The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 04953170308215), received between Dec 21, 2023 and Sep 30, 2026 — classified as 14 malfunctions.

The most frequently reported specific device problem is “Thermal Decomposition of Device” (5 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 1 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 3 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 3 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 27, 2023Class IIOpen, Classified

    Complaint of endoscope model becoming lodged in the endotracheal tube connector due to diameter being too large.

    Z-2493-2023 · Root cause: Labeling Change Control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Thermal Decomposition of Device5
  • Degraded2
  • Erratic or Intermittent Display2
  • No Display/Image2
  • Poor Quality Image2
  • Corroded1
  • Decrease in Suction1
  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions14

Recent adverse event reports

The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Thermal Decomposition of Device

    The customer reported the Bronchovideoscope had a separate issue. During the initial evaluation of the device it was observed, C-cover had a burn. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50947558 ↗
  • MalfunctionSep 30, 2026Thermal Decomposition of Device

    The customer reported C-cover burnt during service/maintenance involving an Olympus Bronchovideoscope. There was no patient injury reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50944260 ↗
  • MalfunctionSep 18, 2026Erratic or Intermittent Display; No Display/Image

    It was reported by the customer that there was no image on the scope. The issue was found during set up for an unspecified procedure. During the initial evaluation of the device it was observed, that image noise occurs. No harm reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50826766 ↗
  • MalfunctionApr 3, 2026Poor Quality Image

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24775689 ↗
  • MalfunctionApr 2, 2026Corroded; Poor Quality Image; Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE BRONCHOVIDEOSCOPE EXHIBITED A CHIP ON DISTAL END, C-COVER HAS CORROSION AND THERE WAS A NOISY IMAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24770949 ↗
  • MalfunctionMar 24, 2026Degraded; Melted

    IT WAS OBSERVED DURING THE DEVICE EVALUATION, THE BRONCHOVIDEOSCOPE EXHIBITED THE CONNECTING TUBE HAD THE COATING PEELING AND THE DISTAL END HAD A BURNED. THERE WAS NO PATIENT INVOLVEMENT

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24681653 ↗
  • MalfunctionMar 9, 2026Degraded; Material Deformation

    IT WAS OBSERVED DURING THE DEVICE EVALUATION, THE BRONCHOVIDEOSCOPE EXHIBITED THE COATING OF THE CONNECTING TUBE WAS PEELING AND THE FORCEPS CHANNEL PORT WAS SHAVED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24546664 ↗
  • MalfunctionMar 6, 2026Thermal Decomposition of Device

    IT WAS REPORTED DURING REPROCESSING, THE PLASTIC DISTAL END COVER OF THE BRONCHOVIDEOSCOPE WAS BURNED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24530477 ↗
  • MalfunctionSep 12, 2025Decrease in Suction; Positioning Failure; Material Discolored; Misfocusing; Low Test Results; Flare or Flash; Gas/Air Leak; Temperature Problem; Decrease in Suction; Positioning Failure; Material Discolored; Misfocusing; Low Test Results; Flare or Flash; Gas/Air Leak; Temperature Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE HAD A BURN ON THE DISTAL END. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23035575 ↗
  • MalfunctionAug 22, 2025No Display/Image; No Display/Image

    IT WAS OBSERVED THAT DURING THE DEVICE INSPECTION, THE BRONCHOVIDEOSCOPE EXHIBITED DUE TO DAMAGE ON THE CCD UNIT, THE IMAGE IS NOT DISPLAYED. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22877475 ↗
  • MalfunctionMay 13, 2025Erratic or Intermittent Display; Erratic or Intermittent Display

    IT WAS REPORTED THAT THE BRONCHOVIDEOSCOPE EXPERIENCED INTERMITTENT IMAGE LOSS. THE ISSUE OCCURRED DURING SET UP / INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22011724 ↗
  • MalfunctionFeb 22, 2024Thermal Decomposition of Device; Leak/Splash; Thermal Decomposition of Device; Leak/Splash

    IT WAS REPORTED THAT THE BRONCHOVIDEOSCOPE HAD A LEAK ON THE BENDING SECTION COVER OR DISTAL SHEATH, AND THE PLASTIC DISTAL END COVER HAD A BURN. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18758551 ↗
  • MalfunctionJan 24, 2024Device Reprocessing Problem; Device Reprocessing Problem

    THE REPORTED ISSUE WAS FOUND AFTER THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18572217 ↗
  • MalfunctionDec 21, 2023Thermal Decomposition of Device; Thermal Decomposition of Device

    THE CUSTOMER REPORTED TO OLYMPUS, THE BRONCHOVIDEOSCOPE DISTAL END SCOPE HAD BURNING. THE CUSTOMER SUSPECTED THAT THE BURNING CAME FROM AN ELECTROSURGERY UNIT WHICH WAS USED DURING THE PREVIOUS CASE. THE ISSUE WAS FOUND DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18377913 ↗

Other NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID) devices

All NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID) devices →

Sources and updates