled/Delaminated
IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE BRONCHOVIDEOSCOPE INSERTION TUBE EXHIBITED EXCESSIVE PEELING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19665724 ↗MalfunctionMay 2, 2024No Display/Image; No Display/Image
IT WAS REPORTED, THE BRONCHOVIDEOSCOPE HAD NO IMAGE. THE ISSUE OCCURRED DURING AN UNSPECIFIED EVENT. THERE WAS NO REPORT OF DELAY, THE PATIENT WAS NOT UNDER ANESTHESIA. THERE WERE NO REPORTS OF PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19234340 ↗MalfunctionMar 23, 2024No Display/Image; No Display/Image
IT WAS REPORTED THAT AFTER THE BRONCHOVIDEOSCOPE WAS PLUGGED IN AND TURNED ON THE IMAGE WENT FUZZY, THEN BLACK AND NO IMAGE RETURNED. IT WAS VERIFIED THAT WITH A DIFFERENT SCOPE THE MACHINE WORKS PROPERLY. THE ISSUE OCCURRED DURING PREPARATION FOR USE FOR A BRONCHOALVEOLAR LAVAGE PROCEDURE THAT WAS COMPLETED USING A SIMILAR DEVICE. THERE WAS NO DELAY REPORTED, AND NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18966336 ↗MalfunctionMar 14, 2024No Display/Image; No Display/Image
IT WAS REPORTED, THE BRONCHOVIDEOSCOPES IMAGE WAS NOT VISIBLE AT ALL WITH A BLACK SCREEN. THE ISSUE OCCURRED DURING PREPARATION FOR A THERAPEUTIC BRONCHOSCOPY. THE PROCEDURE WAS COMPLETED WITH A SIMILAR DEVICE WITH NO DELAY. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18903322 ↗MalfunctionJan 23, 2024No Display/Image; No Display/Image
THE CUSTOMER REPORTED TO OLYMPUS, THE BRONCHOVIDEOSCOPE HAD NO IMAGE. THE ISSUE WAS FOUND DURING PREPARATION FOR USE FOR AN UNSPECIFIED PROCEDURE. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WAS NO DELAY IN THE PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18563318 ↗MalfunctionNov 30, 2023No Display/Image; No Display/Image
THE CUSTOMER REPORTED TO OLYMPUS, THERE WAS NO IMAGE AFTER TURNING ON THE COMPUTER WHEN USING THE BRONCHOVIDEOSCOPE DURING A DIAGNOSTIC TRACHEOSCOPY PROCEDURE. THE DEVICE WAS COMPLETED USING A SIMILAR DEVICE. IT WAS REPORTED THAT AFTER REPLACING THE VIDEO SYSTEM CENTER, THE FAILURE DID NOT REOCCUR. THERE WERE NO REPORTS OF PATIENT HARM ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18235378 ↗MalfunctionNov 15, 2023No Display/Image; No Display/Image
IT WAS REPORTED TO OLYMPUS, THAT THE EVIS EXERA II BRONCHOVIDEOSCOPE DID NOT DISPLAY IMAGE AND THE ENTIRE SCREEN WAS BLACK. THE ISSUE WAS FOUND DURING INSPECTION FOR USE FOR A DIAGNOSTIC ENDOBRONCHIAL ULTRA SOUND AND IT WAS COMPLETED WITH ANOTHER DEVICE. THERE WAS NO PATIENT HARM OR USER INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18144857 ↗MalfunctionOct 31, 2023No Display/Image; No Display/Image
THE CUSTOMER REPORTED TO OLYMPUS, THE EVIS EXERA II BRONCHOVIDEOSCOPE HAD A BLACK SCREEN. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18043915 ↗MalfunctionOct 30, 2023No Display/Image; No Display/Image
THE OLYMPUS FIELD SERVICE ENGINEER REPORTED, DURING ON SITE TROUBLESHOOTING, GENERATION OF NOISE WAS SEEN ON MONITOR WHEN BRONCHOVIDEOSCOPE WAS CONNECTED TO 190 SYSTEM. FLUID INVASION WAS FOUND AT ELECTRICAL CONNECTORS AND WATER-RESISTANCE CAP. ELECTRICAL CONNECTOR WAS CLEANED AND DRIED, THEN RECONNECTED TO 190 SYSTEM, NO IMAGE WAS SHOWN (BLACK SCREEN). THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18035340 ↗MalfunctionOct 5, 2023Dent in Material; Dent in Material
IT WAS REPORTED THAT THE PATIENT BIT DOWN ON THE BRONCHOVIDEOSCOPE IN MULTIPLE PLACES DURING THE PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17883040 ↗MalfunctionOct 2, 2023No Display/Image; No Display/Image
THE CUSTOMER REPORTED TO OLYMPUS THE EVIS EXERA II BRONCHOVIDEOSCOPE, IMAGE PROBLEM, NO CONTACT TAKING PICTURES. THE ISSUE OCCURRED DURING REPROCESSING. THE PROCEDURE WAS UNKNOWN. THERE WERE NO REPORTS OF PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. INSPECTION AND TESTING OF THE RETURNED DEVICE FOUND THAT THERE WAS NO IMAGE. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTIONS FOUND DURING THE EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17855252 ↗
Other NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID) devices
- ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE OLYMPUS LTF-S190-10Olympus Medical Systems Corp. · Nasopharyngoscope (Flexible Or Rigid)785 reports
- NA 242268Medos International Sàrl · Nasopharyngoscope (flexible or rigid)573 reports
- EVIS EXERA II BRONCHOVIDEOSCOPE OLYMPUS BF TYPE 1T180Olympus Medical Systems Corp. · NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID)3 recalls1 death408 reports
- VIVIDEO VNL9-CPHoya Corporation · Nasopharyngoscope (Flexible Or Rigid)389 reports
- PENTAX FNL-10RP3Hoya Corporation · NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID)322 reports
- SINGLE USE BIOPSY VALVE MAJ-210Olympus Medical Systems Corp. · Nasopharyngoscope (Flexible Or Rigid)1 recall221 reports
- SINGLE USE SUCTION VALVE MAJ-209Olympus Medical Systems Corp. · Nasopharyngoscope (Flexible Or Rigid)228 reports
- PENTAX FNL-7RP3Hoya Corporation · NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID)227 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page