Quick summary
This Olympus CYF-5R flexible cystoscope/nephroscope is used to look inside the bladder and kidney. Most of the 55 reports describe cleaning and reprocessing trouble — residue or infection left inside the scope, corroded parts, rust in the instrument channel, chipped glue, and material stuck in a port — with 9 injuries and no deaths. One recall, started December 2023 and still open, updated the manual cleaning instructions after reusable brushes were removed. Reports rose from 10 to 25 over the past 12 months.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 54
- Deaths reported
- 0
- Injuries reported
- 9
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 54 adverse event reports that list this exact device (DI 04953170328817), received between Feb 24, 2019 and Aug 13, 2026 — classified as 9 injuries, 45 malfunctions.
The most frequently reported specific device problem is “Device Reprocessing Problem” (11 reports).
1 FDA recall record name this device's identifier, 1 still open.
2 other recalls cover devices cleared under the same FDA submission (K221690); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Dec 4, 2023Class IIOpen, Classified
Olympus has updated the instructions for use: Reusable cleaning brushes have been removed from the manual cleaning. Single use brushes remain compatible.
Z-0685-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Device Reprocessing Problem11
- Adverse Event Without Identified Device or Use Problem7
- Failure to Clean Adequately6
- Corroded4
- Environmental Particulates3
- Loss of or Failure to Bond3
- Material Integrity Problem3
- Microbial Contamination of Device3
Patient problems
- No Clinical Signs, Symptoms or Conditions42
- Urinary Tract Infection5
- Dysuria3
- No Patient Involvement3
- Bacterial Infection2
- Unspecified Infection2
- Urinary Frequency / Polyuria2
- Urinary Retention2
Recent adverse event reports
The 25 most recent of 54 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 13, 2026Detachment of Device or Device Component
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50244058 ↗MalfunctionAug 13, 2026Failure to Clean Adequately
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50240835 ↗MalfunctionAug 12, 2026Corroded
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50229710 ↗MalfunctionJun 15, 2026Break
THE CUSTOMER RETURNED THE CYSTONEPHROFIBERSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE SERVICE REPAIR TECHNICIAN INDICATES THAT BROKEN/CHIPPED GLUE WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25505567 ↗MalfunctionJun 9, 2026Failure to Clean Adequately
THE CUSTOMER REPORTED SOMETHING STUCK IN THE PORT DURING REPROCESSING INVOLVING AN OLYMPUS CYSTONEPHROFIBERSCOPE. THERE WAS NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25451852 ↗MalfunctionJun 8, 2026Corroded
THE CUSTOMER REPORTED RUST INSIDE INSTRUMENT CHANNEL AND VENTING CONNECTOR DURING REPROCESSING INVOLVING AN OLYMPUS CYSTONEPHROFIBERSCOPE. THERE WAS NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25436417 ↗MalfunctionMay 22, 2026Residue After Decontamination
THE FIBERSCOPE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT OF INTERNAL DISCOLORATION. THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER. DURING INSPECTION OF THE DEVICE, IT WAS FOUND THAT THE PLASTIC DISTAL END CONTROL BODY CONTAINED A RESIDUE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25260070 ↗MalfunctionApr 15, 2026Loss of or Failure to Bond
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE BENDING SECTION COVER ADHESIVE OF THE CYSTO-NEPHRO VIDEOSCOPE WAS MISSING. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24889777 ↗MalfunctionApr 13, 2026Environmental Particulates
THE CYSTONEPHROFIBERSCOPE WAS RETURNED TO THE SERVICE CENTER WITH A COMPLAINT OF LEAKING AND A BROKEN LENS. THIS DOES NOT INDICATE A REPORTABLE EVENT. HOWEVER, A REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER. DURING INSPECTION OF THE DEVICE, IT WAS OBSERVED THAT THE DEVICE HAD A DIRTY LENS (ENVIRONMENTAL PARTICULATES, LENSES). THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24857385 ↗MalfunctionApr 3, 2026Corroded
THE CYSTONEPHROFIBERSCOPE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT OF A FAILED LEAK TEST. THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER. DURING INSPECTION OF THE DEVICE, A CORRODED ETHYLENE OXIDE (GAS) VALVE CONNECTOR WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24781668 ↗MalfunctionMar 31, 2026Crack
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE CYSTONEPHROFIBERSCOPE EXHIBITED A CHIP ON THE END OF THE PLASTIC END DISTAL COVER. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24742228 ↗MalfunctionMar 2, 2026Loss of or Failure to Bond; Crack; Scratched Material
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE ADHESIVE OF THE FIBERSCOPE'S CURVED RUBBER WAS CHIPPED, LIFTED, AND SCRATCHED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24491993 ↗MalfunctionFeb 25, 2026Material Deformation
IT WAS OBSERVED DURING THE DEVICE EVALUATION, THE CYSTO-NEPHRO FIBERSCOPE EXHIBITED THE A-RUBBER GLUE CHIPPED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24442243 ↗InjuryFeb 20, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT AFTER A CYSTOSCOPY PROCEDURE, A PATIENT REPORTED URINARY TRACT INFECTION (UTI) SYMPTOMS 3 DAYS AFTER THE PROCEDURE DATE WITH THE USE OF THIS CYSTONEPHROFIBROSCOPE. ADDITIONAL INFORMATION WAS RECEIVED FROM THE CUSTOMER THAT THE PATIENT EXHIBITED BURNING SENSATION WITH URINATION, URINARY FREQUENCY AND FEELING OF INCOMPLETE VOIDING. THE PATIENT WAS THEN TREATED WITH ANTIBIOTIC FOR 7 DAYS. THERE WERE NO CULTURE TESTS DONE WITH THE PATIENT AND THE CYSTONEPHROFIBROSCOPE. THE CUSTOMER NOTED IT WAS QUITE UNUSUAL TO HAVE UTIS IN THAT SPAN OF TIME. THERE WERE NO FURTHER REPORTS OF PATIENT HARM. THERE WAS NO ALLEGATION OF DEVICE MALFUNCTION. THIS REPORT IS FOR PATIENT 4 OF 4.
Patient problems: Dysuria; Urinary Frequency / Polyuria; Urinary Retention; Urinary Tract Infection
MAUDE report 24406150 ↗InjuryFeb 20, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT AFTER A CYSTOSCOPY PROCEDURE, A PATIENT REPORTED URINARY TRACT INFECTION (UTI) SYMPTOMS 1 DAY AFTER THE PROCEDURE DATE WITH THE USE OF THIS CYSTONEPHROFIBROSCOPE. ADDITIONAL INFORMATION WAS RECEIVED FROM THE CUSTOMER THAT THE PATIENT EXHIBITED BURNING SENSATION WITH URINATION, ABDOMINAL PAIN, AND URINARY URGENCY. THE PATIENT WAS THEN TREATED WITH ANTIBIOTIC FOR 7 DAYS. THERE WERE NO CULTURE TESTS DONE WITH THE PATIENT AND THE CYSTONEPHROFIBROSCOPE. THE CUSTOMER NOTED IT WAS QUITE UNUSUAL TO HAVE UTIS IN THAT SPAN OF TIME. THERE WERE NO FURTHER REPORTS OF PATIENT HARM. THERE WAS NO ALLEGATION OF DEVICE MALFUNCTION. THIS REPORT IS FOR PATIENT 3 OF 4.
Patient problems: Abdominal Pain; Bladder Dysfunction; Dysuria; Urinary Tract Infection
MAUDE report 24406010 ↗InjuryFeb 20, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT AFTER A CYSTOSCOPY PROCEDURE, A PATIENT REPORTED URINARY TRACT INFECTION (UTI) SYMPTOMS 2 DAYS AFTER THE PROCEDURE DATE WITH THE USE OF THIS CYSTONEPHROFIBROSCOPE. ADDITIONAL INFORMATION WAS RECEIVED FROM THE CUSTOMER THAT THE PATIENT EXHIBITED BURNING SENSATION WITH URINATION, URINARY FREQUENCY AND FEELING OF INCOMPLETE VOIDING. THE PATIENT WAS THEN TREATED WITH ANTIBIOTIC FOR 7 DAYS. THERE WERE NO CULTURE TESTS DONE WITH THE PATIENT AND THE CYSTONEPHROFIBROSCOPE. THE CUSTOMER NOTED IT WAS QUITE UNUSUAL TO HAVE UTIS IN THAT SPAN OF TIME. THERE WERE NO FURTHER REPORTS OF PATIENT HARM. THERE WAS NO ALLEGATION OF DEVICE MALFUNCTION. THIS REPORT IS FOR PATIENT 1 OF 4.
Patient problems: Dysuria; Urinary Frequency / Polyuria; Urinary Retention; Urinary Tract Infection
MAUDE report 24405487 ↗MalfunctionFeb 4, 2026Material Separation
IT WAS OBSERVED DURING THE DEVICE INVESTIGATION THAT THE CYSTO-NEPHRO FIBERSCOPE HAD A DETACHED GLUE AT THE DISTAL END OF THE BENDING SECTION. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24259393 ↗MalfunctionFeb 2, 2026Environmental Particulates
IT WAS OBSERVED DURING THE DEVICE EVALUATION, THE CYSTO NEPHRO FIBERSCOPE EXHIBITED A DIRTY OBJECTIVE LENS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24240077 ↗MalfunctionJan 9, 2026Degraded; Material Discolored; Poor Quality Image; Residue After Decontamination; Optical Distortion
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE CYSTONEPHROFIBERSCOPE HAD DIRT ON THE LENS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24027034 ↗MalfunctionDec 16, 2025Material Integrity Problem; Material Integrity Problem
NO ADDITIONAL INFORMATION REC FROM CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23824813 ↗MalfunctionDec 1, 2025Poor Quality Image; Poor Quality Image
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE CYSTO-NEPHRO FIBERSCOPE EXHIBITED A DIRTY OBJECTIVE LENS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23681028 ↗MalfunctionNov 19, 2025Failure to Clean Adequately; Failure to Clean Adequately
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE CYSTONEPHROFIBERSCOPE EXHIBITED RUST INSIDE THE DISTAL END. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23600950 ↗MalfunctionNov 4, 2025Corroded; Corroded
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE FLEXIBLE SCOPE EXHIBITED RUST AND A TEAR INSIDE THE CHANNEL. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23466991 ↗MalfunctionNov 3, 2025Separation Failure; Separation Failure
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE EXHIBITED A BIOPSY PORT THAT FELL OFF. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23453009 ↗MalfunctionOct 29, 2025Failure to Clean Adequately; Failure to Clean Adequately
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE OBJECTIVE LENS OF THE CYSTONEPHROFIBERSCOPE WAS DIRTY. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23415731 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K221690. One submission often covers many models, so check the product and lot details in each record.
- Z-0684-2024 · Dec 4, 2023 · Open, ClassifiedModel No. CYF-5, Cystonephrofiberscope
- Z-0331-2024 · Sep 25, 2023 · Open, ClassifiedCystonephroscope-endoscopic diagnosis and treatment within the bladder, urethra, and kidney Model Numbers: CYF-5, CYF-VH, CYF-VHR.
Other Cystoscope and accessories, flexible/rigid devices
- CYSTO-NEPHRO VIDEOSCOPE OLYMPUS CYF-VHOlympus Medical Systems Corp. · Cystoscope And Accessories, Flexible/Rigid2 recalls1 death1,380 reports
- CYSTO-NEPHRO VIDEOSCOPE OLYMPUS CYF-VHROlympus Medical Systems Corp. · Cystoscope And Accessories, Flexible/Rigid2 recalls377 reports
- VISERA CYSTO-NEPHRO VIDEOSCOPE OLYMPUS CYF TYPE V2Olympus Medical Systems Corp. · Cystoscope And Accessories, Flexible/Rigid1 recall418 reports
- OES CYSTONEPHROFIBERSCOPE OLYMPUS CYF-5Olympus Medical Systems Corp. · Cystoscope and accessories, flexible/rigid1 recall353 reports
- OES CYSTONEPHROFIBERSCOPE OLYMPUS CYF-5Olympus Medical Systems Corp. · Cystoscope and accessories, flexible/rigid1 recall1 death332 reports
- OES CYSTONEPHROFIBERSCOPE OLYMPUS CYF-5Olympus Medical Systems Corp. · Cystoscope and accessories, flexible/rigid2 recalls75 reports
- OES CYSTONEPHROFIBERSCOPE OLYMPUS CYF-5ROlympus Medical Systems Corp. · Cystoscope and accessories, flexible/rigid114 reports
- HOPKINS 27293AAKarl Storz GmbH & Co. KG · Cystoscope and accessories, flexible/rigid2 recalls2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page