EVIS EXERA II COLONOVIDEOSCOPE OLYMPUS PCF TYPE H180AI (UDI 04953170339882)Recalls & reported problems

Olympus Medical Systems Corp. · "PCF-H180AI EVIS EXERA II ULTRA THIN CLNV" · Class 2

All models of EVIS EXERA II COLONOVIDEOSCOPE (28)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 04953170339882), received between Aug 8, 2023 and Dec 12, 2025 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Failure to Clean Adequately” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 3 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Clean Adequately5
  • Device Reprocessing Problem1
  • Mechanical Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 12, 2025Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE EVALUATION, THE AIR/WATER CHANNEL OF THE COLONOVIDEOSCOPE HAD WHITE RESIDUE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23790237 ↗
  • MalfunctionJan 29, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED FOREIGN OBJECTS INSIDE THE AIR/WATER TUBE AND AIR/WATER CYLINDER. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18595550 ↗
  • MalfunctionJan 14, 2024Failure to Clean Adequately; Failure to Clean Adequately

    THE CUSTOMER REPORTED TO OLYMPUS, THE COLONOVIDEOSCOPE HAD AN ANGULATION MALFUNCTION. THE DEVICE WAS RETURNED FOR EVALUATION. DURING THE DEVICE EVALUATION, THERE WAS A FOREIGN MATERIAL IN THE AIR/WATER TUBE, THE AIR/WATER CYLINDER AND IN THE AUXILIARY JET CHANNEL. THERE WERE NO REPORTS OF PATIENT HARM. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18509615 ↗
  • MalfunctionJan 2, 2024Failure to Clean Adequately; Failure to Clean Adequately

    A USER FACILITY RETURNED THE OLYMPUS ASSET, EVIS EXERA II COLONOVIDEOSCOPE TO THE OLYMPUS SERVICE CENTER. UPON INSPECTION AND TESTING OF THE RETURNED DEVICE, IT WAS OBSERVED THAT THERE THE A/W (AIR/WATER) TUBE AND CYLINDER HAD FOREIGN MATERIALS AND THE J-TUBE (JET TUBE) WAS CLOGGED WITH FOREIGN MATERIAL. THIS REPORT IS BEING SUBMITTED FOR THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18432053 ↗
  • MalfunctionOct 30, 2023Failure to Clean Adequately; Failure to Clean Adequately

    THE CUSTOMER RETURNED THE EVIS EXERA II COLONOVIDEOSCOPE, FOR THE ANNUAL INSPECTION WITHOUT REPORTING ANY ISSUES. THERE WERE NO REPORTS OF PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. INSPECTION AND TESTING OF THE RETURNED DEVICE FOUND THAT THE AIR/WATER TUBE AND AIR/WATER CYLINDER HAD FOREIGN MATERIAL. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTIONS FOUND DURING THE EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18031932 ↗
  • MalfunctionOct 20, 2023Device Reprocessing Problem; Device Reprocessing Problem

    THE CUSTOMER REPORTED TO OLYMPUS, THE EVIS EXERA II COLONOVIDEOSCOPE HAD A BROKEN BODY CONTROL UNIT RUBBER. THERE WERE NO REPORTS OF PATIENT HARM. THE DEVICE WAS RETURNED FOR EVALUATION. DURING THE DEVICE EVALUATION, THE FOLLOWING REPORTABLE MALFUNCTION WAS FOUND: AIR OR WATER CYLINDER HAD FOREIGN OBJECTS. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17980866 ↗
  • MalfunctionAug 8, 2023Mechanical Problem; Mechanical Problem

    THE CUSTOMER REPORTED TO OLYMPUS THAT THE EVIS EXERA II COLONOVIDEOSCOPE WAS DUE FOR AN ANNUAL INSPECTION. NO PATIENT HARM WAS ASSOCIATED WITH THIS EVENT/REQUEST. THE DEVICE WAS RETURNED TO OLYMPUS FOR AN ANNUAL INSPECTION AND FOUND DUE TO DAMAGE THE ANGULATION BECOMES LOCKED AND CANNOT DISENGAGE, DUE TO DEFORMATION OF RIGHT/LEFT SPROCKET THE BENDING SECTION COULD NOT BE CONTROLLED AT ALL, AND THE RIGHT/LEFT HOLDS THE POSITION WITH NO AUTOMATIC RETURN. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17485153 ↗

Other Gastroscope and accessories, flexible/rigid devices

All Gastroscope and accessories, flexible/rigid devices →