Skip to content

EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF TYPE H180J (UDI 04953170339110)Recalls & reported problems

Olympus Medical Systems Corp. · GIF-H180J GASTROSCOPE HDTV, WATER JET · Class 2

All models of EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE (20)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Olympus EVIS EXERA II gastroscope has 248 reports on file, with no deaths and 4 injuries. Many describe the scope failing to clean adequately after reprocessing, and others describe hazy, distorted or missing images. No recalls name this device, and reports rose to 177 in the latest year from 21 in the year before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
231
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 231 adverse event reports that list this exact device (DI 04953170339110), received between Oct 11, 2017 and Sep 24, 2026 — classified as 4 injuries, 227 malfunctions.

The most frequently reported specific device problem is “Failure to Clean Adequately” (65 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 5 reportsDec 2023: 2 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 2 reportsJun 2024: 1 reportsJul 2024: 1 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 3 reportsDec 2024: 0 reportsJan 2025: 3 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 2 reportsJun 2025: 3 reportsJul 2025: 1 reportsAug 2025: 1 reportsSep 2025: 2 reportsOct 2025: 5 reportsNov 2025: 21 reportsDec 2025: 44 reportsJan 2026: 19 reportsFeb 2026: 18 reportsMar 2026: 8 reportsApr 2026: 17 reportsMay 2026: 15 reportsJun 2026: 14 reportsJul 2026: 3 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports20242025202644
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Clean Adequately65
  • Poor Quality Image49
  • No Display/Image47
  • Erratic or Intermittent Display28
  • Device Reprocessing Problem15
  • Microbial Contamination of Device8
  • Contamination6
  • Communication or Transmission Problem5

Patient problems

  • No Clinical Signs, Symptoms or Conditions224
  • No Known Impact Or Consequence To Patient3
  • No Patient Involvement3
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 25 most recent of 231 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 24, 2026Poor Quality Image

    It was reported that during a diagnostic Upper Gastrointestinal procedure (UGD) with use of the gastrointestinal videoscope, the screen would not work. The customer reported that the image was distorted and the freeze control button was not working. The procedure was completed with the same device. There was no report of patient injury or adverse patient effects.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50884394 ↗
  • MalfunctionJul 20, 2026Thermal Decomposition of Device

    THE CUSTOMER RETURNED THE GASTROINTESTINAL VIDEOSCOPE TO THE SERVICE CENTER WITH NO REPORTED COMPLAINT. DURING DEVICE ANALYSIS, THE SERVICE TECHNICIAN IDENTIFIED THAT THE DEVICE HAD A BURN ON THE DISTAL END PLASTIC COVER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25882741 ↗
  • MalfunctionJul 9, 2026No Display/Image

    THE CUSTOMER REPORTED NO IMAGE DURING REPROCESSING INVOLVING A GASTROINTESTINAL VIDEOSCOPE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25769220 ↗
  • MalfunctionJul 7, 2026Crack; Dent in Material; Communication or Transmission Problem; Gas/Air Leak; Positioning Problem; Material Split, Cut or Torn

    IT WAS REPORTED THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED AN UNKNOWN COMMUNICATION ERROR CODE DURING INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25742294 ↗
  • MalfunctionJun 30, 2026Failure to Clean Adequately

    THE CUSTOMER RETURNED THE GASTROINTESTINAL VIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING DEVICE ANALYSIS, THE SERVICE TECHNICIAN IDENTIFIED THAT THE DEVICE HAD FOREIGN MATERIAL IN THE LIGHT GUIDE LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25670223 ↗
  • MalfunctionJun 29, 2026No Device Output

    THE CUSTOMER RETURNED THE GASTROINTESTINAL VIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING DEVICE ANALYSIS, THE SERVICE TECHNICIAN IDENTIFIED THAT THE DEVICE HAD NO LIGHT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25653635 ↗
  • MalfunctionJun 29, 2026Failure to Clean Adequately

    THE CUSTOMER RETURNED THE GASTROINTESTINAL VIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING DEVICE ANALYSIS, THE SERVICE TECHNICIAN IDENTIFIED THAT THE DEVICE HAD WHITE FOREIGN MATERIAL IN THE AIR/WATER CHANNEL AND CYLINDER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25653154 ↗
  • MalfunctionJun 25, 2026Erratic or Intermittent Display; No Display/Image

    IT WAS REPORTED THE GASTROINTESTINAL VIDEOSCOPE HAD NO IMAGE AND ONLY WAVY LINES WERE DISPLAYED. THE ISSUE OCCURRED DURING AN UNSPECIFIED THERAPEUTIC PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25622621 ↗
  • MalfunctionJun 25, 2026Failure to Clean Adequately

    THE CUSTOMER RETURNED THE GASTROINTESTINAL VIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING DEVICE ANALYSIS, THE SERVICE TECHNICIAN IDENTIFIED THAT THE DEVICE HAD FOREIGN MATERIAL ON THE OBJECTIVE LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25619940 ↗
  • MalfunctionJun 15, 2026No Display/Image

    THE CUSTOMER RETURNED THE GASTROINTESTINAL VIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING DEVICE ANALYSIS, THE SERVICE TECHNICIAN IDENTIFIED THAT THE DEVICE HAD A BLACK SCREEN WITH NO IMAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25505351 ↗
  • MalfunctionJun 11, 2026Thermal Decomposition of Device

    THE CUSTOMER RETURNED THE GASTROINTESTINAL VIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING DEVICE ANALYSIS, THE SERVICE TECHNICIAN IDENTIFIED THAT THE DEVICE HAD A BURN ON THE DISTAL END PLASTIC COVER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25480138 ↗
  • MalfunctionJun 11, 2026Failure to Clean Adequately

    THE CUSTOMER RETURNED THE GASTROINTESTINAL VIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE EVALUATION, THE REPAIR TECHNICIAN INDICATES THAT A DIRTY OBJECTIVE LENS WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25478330 ↗
  • MalfunctionJun 10, 2026No Display/Image

    THE CUSTOMER RETURNED THE GASTROINTESTINAL VIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING DEVICE ANALYSIS, THE SERVICE TECHNICIAN IDENTIFIED THAT THE DEVICE HAD NO IMAGE AND THE SCREEN WAS BLACK. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25460439 ↗
  • MalfunctionJun 10, 2026Failure to Clean Adequately

    THE CUSTOMER RETURNED THE GASTROINTESTINAL VIDEOSCOPE TO THE SERVICE CENTER WITH NO COMPLAINT. DURING DEVICE ANALYSIS, THE SERVICE TECHNICIAN IDENTIFIED THAT THE DEVICE HAD BLACK FOREIGN MATERIAL IN THE LIGHT GUIDE LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25457673 ↗
  • MalfunctionJun 9, 2026Thermal Decomposition of Device

    THE CUSTOMER DID NOT REPORT A MALFUNCTION. DURING ONSITE INSPECTION OF GASTROINTESTINAL VIDEOSCOPE, A BURNED C-COVER WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25449853 ↗
  • MalfunctionJun 9, 2026Failure to Clean Adequately

    THE CUSTOMER RETURNED THE GASTROINTESTINAL VIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING DEVICE ANALYSIS, THE SERVICE TECHNICIAN IDENTIFIED THAT THE DEVICE HAD FOREIGN MATERIAL IN THE LIGHT GUIDED LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25443918 ↗
  • MalfunctionJun 8, 2026Erratic or Intermittent Display

    THE CUSTOMER RETURNED THE GASTROINTESTINAL VIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING DEVICE ANALYSIS, THE SERVICE TECHNICIAN IDENTIFIED THAT THE DEVICE HAD A FLICKERING IMAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25429988 ↗
  • MalfunctionJun 2, 2026Communication or Transmission Problem

    THE CUSTOMER REPORTED SCOPE GIVING ERROR DURING AN UNSPECIFIED PROCEDURE. INVOLVING AN OLYMPUS GASTROINTESTINAL VIDEOSCOPE. THERE WAS NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25364603 ↗
  • MalfunctionMay 29, 2026Poor Quality Image

    THE CUSTOMER RETURNED THE GASTROINTESTINAL VIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE ANALYSIS, IT WAS OBSERVED THAT THE IMAGE WAS DISTORTED. IT WAS DETERMINED THAT THE ELECTRICAL CONNECTOR WAS DEFECTIVE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25334041 ↗
  • MalfunctionMay 29, 2026Failure to Clean Adequately

    THE CUSTOMER RETURNED THE GASTROINTESTINAL VIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE ANALYSIS, IT WAS OBSERVED THAT THE AIR/WATER CHANNEL HAD WHITE FOREIGN MATERIAL. THE ROOT CAUSE OF THE WHITE FOREIGN MATERIAL COULD NOT BE IDENTIFIED. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25334033 ↗
  • MalfunctionMay 27, 2026Failure to Clean Adequately

    THE GASTROINTESTINAL VIDEOSCOPE WAS RETURNED TO OLYMPUS WITH A REPORTED COMPLAINT OF LEAK AT THE DISTAL TIP; THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, AN MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER, THE AIR/WATER CHANNEL HAD WHITE FOREIGN MATERIAL. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25299206 ↗
  • MalfunctionMay 27, 2026Failure to Clean Adequately

    THE GASTROINTESTINAL VIDEOSCOPE WAS RETURNED TO OLYMPUS WITH A REPORTED COMPLAINT OF LOOSE ANGULATION FOR BOTH CONTROL KNOBS; THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, AN MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER, THE LIGHT GUIDE LENS HAS DIRTY, FOREIGN BODIES. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25298195 ↗
  • MalfunctionMay 25, 2026Erratic or Intermittent Display; No Display/Image

    THE CUSTOMER REPORTED INTERMITTENT/ BLACK IMAGE INVOLVING AN OLYMPUS GASTROINTESTINAL VIDEOSCOPE. THERE WAS NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25273176 ↗
  • MalfunctionMay 25, 2026No Display/Image

    THE CUSTOMER REPORTED NO IMAGE DURING ENDOSCOPY PROCEDURE BEFORE SEDATION INVOLVING AN OLYMPUS GASTROINTESTINAL VIDEOSCOPE. THERE WAS NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25269293 ↗
  • MalfunctionMay 22, 2026Melted

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED A BURN ON THE DISTAL END COVER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25257809 ↗

Other Colonoscope and accessories, flexible/rigid devices

All Colonoscope and accessories, flexible/rigid devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page