EVIS EXERA II BRONCHOVIDEOSCOPE OLYMPUS BF TYPE 1T180 (UDI 04953170308215)Recalls & reported problems
Olympus Medical Systems Corp. · Class 2
All models of EVIS EXERA II BRONCHOVIDEOSCOPE (12)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 04953170308215), received between Dec 21, 2023 and Sep 30, 2026 — classified as 14 malfunctions.
The most frequently reported specific device problem is “Thermal Decomposition of Device” (5 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Jul 27, 2023Class IIOpen, Classified
Complaint of endoscope model becoming lodged in the endotracheal tube connector due to diameter being too large.
Z-2493-2023 · Root cause: Labeling Change Control · Recall details · FDA record ↗
Most reported problems
Device problems
- Thermal Decomposition of Device5
- Degraded2
- Erratic or Intermittent Display2
- No Display/Image2
- Poor Quality Image2
- Corroded1
- Decrease in Suction1
- Detachment of Device or Device Component1
Patient problems
- No Clinical Signs, Symptoms or Conditions14
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Thermal Decomposition of Device
The customer reported the Bronchovideoscope had a separate issue. During the initial evaluation of the device it was observed, C-cover had a burn. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50947558 ↗MalfunctionSep 30, 2026Thermal Decomposition of Device
The customer reported C-cover burnt during service/maintenance involving an Olympus Bronchovideoscope. There was no patient injury reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50944260 ↗MalfunctionSep 18, 2026Erratic or Intermittent Display; No Display/Image
It was reported by the customer that there was no image on the scope. The issue was found during set up for an unspecified procedure. During the initial evaluation of the device it was observed, that image noise occurs. No harm reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50826766 ↗MalfunctionApr 3, 2026Poor Quality Image
NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24775689 ↗MalfunctionApr 2, 2026Corroded; Poor Quality Image; Detachment of Device or Device Component
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE BRONCHOVIDEOSCOPE EXHIBITED A CHIP ON DISTAL END, C-COVER HAS CORROSION AND THERE WAS A NOISY IMAGE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24770949 ↗MalfunctionMar 24, 2026Degraded; Melted
IT WAS OBSERVED DURING THE DEVICE EVALUATION, THE BRONCHOVIDEOSCOPE EXHIBITED THE CONNECTING TUBE HAD THE COATING PEELING AND THE DISTAL END HAD A BURNED. THERE WAS NO PATIENT INVOLVEMENT
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24681653 ↗MalfunctionMar 9, 2026Degraded; Material Deformation
IT WAS OBSERVED DURING THE DEVICE EVALUATION, THE BRONCHOVIDEOSCOPE EXHIBITED THE COATING OF THE CONNECTING TUBE WAS PEELING AND THE FORCEPS CHANNEL PORT WAS SHAVED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24546664 ↗MalfunctionMar 6, 2026Thermal Decomposition of Device
IT WAS REPORTED DURING REPROCESSING, THE PLASTIC DISTAL END COVER OF THE BRONCHOVIDEOSCOPE WAS BURNED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24530477 ↗MalfunctionSep 12, 2025Decrease in Suction; Positioning Failure; Material Discolored; Misfocusing; Low Test Results; Flare or Flash; Gas/Air Leak; Temperature Problem; Decrease in Suction; Positioning Failure; Material Discolored; Misfocusing; Low Test Results; Flare or Flash; Gas/Air Leak; Temperature Problem
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE HAD A BURN ON THE DISTAL END. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23035575 ↗MalfunctionAug 22, 2025No Display/Image; No Display/Image
IT WAS OBSERVED THAT DURING THE DEVICE INSPECTION, THE BRONCHOVIDEOSCOPE EXHIBITED DUE TO DAMAGE ON THE CCD UNIT, THE IMAGE IS NOT DISPLAYED. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22877475 ↗MalfunctionMay 13, 2025Erratic or Intermittent Display; Erratic or Intermittent Display
IT WAS REPORTED THAT THE BRONCHOVIDEOSCOPE EXPERIENCED INTERMITTENT IMAGE LOSS. THE ISSUE OCCURRED DURING SET UP / INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22011724 ↗MalfunctionFeb 22, 2024Thermal Decomposition of Device; Leak/Splash; Thermal Decomposition of Device; Leak/Splash
IT WAS REPORTED THAT THE BRONCHOVIDEOSCOPE HAD A LEAK ON THE BENDING SECTION COVER OR DISTAL SHEATH, AND THE PLASTIC DISTAL END COVER HAD A BURN. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18758551 ↗MalfunctionJan 24, 2024Device Reprocessing Problem; Device Reprocessing Problem
THE REPORTED ISSUE WAS FOUND AFTER THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18572217 ↗MalfunctionDec 21, 2023Thermal Decomposition of Device; Thermal Decomposition of Device
THE CUSTOMER REPORTED TO OLYMPUS, THE BRONCHOVIDEOSCOPE DISTAL END SCOPE HAD BURNING. THE CUSTOMER SUSPECTED THAT THE BURNING CAME FROM AN ELECTROSURGERY UNIT WHICH WAS USED DURING THE PREVIOUS CASE. THE ISSUE WAS FOUND DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18377913 ↗
Other NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID) devices
- ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE OLYMPUS LTF-S190-10Olympus Medical Systems Corp. · Nasopharyngoscope (Flexible Or Rigid)785 reports
- NA 242268Medos International Sàrl · Nasopharyngoscope (flexible or rigid)573 reports
- EVIS EXERA II BRONCHOVIDEOSCOPE OLYMPUS BF TYPE 1T180Olympus Medical Systems Corp. · NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID)3 recalls1 death408 reports
- VIVIDEO VNL9-CPHoya Corporation · Nasopharyngoscope (Flexible Or Rigid)389 reports
- PENTAX FNL-10RP3Hoya Corporation · NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID)322 reports
- SINGLE USE BIOPSY VALVE MAJ-210Olympus Medical Systems Corp. · Nasopharyngoscope (Flexible Or Rigid)1 recall221 reports
- SINGLE USE SUCTION VALVE MAJ-209Olympus Medical Systems Corp. · Nasopharyngoscope (Flexible Or Rigid)228 reports
- PENTAX FNL-7RP3Hoya Corporation · NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID)227 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page