OES CYSTONEPHROFIBERSCOPE OLYMPUS CYF-5 (UDI 04953170411144)Recalls & reported problems

Olympus Medical Systems Corp. · OES CYSTONEPHROFIBERSCOPE · Class 2

All models of OES CYSTONEPHROFIBERSCOPE (9)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Olympus cystonephrofiberscope has 349 adverse event reports on file, with no deaths and 26 injuries. Reports describe parts detaching, glue peeling inside the scope and trouble cleaning it fully, and some mention urinary infections. 1 open recall names this device: it began in December 2023 and concerns updated cleaning instructions that remove reusable brushes from manual cleaning. Reports in the last 12 months rose to 193 from 72 in the prior 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
353
Deaths reported
0
Injuries reported
26
Recalls naming it
1

Summary

The FDA's MAUDE database contains 353 adverse event reports that list this exact device (DI 04953170411144), received between Jun 4, 2020 and Sep 30, 2026 — classified as 26 injuries, 327 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (64 reports).

1 FDA recall record name this device's identifier, 1 still open.

2 other recalls cover devices cleared under the same FDA submission (K221690); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 6 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 2 reportsOct 2024: 1 reportsNov 2024: 3 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 2 reportsMay 2025: 1 reportsJun 2025: 4 reportsJul 2025: 4 reportsAug 2025: 8 reportsSep 2025: 12 reportsOct 2025: 36 reportsNov 2025: 37 reportsDec 2025: 10 reportsJan 2026: 12 reportsFeb 2026: 21 reportsMar 2026: 15 reportsApr 2026: 24 reportsMay 2026: 23 reportsJun 2026: 17 reportsJul 2026: 8 reportsAug 2026: 18 reportsSep 2026: 12 reportsOct 2026: 0 reports20242025202637
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 4, 2023Class IIOpen, Classified

    Olympus has updated the instructions for use: Reusable cleaning brushes have been removed from the manual cleaning. Single use brushes remain compatible.

    Z-0684-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Detachment of Device or Device Component64
  • Failure to Clean Adequately57
  • Microbial Contamination of Device40
  • Device Reprocessing Problem29
  • Environmental Particulates19
  • Material Split, Cut or Torn19
  • Peeled/Delaminated15
  • Break14

Patient problems

  • No Clinical Signs, Symptoms or Conditions322
  • Unspecified Infection16
  • Urinary Tract Infection10
  • Bacterial Infection5
  • Hematuria5
  • No Patient Involvement5
  • Urinary Retention4
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available3

Recent adverse event reports

The 25 most recent of 353 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Detachment of Device or Device Component

    The customer returned the OES Cystonephrofiberscope for a separate issue. During the initial evaluation of the device it was observed that the bending section cover glue was missing. No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50951656 ↗
  • MalfunctionSep 29, 2026Environmental Particulates

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50934256 ↗
  • MalfunctionSep 25, 2026Corroded

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50903380 ↗
  • MalfunctionSep 25, 2026Stretched

    The customer reported the seal/sheath was hanging over the tip of the scope involving an Olympus Cysto Nephro Fiberscope. The issue occurred during reprocessing. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50897318 ↗
  • MalfunctionSep 23, 2026Naturally Worn; Scratched Material

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50876207 ↗
  • MalfunctionSep 23, 2026Detachment of Device or Device Component

    The customer reported the Cystonephrofiberscope had a separate issue. During the initial evaluation of the device it was observed, that the bending section rubber cover had blown off and was missing. No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50871585 ↗
  • MalfunctionSep 21, 2026Loss of or Failure to Bond

    The customer returned the Cysto Nephro Fiberscope for a separate issue. During the initial evaluation of returned device it was found the bending section cover/distal sheath torn off. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50841488 ↗
  • MalfunctionSep 21, 2026Stretched

    The customer returned the Cysto Nephro Fiberscope for a separate issue. During the initial evaluation of returned device it was found the bending section cover/distal sheath is stretched, bulging. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50841325 ↗
  • MalfunctionSep 21, 2026Material Fragmentation

    The customer returned the Cysto Nephro Fiberscope for a separate issue. During the initial evaluation of returned device it was found the adhesive of the bending section cover/distal sheath has a chip. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50841239 ↗
  • MalfunctionSep 3, 2026Peeled/Delaminated; Environmental Particulates

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50588199 ↗
  • MalfunctionSep 3, 2026Material Fragmentation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50576972 ↗
  • MalfunctionSep 2, 2026Peeled/Delaminated

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50570537 ↗
  • MalfunctionAug 27, 2026Peeled/Delaminated

    The customer returned the Flex Deflectable Videoscope to the service center for evaluation related to a separate issue. During testing, the service repair technician identified the A-rubber glue was peeled. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50481377 ↗
  • MalfunctionAug 24, 2026Obstruction of Flow

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50419473 ↗
  • MalfunctionAug 24, 2026Contamination

    The customer reported the Cystonephrofiberscope had a separate issue. During the initial evaluation of the device it was observed, that the objective lens was dirty. No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50407209 ↗
  • MalfunctionAug 21, 2026Environmental Particulates

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50375373 ↗
  • MalfunctionAug 21, 2026Peeled/Delaminated

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50372950 ↗
  • MalfunctionAug 18, 2026Failure to Clean Adequately

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50303244 ↗
  • MalfunctionAug 17, 2026Environmental Particulates

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50289709 ↗
  • MalfunctionAug 14, 2026Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50261200 ↗
  • MalfunctionAug 14, 2026Corroded; Material Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50257690 ↗
  • MalfunctionAug 12, 2026Loss of or Failure to Bond

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50221788 ↗
  • MalfunctionAug 6, 2026Stretched

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50168130 ↗
  • MalfunctionAug 6, 2026Failure to Clean Adequately

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50168027 ↗
  • MalfunctionAug 6, 2026Failure to Clean Adequately

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50167868 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K221690. One submission often covers many models, so check the product and lot details in each record.

  • Z-0685-2024 · Dec 4, 2023 · Open, Classified
    Model No. CYF-5R, Cystonephrofiberscope
  • Z-0331-2024 · Sep 25, 2023 · Open, Classified
    Cystonephroscope-endoscopic diagnosis and treatment within the bladder, urethra, and kidney Model Numbers: CYF-5, CYF-VH, CYF-VHR.

Other Cystoscope and accessories, flexible/rigid devices

All Cystoscope and accessories, flexible/rigid devices →