CAMERA HEAD OTV-S7H-1D-L08ERecalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of CAMERA HEAD (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Olympus laparoscope camera head is the video unit used with a laparoscope in surgery. Its 182 reports include no deaths and no injuries. Many describe no image or poor image quality, and some describe water leaking in, corrosion or torn cables. No recalls name this device, and there are no reports in the last 12 months, compared with 6 in the prior 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
182
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 182 adverse event reports that list this exact device (DI 04953170323584), received between Jan 7, 2021 and Feb 5, 2025 — classified as 182 malfunctions.

The most frequently reported specific device problem is “No Display/Image” (86 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 3 reportsDec 2023: 4 reportsJan 2024: 24 reportsFeb 2024: 6 reportsMar 2024: 5 reportsApr 2024: 8 reportsMay 2024: 13 reportsJun 2024: 11 reportsJul 2024: 9 reportsAug 2024: 7 reportsSep 2024: 3 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 4 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202624
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • No Display/Image86
  • Poor Quality Image37
  • Leak/Splash12
  • Corroded9
  • Crack9
  • Material Split, Cut or Torn9
  • Disconnection8
  • Break6

Patient problems

  • No Clinical Signs, Symptoms or Conditions179
  • No Consequences Or Impact To Patient2
  • Insufficient Information1
  • No Patient Involvement1

Recent adverse event reports

The 25 most recent of 182 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 5, 2025Poor Quality Image; Poor Quality Image

    IT WAS REPORTED THAT THE SUBJECT DEVICE HAD ABNORMAL IMAGE. THE ISSUE OCCURRED DURING OUT OF THE BOX FAILURE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21314658 ↗
  • MalfunctionJan 21, 2025Leak/Splash; Leak/Splash

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE EXHIBITED A FAILED LEAK TEST. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21200356 ↗
  • MalfunctionJan 13, 2025Gas/Air Leak; Gas/Air Leak

    IT WAS OBSERVED DURING THE DEVICE EVALUATION THAT THE SCOPE MOUNT AND THE IMAGING SYSTEM OF THE SUBJECTED DEVICE WERE FLOODED. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21146392 ↗
  • MalfunctionJan 7, 2025Corroded; Leak/Splash; Material Deformation; Corroded; Leak/Splash; Material Deformation

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE EXHIBITED WATER TIGHTNESS IS LOST, ELECTRICAL CONTACT WAS CORRODED AND CABLE HAD DEFORMATION. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21093076 ↗
  • MalfunctionJan 7, 2025Poor Quality Image; Poor Quality Image

    IT WAS REPORTED THAT THE SUBJECTED DEVICE EXHIBITED A NOISY IMAGE. THE ISSUE WAS FOUND DURING MAINTENANCE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21093000 ↗
  • MalfunctionNov 8, 2024Break; No Display/Image; Mechanical Problem; Break; No Display/Image; Mechanical Problem

    IT WAS REPORTED THAT THE IMAGE ON THE SUBJECT DEVICE DID NOT DISPLAY. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20642707 ↗
  • MalfunctionOct 17, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT THE RUBBER LINING OF THE CAMERA HEAD'S CABLE HAD BEEN RIPPED OFF WHICH WAS FOUND DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20470416 ↗
  • MalfunctionSep 19, 2024No Display/Image; No Display/Image

    IT WAS REPORTED THAT THE CAMERA HEAD DID NOT DISPLAY THE IMAGE. THE PROBLEM WAS FOUND DURING EQUIPMENT INSPECTION. THERE WAS NO REPORT OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20263390 ↗
  • MalfunctionSep 11, 2024No Display/Image; No Display/Image

    IT WAS REPORTED FOR CAMERA HEAD THE IMAGE WAS BLACK. THE ISSUE WAS FOUND DURING PREPARATION FOR USE FOR AN UNSPECIFIED PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20191609 ↗
  • MalfunctionSep 10, 2024Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT THE SUBJECT DEVICE'S CABLE MATERIAL WAS DISINTEGRATED. THE ISSUE WAS IDENTIFIED DURING PREPARATION FOR USE FOR AN UNKNOWN DIAGNOSTIC PROCEDURE. THERE WAS NO SURGICAL DELAY, AND THE PROCEDURE WAS COMPLETED WITH A SIMILAR DEVICE. NO OTHER DEVICES WERE CONNECTED TO THE SUBJECT DEVICE WHEN THE FAILURE OCCURRED, AND NO ERROR MESSAGES WERE OBSERVED. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20180407 ↗
  • MalfunctionAug 28, 2024Leak/Splash; Material Split, Cut or Torn; Leak/Splash; Material Split, Cut or Torn

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE CAMERA HEAD EXHIBITED UNIT FAILED LEAK TEST AND A PUNCTURE ON CABLE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20085488 ↗
  • MalfunctionAug 23, 2024Disconnection; No Display/Image; Disconnection; No Display/Image

    IT WAS REPORTED THAT THE SUBJECT DEVICE HAD CORD DISCONNECTION AND DID NOT DISPLAY IMAGE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20055399 ↗
  • MalfunctionAug 22, 2024Disconnection; Disconnection

    IT WAS REPORTED THAT THE CAMERA HEAD HAD CORD DISCONNECTION. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20042162 ↗
  • MalfunctionAug 22, 2024Crack; Crack

    IT WAS REPORTED THAT THE SUBJECT DEVICE HAD A CRACK WHERE THEY PLUG IT IN. THE EVENT OCCURRED DURING MAINTENANCE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20040883 ↗
  • MalfunctionAug 9, 2024Poor Quality Image; Poor Quality Image

    IT WAS REPORTED THE SUBJECT DEVICE HAD A BLURRY IMAGE. THE ISSUE WAS IDENTIFIED DURING PREPARATION OF USE FOR AN UNSPECIFIED DIAGNOSTIC PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19952064 ↗
  • MalfunctionAug 8, 2024Poor Quality Image; Poor Quality Image

    IT WAS REPORTED THAT WHEN PLUGGED INTO THE CAMERA HEAD, THE SCOPE WAS DIM DESPITE WHITE BALANCING AND THE IMAGE WAS TOO DARK. THE ISSUE OCCURRED DURING AN UNSPECIFIED THERAPEUTIC PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19939899 ↗
  • MalfunctionAug 8, 2024Poor Quality Image; Poor Quality Image

    IT WAS REPORTED THAT THE SUBJECT DEVICE HAD NOISE WHEN USING THE ELECTRIC KNIFE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19939700 ↗
  • MalfunctionJul 19, 2024Corroded; Fracture; Leak/Splash; Gas/Air Leak; Material Split, Cut or Torn; Corroded; Fracture; Leak/Splash; Gas/Air Leak; Material Split, Cut or Torn

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE EXHIBITED THE FOLLOWING ISSUES: THE CAMERA CABLE WAS BROKEN (BURN), THERE WAS AN AIR LEAK IN THE CAMERA CABLE, WATER IMMERSION IN THE MAIN UNIT IMAGING, AS WELL IN THE CAMERA CABLE, THE MAIN UNIT WAS FLOODED, THE SCOPE MOUNT AND THE FREE LOCK LEVER WERE CORRODED, AND THERE WAS A SCRATCH ON THE LENS OF THE MAIN BODY. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19780510 ↗
  • MalfunctionJul 18, 2024Poor Quality Image; Poor Quality Image

    IT WAS REPORTED THAT THE SUBJECT DEVICE DISPLAYED A COMPLETELY DARK IMAGE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19774121 ↗
  • MalfunctionJul 17, 2024Poor Quality Image; Poor Quality Image

    IT WAS REPORTED THAT THE SUBJECT DEVICE HAD IMAGE RUSTLING. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19765255 ↗
  • MalfunctionJul 16, 2024Material Puncture/Hole; Material Puncture/Hole

    IT WAS REPORTED THE SUBJECT DEVICE CABLE COATING WAS BROKEN. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19749594 ↗
  • MalfunctionJul 15, 2024Corroded; Crack; Loose or Intermittent Connection; Poor Quality Image; Material Integrity Problem; Corroded; Crack; Loose or Intermittent Connection; Poor Quality Image; Material Integrity Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE EXHIBITED CONNECTOR CRACKS, ELECTRICAL CONTACT CORROSION, FREE LOCK LEVER CORROSION, AND SCOPE MOUNT CORROSION, FILM RUPTURE, AND IMAGING DETERIORATION. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19742552 ↗
  • MalfunctionJul 9, 2024Gas/Air Leak; Gas/Air Leak

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE EXHIBITED FLOODING DUE TO CABLE SCRATCHES. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19702562 ↗
  • MalfunctionJul 9, 2024No Display/Image; No Display/Image

    IT WAS REPORTED THAT THE CAMERA HEAD SUDDENLY STOPPED DISPLAYING IMAGES ON THE MONITOR AND BECAME UNUSABLE DURING AN UNSPECIFIED TYPE OF PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19700730 ↗
  • MalfunctionJul 3, 2024Disconnection; Disconnection

    IT WAS REPORTED THAT IN THE MIDDLE OF AN UNSPECIFIED THERAPEUTIC PROCEDURE, THE CAMERA HEAD PRESENTED A DISCONNECTION. THE PROCEDURE WAS COMPLETED WITH A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19668482 ↗