EVIS EXERA II COLONOVIDEOSCOPE OLYMPUS PCF TYPE H180AI (UDI 04953170339882)Recalls & reported problems
Olympus Medical Systems Corp. · "PCF-H180AI EVIS EXERA II ULTRA THIN CLNV" · Class 2
All models of EVIS EXERA II COLONOVIDEOSCOPE (28)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 04953170339882), received between Aug 8, 2023 and Dec 12, 2025 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Failure to Clean Adequately” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Clean Adequately5
- Device Reprocessing Problem1
- Mechanical Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 12, 2025Failure to Clean Adequately; Failure to Clean Adequately
IT WAS OBSERVED DURING THE DEVICE EVALUATION, THE AIR/WATER CHANNEL OF THE COLONOVIDEOSCOPE HAD WHITE RESIDUE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23790237 ↗MalfunctionJan 29, 2024Failure to Clean Adequately; Failure to Clean Adequately
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED FOREIGN OBJECTS INSIDE THE AIR/WATER TUBE AND AIR/WATER CYLINDER. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18595550 ↗MalfunctionJan 14, 2024Failure to Clean Adequately; Failure to Clean Adequately
THE CUSTOMER REPORTED TO OLYMPUS, THE COLONOVIDEOSCOPE HAD AN ANGULATION MALFUNCTION. THE DEVICE WAS RETURNED FOR EVALUATION. DURING THE DEVICE EVALUATION, THERE WAS A FOREIGN MATERIAL IN THE AIR/WATER TUBE, THE AIR/WATER CYLINDER AND IN THE AUXILIARY JET CHANNEL. THERE WERE NO REPORTS OF PATIENT HARM. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18509615 ↗MalfunctionJan 2, 2024Failure to Clean Adequately; Failure to Clean Adequately
A USER FACILITY RETURNED THE OLYMPUS ASSET, EVIS EXERA II COLONOVIDEOSCOPE TO THE OLYMPUS SERVICE CENTER. UPON INSPECTION AND TESTING OF THE RETURNED DEVICE, IT WAS OBSERVED THAT THERE THE A/W (AIR/WATER) TUBE AND CYLINDER HAD FOREIGN MATERIALS AND THE J-TUBE (JET TUBE) WAS CLOGGED WITH FOREIGN MATERIAL. THIS REPORT IS BEING SUBMITTED FOR THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18432053 ↗MalfunctionOct 30, 2023Failure to Clean Adequately; Failure to Clean Adequately
THE CUSTOMER RETURNED THE EVIS EXERA II COLONOVIDEOSCOPE, FOR THE ANNUAL INSPECTION WITHOUT REPORTING ANY ISSUES. THERE WERE NO REPORTS OF PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. INSPECTION AND TESTING OF THE RETURNED DEVICE FOUND THAT THE AIR/WATER TUBE AND AIR/WATER CYLINDER HAD FOREIGN MATERIAL. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTIONS FOUND DURING THE EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18031932 ↗MalfunctionOct 20, 2023Device Reprocessing Problem; Device Reprocessing Problem
THE CUSTOMER REPORTED TO OLYMPUS, THE EVIS EXERA II COLONOVIDEOSCOPE HAD A BROKEN BODY CONTROL UNIT RUBBER. THERE WERE NO REPORTS OF PATIENT HARM. THE DEVICE WAS RETURNED FOR EVALUATION. DURING THE DEVICE EVALUATION, THE FOLLOWING REPORTABLE MALFUNCTION WAS FOUND: AIR OR WATER CYLINDER HAD FOREIGN OBJECTS. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17980866 ↗MalfunctionAug 8, 2023Mechanical Problem; Mechanical Problem
THE CUSTOMER REPORTED TO OLYMPUS THAT THE EVIS EXERA II COLONOVIDEOSCOPE WAS DUE FOR AN ANNUAL INSPECTION. NO PATIENT HARM WAS ASSOCIATED WITH THIS EVENT/REQUEST. THE DEVICE WAS RETURNED TO OLYMPUS FOR AN ANNUAL INSPECTION AND FOUND DUE TO DAMAGE THE ANGULATION BECOMES LOCKED AND CANNOT DISENGAGE, DUE TO DEFORMATION OF RIGHT/LEFT SPROCKET THE BENDING SECTION COULD NOT BE CONTROLLED AT ALL, AND THE RIGHT/LEFT HOLDS THE POSITION WITH NO AUTOMATIC RETURN. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17485153 ↗
Other Gastroscope and accessories, flexible/rigid devices
- PENTAX Medical EG29-i10cHoya Corporation · Gastroscope And Accessories, Flexible/Rigid1,623 reports
- EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-1TH190Olympus Medical Systems Corp. · Gastroscope And Accessories, Flexible/Rigid4 recalls6 deaths1,025 reports
- EVIS EXERA lll GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-XP190NOlympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid1 recall559 reports
- EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE GIF-Q165Olympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid349 reports
- EVIS EXERA II GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF TYPE H180Olympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid1 death302 reports
- PENTAX EG27-i10Hoya Corporation · Gastroscope and accessories, flexible/rigid261 reports
- EVIS EXERA II COLONOVIDEOSCOPE OLYMPUS CF TYPE H180ALOlympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid2 deaths252 reports
- BALLOON MAJ-1351Olympus Medical Systems Corp. · Gastroscope and accessories, flexible/rigid174 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page