TROCAR INTRODUCER FOR LTF 5 MM SCOPE MAJ-1379Recalls & reported problems
Olympus Medical Systems Corp. · MAJ-1379 TROCAR INTRODUCER FOR LTF 5MM S · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04953170316159), received between Jul 26, 2022 and Jul 26, 2022 — classified as 1 injury.
The most frequently reported specific device problem is “Material Fragmentation” (1 report).
1 other recall cover devices cleared under the same FDA submission (K201832); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Material Fragmentation1
Patient problems
- Foreign Body In Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryJul 26, 2022Material Fragmentation; Material Fragmentation
THE CUSTOMER REPORTED TO OLYMPUS, NEAR THE END OF THE THERAPEUTIC LAPAROSCOPY, PART OF THE SUBJECT DEVICE WAS DAMAGED. THE SUBJECT DEVICE WAS ATTACHED TO OLYMPUS ENDOSCOPE LTF-S190-5 AND USED AS AN INSERTION AID. A PORTABLE X-RAY DID NOT CONFIRM ANY DEBRIS IN ABDOMINAL CAVITY, NO SHADOWS SEEN. AFTER THE PROCEDURE WAS COMPLETED, A CT SCAN CONFIRMED PIECES FELL INTO THE ABDOMINAL CAVITY. A REOPERATION WAS PERFORMED (B)(6) 2022 AND THE FOREIGN MATERIAL WAS REMOVED. IT IS UNCLEAR WHEN C0R47 WAS USED BY THE CUSTOMER. THE PATIENT WAS DISCHARGED WITHOUT FURTHER ISSUES (B)(6) 2022.
Patient problems: Foreign Body In Patient
MAUDE report 15106218 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K201832. One submission often covers many models, so check the product and lot details in each record.
- Z-0644-2024 · Nov 20, 2023 · Open, ClassifiedOLYMPUS ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE (5.4 mm diameter) Model: LTF-S190-5 Used with a video system center, light source, documentation equipment, monitor, hand instruments
Other Laparoscope, general & plastic surgery devices
- HIGH DEFINITION LCD MONITOR OEV262HOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,937 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-12EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,886 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-10EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,086 reports
- Auto Suture OMS-T10BTCovidien LP · Laparoscope, general & plastic surgery1 recall955 reports
- PDB OMS-PDB1000Covidien LP · Laparoscope, general & plastic surgery859 reports
- VISERA 4K UHD CAMERA CONTROL UNIT OLYMPUS OTV-S400Olympus Medical Systems Corp. · Laparoscope, general & plastic surgery777 reports
- Auto Suture OMS-T10SBCovidien LP · Laparoscope, general & plastic surgery1 recall697 reports
- Spacemaker SMBTTRNDXCovidien LP · Laparoscope, general & plastic surgery590 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page