Skip to content

EVIS EXERA II COLONOVIDEOSCOPE OLYMPUS PCF TYPE Q180AL (UDI 04953170339189)Recalls & reported problems

Olympus Medical Systems Corp. · "PCF-Q180AL EVIS EXERA II COLONOSCOPE" · Class 2

All models of EVIS EXERA II COLONOVIDEOSCOPE (28)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Olympus EVIS EXERA II colonoscope has 118 adverse event reports, with 8 injuries and no deaths. Many describe image problems such as no image, black or grainy pictures, and some describe cleaning problems or residue after reprocessing. Reports rose from 22 in the prior 12 months to 78 in the last 12 months. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
119
Deaths reported
0
Injuries reported
8
Recalls naming it
0

Summary

The FDA's MAUDE database contains 119 adverse event reports that list this exact device (DI 04953170339189), received between May 5, 2021 and Sep 24, 2026 — classified as 8 injuries, 111 malfunctions.

The most frequently reported specific device problem is “Failure to Clean Adequately” (31 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 3 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 3 reportsFeb 2025: 3 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 4 reportsJun 2025: 1 reportsJul 2025: 2 reportsAug 2025: 3 reportsSep 2025: 1 reportsOct 2025: 5 reportsNov 2025: 8 reportsDec 2025: 31 reportsJan 2026: 10 reportsFeb 2026: 7 reportsMar 2026: 8 reportsApr 2026: 7 reportsMay 2026: 3 reportsJun 2026: 4 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports20242025202631
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Clean Adequately31
  • No Display/Image26
  • Poor Quality Image20
  • Erratic or Intermittent Display12
  • Device Reprocessing Problem10
  • Residue After Decontamination6
  • Adverse Event Without Identified Device or Use Problem4
  • Contamination3

Patient problems

  • No Clinical Signs, Symptoms or Conditions115
  • Bowel Perforation2
  • Hemorrhage/Blood Loss/Bleeding2

Recent adverse event reports

The 25 most recent of 119 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 24, 2026Failure to Clean Adequately

    The customer returned the colonovideoscope for a separate issue. During the initial evaluation of the returned device, white residue in air water channel and cylinder. No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50888874 ↗
  • MalfunctionJun 27, 2026No Display/Image

    THE CUSTOMER RETURNED THE COLONOVIDEOSCOPE, WHICH IS AN OLYMPUS ASSET WITH NO REPORTED COMPLAINT. DURING THE EVALUATION OF THE DEVICE, NO IMAGE AND BLACK IMAGE WAS OBSERVED. THE SERVICE REPAIR TECHNICIAN INDICATED THAT THE MOST LIKELY CAUSE OF THE NO IMAGE AND BLACK IMAGE WAS DUE TO COMPONENT FAILURE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25645988 ↗
  • MalfunctionJun 12, 2026Failure to Clean Adequately

    THE CUSTOMER RETURNED THE COLONOVIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING DEVICE ANALYSIS, THE SERVICE TECHNICIAN IDENTIFIED THAT THE DEVICE HAD BLACK FOREIGN MATERIAL IN THE INSTRUMENT CHANNEL. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25487856 ↗
  • MalfunctionJun 9, 2026No Display/Image; Residue After Decontamination

    THE CUSTOMER RETURNED THE COLONOVIDEOSCOPE, WHICH IS AN OLYMPUS ASSET FOR A SEPARATE ISSUE. DURING THE EVALUATION OF THE DEVICE, NO IMAGE AND WHITE RESIDUE IN AIR/ WATER CHANNEL WAS OBSERVED. THE SERVICE REPAIR TECHNICIAN INDICATED THAT THE MOST LIKELY CAUSE OF THE WHITE RESIDUE WAS NOT ESTABLISHED AND FOR THE NO IMAGE FINDING WAS DUE TO COMPONENT FAILURE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25446808 ↗
  • MalfunctionJun 1, 2026Poor Quality Image

    THE CUSTOMER RETURNED THE COLONOVIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE SERVICE REPAIR TECHNICIAN INDICATES THAT GRAINY AND DUSTY IMAGE WAS FOUND.  THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25351917 ↗
  • MalfunctionMay 29, 2026Degraded

    THE CUSTOMER RETURNED THE COLONOVIDEOSCOPE TO THE SERVICE CENTER FOR EVALUATION RELATED TO WATER INVASION. DURING TESTING, THE REPAIR CENTER TECHNICIAN FOUND CORROSION OBSERVED IN ONE LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25333179 ↗
  • MalfunctionMay 24, 2026No Display/Image

    THE COLONOVIDEOSCOPE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT OF WATER INTRUSION. THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER. DURING INSPECTION OF THE DEVICE, NO IMAGE WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25267267 ↗
  • MalfunctionMay 20, 2026Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE AIR WATER CHANNEL HAD WHITE RESIDUE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25228791 ↗
  • MalfunctionApr 27, 2026Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED DEBRIS INSIDE LIGHT GUIDE LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25000708 ↗
  • MalfunctionApr 15, 2026No Display/Image

    IT WAS REPORTED THAT THE COLONOVIDEOSCOPE HAD A BLACK SCREEN. THE EVENT OCCURRED DURING INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24888099 ↗
  • MalfunctionApr 13, 2026Poor Quality Image

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE HAD FLICKERING IMAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24856388 ↗
  • MalfunctionApr 10, 2026No Display/Image

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED NO IMAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24840684 ↗
  • MalfunctionApr 8, 2026Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED DEBRIS ON OBJECTIVE AND LIGHT GUIDE LENSES. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24815218 ↗
  • MalfunctionApr 7, 2026Poor Quality Image

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE IMAGE WAS DISTORTION. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24799458 ↗
  • MalfunctionApr 6, 2026Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED WHITE RESIDUES ON AUXILIARY CHANNEL AND AIR/WATER CHANNEL. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24788324 ↗
  • MalfunctionMar 30, 2026Residue After Decontamination

    IT WAS OBSERVED DURING THE DEVICE EVALUATION THAT THE COLONOVIDEOSCOPE HAD WHITE RESIDUE INSIDE THE AIR/WATER CYLINDER, AIR/WATER CHANNEL, AND NOZZLE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24718991 ↗
  • MalfunctionMar 17, 2026Melted

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE HAD BURNT C-COVER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24618810 ↗
  • MalfunctionMar 16, 2026Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED WHITE RESIDUE IN THE AIR WATER CHANNEL. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24604789 ↗
  • MalfunctionMar 16, 2026Erratic or Intermittent Display

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED AN INTERMITTENT FLICKERING IMAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24603324 ↗
  • MalfunctionMar 13, 2026Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE HAD WHITE RESIDUE IN THE AUXILIARY CHANNEL AND IN THE AIR/WATER CHANNEL. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24595749 ↗
  • MalfunctionMar 13, 2026Erratic or Intermittent Display; Poor Quality Image

    IT WAS REPORTED THAT THE COLONOVIDEOSCOPE EXHIBITED IMAGE WAS WARPED, STATIC AND SNOWY. THE ISSUE WAS FOUND DURING INSPECTION FOR USE OF THE DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24592219 ↗
  • MalfunctionMar 5, 2026No Display/Image

    IT WAS REPORTED THAT THE COLONOVIDEOSCOPE HAD A NO IMAGE. THE ISSUE OCCURRED DURING THE INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24517505 ↗
  • MalfunctionMar 3, 2026Poor Quality Image

    IT WAS REPORTED THAT THE COLONOVIDEOSCOPE EXHIBITED A DISTORTED IMAGE DURING INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24494796 ↗
  • MalfunctionFeb 23, 2026Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE HAD DEBRIS IN NOZZLE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24412895 ↗
  • MalfunctionFeb 19, 2026Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED WHITE RESIDUE INSIDE THE CONTROL BODY. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24389522 ↗

Other Colonoscope and accessories, flexible/rigid devices

All Colonoscope and accessories, flexible/rigid devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page