Skip to content

CYSTO-NEPHRO VIDEOSCOPE OLYMPUS CYF-VHR (REF N6006240)Recalls & reported problems

Olympus Medical Systems Corp. · CYSTO-NEPHRO VIDEOSCOPE · Class 2

All models of CYSTO-NEPHRO VIDEOSCOPE (7)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Olympus cysto-nephro videoscope views the bladder and kidneys. Its 378 reports mostly describe parts detaching from the scope. The reports include 19 injuries and no deaths. Reports rose from 111 in the previous 12 months to 219 in the last 12 months. There are 2 open recalls about possible water left in scope channels after repair and about cleaning brush instructions.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
377
Deaths reported
0
Injuries reported
19
Recalls naming it
2

Summary

The FDA's MAUDE database contains 377 adverse event reports that list this exact device (DI 04953170411298), received between Oct 2, 2019 and Sep 25, 2026 — classified as 19 injuries, 358 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (215 reports).

2 FDA recall records name this device's identifier, 2 still open.

3 other recalls cover devices cleared under the same FDA submission (K221683); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 2 reportsJun 2024: 4 reportsJul 2024: 2 reportsAug 2024: 6 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 3 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 2 reportsMar 2025: 3 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 3 reportsJul 2025: 2 reportsAug 2025: 12 reportsSep 2025: 10 reportsOct 2025: 74 reportsNov 2025: 28 reportsDec 2025: 8 reportsJan 2026: 24 reportsFeb 2026: 19 reportsMar 2026: 6 reportsApr 2026: 30 reportsMay 2026: 32 reportsJun 2026: 20 reportsJul 2026: 18 reportsAug 2026: 25 reportsSep 2026: 8 reportsOct 2026: 0 reports20242025202674
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 4, 2023Class IIOpen, Classified

    Olympus has updated the instructions for use: Reusable cleaning brushes have been removed from the manual cleaning. Single use brushes remain compatible.

    Z-0690-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Sep 25, 2023Class IIOpen, Classified

    Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

    Z-0331-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Detachment of Device or Device Component215
  • Corroded39
  • Device Reprocessing Problem25
  • Crack18
  • Failure to Clean Adequately18
  • Loss of or Failure to Bond17
  • Peeled/Delaminated14
  • Degraded12

Patient problems

  • No Clinical Signs, Symptoms or Conditions355
  • Bacterial Infection7
  • Dysuria6
  • Fever4
  • Unspecified Infection4
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
  • Foreign Body In Patient2
  • Urinary Tract Infection2

Recent adverse event reports

The 25 most recent of 377 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 25, 2026Stretched

    The customer returned the Cysto-Nephro Videoscope for a separate issue. During the initial evaluation it was observed that the adhesive cover was stretched at the bending section. No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50892998 ↗
  • MalfunctionSep 23, 2026Material Deformation

    The customer reported that the Cysto-Nephro Videoscope had a separate issue. During the initial evaluation of the device, it was observed that the bending section rubber was stretchy. No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50869864 ↗
  • MalfunctionSep 22, 2026Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50859282 ↗
  • MalfunctionSep 4, 2026Moisture Damage

    The customer returned the Cysto-Nephro Videoscope for a separate issue. During the evaluation it was observed that the objective lens had corrosion. No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50604273 ↗
  • MalfunctionSep 3, 2026Collapse; Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50595481 ↗
  • MalfunctionSep 3, 2026Failure to Clean Adequately

    The customer returned the videoscope for a separate issue. During the initial evaluation of the device, it was observed that the objective lens was dirty. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50594187 ↗
  • MalfunctionSep 2, 2026Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50571027 ↗
  • MalfunctionSep 2, 2026Degraded

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50557747 ↗
  • MalfunctionAug 27, 2026Detachment of Device or Device Component

    The customer reported the Cystonephrofiberscope had a separate issue. During the initial evaluation of the device it was observed, that the distal end detached. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50478019 ↗
  • MalfunctionAug 20, 2026Degraded

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50356339 ↗
  • MalfunctionAug 20, 2026Crack; Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50350230 ↗
  • MalfunctionAug 20, 2026Corroded

    It was reported by the customer that the Cysto-Nephro Videoscope distal tip is rusted. It is unknown when the issue was found. There was no patient involvement or patient injury reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50336652 ↗
  • MalfunctionAug 20, 2026Corroded

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50336577 ↗
  • MalfunctionAug 20, 2026Corroded

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50336279 ↗
  • MalfunctionAug 20, 2026Corroded

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50335636 ↗
  • MalfunctionAug 20, 2026Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50335213 ↗
  • MalfunctionAug 19, 2026Detachment of Device or Device Component

    The customer returned the Cysto-Nephro Videoscope for a separate issue. During the initial evaluation of the device, chipped bending section cover glue was observed. No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50334540 ↗
  • MalfunctionAug 19, 2026Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50330279 ↗
  • MalfunctionAug 18, 2026Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50308831 ↗
  • MalfunctionAug 18, 2026Corroded; Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50301918 ↗
  • MalfunctionAug 14, 2026Corroded

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50246525 ↗
  • MalfunctionAug 11, 2026Peeled/Delaminated

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50205927 ↗
  • MalfunctionAug 11, 2026Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50204495 ↗
  • MalfunctionAug 11, 2026Detachment of Device or Device Component

    The customer returned the cysto-nephro videoscope to the service center for a separate issue. During the device analysis, the service repair technician indicated a detached distal end was found. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50202500 ↗
  • MalfunctionAug 10, 2026Detachment of Device or Device Component

    The customer returned the Cysto-Nephro Videoscope to the service center for a separate issue. During the device analysis, the service repair technician indicates that the objective lens glue is missing, and the c-body is detached. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50196365 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K221683. One submission often covers many models, so check the product and lot details in each record.

  • Z-0691-2024 · Dec 4, 2023 · Open, Classified
    Model No. CYF-V2, VISERA Cysto-Nephro Videoscope
  • Z-0692-2024 · Dec 4, 2023 · Open, Classified
    Model No. CYF-V2R, VISERA Cysto-Nephro Videoscope
  • Z-0689-2024 · Dec 4, 2023 · Open, Classified
    Model No. CYF-VH, CYSTO-NEPHRO VIDEOSCOPE

Other Cystoscope And Accessories, Flexible/Rigid devices

All Cystoscope And Accessories, Flexible/Rigid devices →