EVIS EXERA II BRONCHOVIDEOSCOPERecalls & reported problems across all models
Olympus Medical Systems Corp. · 12 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
10 open recalls name at least one EVIS EXERA II BRONCHOVIDEOSCOPE model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 671
- Deaths reported
- 1
- Injuries reported
- 68
- Recalls
- 10
We found 12 models of EVIS EXERA II BRONCHOVIDEOSCOPE by Olympus Medical Systems Corp. in the FDA's device registry (GUDID). Together they are named in 671 adverse event reports in the FDA's MAUDE database, received between Jul 13, 2018 and Sep 30, 2026 — classified as 1 death, 68 injuries, 602 malfunctions.
The most frequently reported specific device problem is “No Display/Image” (269 reports).
10 FDA recall records name at least one of these models by its device identifier, 10 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
EVIS EXERA II BRONCHOVIDEOSCOPE recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Sep 11, 2025Open, Classified
Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.
Z-0062-2026 · Root cause: Device Design · Recall details · FDA record ↗
- Sep 11, 2025Open, Classified
Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.
Z-0061-2026 · Root cause: Device Design · Recall details · FDA record ↗
- Sep 11, 2025Open, Classified
Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.
Z-0056-2026 · Root cause: Device Design · Recall details · FDA record ↗
- Sep 11, 2025Open, Classified
Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.
Z-0051-2026 · Root cause: Device Design · Recall details · FDA record ↗
- Sep 11, 2025Open, Classified
Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.
Z-0046-2026 · Root cause: Device Design · Recall details · FDA record ↗
- Jul 27, 2023Open, Classified
Complaint of endoscope model becoming lodged in the endotracheal tube connector due to diameter being too large.
Z-2493-2023 · Root cause: Labeling Change Control · names 10 of these models · Recall details · FDA record ↗
- Jul 27, 2023Open, Classified
Complaint of endoscope model becoming lodged in the endotracheal tube connector due to diameter being too large.
Z-2492-2023 · Root cause: Labeling Change Control · Recall details · FDA record ↗
- Jun 8, 2023Open, Classified
There have been complaints of endobronchial combustion during therapeutic laser procedures with the bronchoscope.
Z-2016-2023 · Root cause: Under Investigation by firm · names 5 of these models · Recall details · FDA record ↗
- Jun 8, 2023Open, Classified
There have been complaints of endobronchial combustion during therapeutic laser procedures with the bronchoscope.
Z-2011-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Mar 8, 2022Open, Classified
Bronchofiberscope Updated instructions on the use of certain chemicals to be used for reprocessing, conditions for ETO gas sterilization, rinsing steps after high-level disinfection, and a recommendation on using sterilization
Z-1049-2022 · Root cause: Other · Recall details · FDA record ↗
Most reported EVIS EXERA II BRONCHOVIDEOSCOPE problems
Device problems
- No Display/Image269
- Thermal Decomposition of Device110
- Peeled/Delaminated62
- Microbial Contamination of Device53
- Failure to Clean Adequately35
- Device Reprocessing Problem26
- Erratic or Intermittent Display22
- Detachment of Device or Device Component21
Patient problems
- No Clinical Signs, Symptoms or Conditions590
- Bacterial Infection31
- Unspecified Infection18
- No Patient Involvement10
- Respiratory Tract Infection6
- Fungal Infection5
- No Known Impact Or Consequence To Patient3
- Pneumothorax3
EVIS EXERA II BRONCHOVIDEOSCOPE models
12 models, most reported first. Each model has its own page with its reports, recalls and device details.
- EVIS EXERA II BRONCHOVIDEOSCOPE OLYMPUS BF TYPE 1T180NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID)3 recalls1 death408 reports
- EVIS EXERA II BRONCHOVIDEOSCOPE OLYPUS BF TYPE P180Endoscope, accessories, narrow band spectrum4 recalls85 reports
- EVIS EXERA II BRONCHOVIDEOSCOPE OLYMPUS BF TYPE 1TQ180BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls64 reports
- EVIS EXERA II BRONCHOVIDEOSCOPE OLYMPUS BF TYPE Q180-ACBRONCHOSCOPE (FLEXIBLE OR RIGID)3 recalls27 reports
- EVIS EXERA II BRONCHOVIDEOSCOPE OLYMPUS BF TYPE Q180BRONCHOSCOPE (FLEXIBLE OR RIGID)2 recalls12 reports
- EVIS EXERA II BRONCHOVIDEOSCOPE OLYPUS BF TYPE P180NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID)1 recall48 reports
- EVIS EXERA II BRONCHOVIDEOSCOPE OLYMPUS BF TYPE 1T180NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID)1 recall14 reports
- EVIS EXERA II BRONCHOVIDEOSCOPE OLYMPUS BF TYPE 1T180BRONCHOSCOPE (FLEXIBLE OR RIGID)1 recall8 reports
- EVIS EXERA II BRONCHOVIDEOSCOPE OLYMPUS BF TYPE Q180BRONCHOSCOPE (FLEXIBLE OR RIGID)1 recall2 reports
- EVIS EXERA II BRONCHOVIDEOSCOPE OLYMPUS BF TYPE Q180BRONCHOSCOPE (FLEXIBLE OR RIGID)1 recall2 reports
- EVIS EXERA II BRONCHOVIDEOSCOPE OLYMPUS BF TYPE 1TQ180BRONCHOSCOPE (FLEXIBLE OR RIGID)1 recall1 report
- EVIS EXERA II BRONCHOVIDEOSCOPE OLYMPUS BF TYPE 1TQ180NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID)1 recall0 reports
Recent EVIS EXERA II BRONCHOVIDEOSCOPE adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Thermal Decomposition of Device
The customer reported the Bronchovideoscope had a separate issue. During the initial evaluation of the device it was observed, C-cover had a burn. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50947558 ↗MalfunctionSep 30, 2026Thermal Decomposition of Device
The customer reported C-cover burnt during service/maintenance involving an Olympus Bronchovideoscope. There was no patient injury reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50944260 ↗MalfunctionSep 25, 2026Material Deformation
The customer reported the Bronchovideoscope had a separate issue. During the initial evaluation of the device it was observed, distal end plastic cover was chipped. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50897877 ↗MalfunctionSep 24, 2026Break; Material Separation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50889001 ↗MalfunctionSep 22, 2026Device Reprocessing Problem
It was reported that the Bronchovideoscope was sterilized by SPD and the cap was not removed, resulting in a damaged Scope. There was no report of patient harm or involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50849930 ↗MalfunctionSep 18, 2026Erratic or Intermittent Display; No Display/Image
It was reported by the customer that there was no image on the scope. The issue was found during set up for an unspecified procedure. During the initial evaluation of the device it was observed, that image noise occurs. No harm reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50826766 ↗MalfunctionSep 2, 2026Detachment of Device or Device Component
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50576360 ↗MalfunctionSep 2, 2026No Display/Image; Loose or Intermittent Connection
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50563525 ↗MalfunctionSep 1, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50542661 ↗MalfunctionAug 31, 2026No Display/Image
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50521299 ↗MalfunctionAug 26, 2026Corroded
The customer returned the BRONCHOVIDEOSCOPE to the service center for a separate issue. During the device analysis, the investigator indicated the connecting tube had corrosion. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50456827 ↗MalfunctionAug 25, 2026Device Reprocessing Problem; Corroded; Improper Flow or Infusion
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50436483 ↗MalfunctionAug 13, 2026No Display/Image
Patient problems: Insufficient Information
MAUDE report 50230619 ↗MalfunctionAug 10, 2026No Display/Image
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50189504 ↗MalfunctionAug 5, 2026Corroded
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50142805 ↗
Other Olympus Medical Systems Corp. products
- EVIS EXERA lll COLONOVIDEOSCOPE16 models · 15,013 reports · 7 recalls
- EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE7 models · 13,448 reports · 7 recalls
- EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE9 models · 8,317 reports · 2 recalls
- EVIS EXERA III BRONCHOVIDEOSCOPE11 models · 6,060 reports · 16 recalls
- EVIS EXERA III XENON LIGHT SOURCE2 models · 6,694 reports
- EVIS EXERA III COLONOVIDEOSCOPE6 models · 6,117 reports · 5 recalls
- EVIS EXERA III DUODENOVIDEOSCOPE OLYMPUS TJF-Q190V2 models · 5,281 reports · 4 recalls
- HD CAMERA HEAD2 models · 5,217 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page