TRACHEAL INTUBATION FIBERSCOPE OLYMPUS LF-GP (UDI 04953170340215)Recalls & reported problems

Olympus Medical Systems Corp. · LF-GP TRACHEAL INTUBATION FIBE RSCOPE-PO · Class 2

All models of TRACHEAL INTUBATION FIBERSCOPE (7)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Olympus LF-GP tracheal intubation fiberscope is a thin flexible scope used to guide breathing tubes into the airway. Most of the 70 reports describe the scope's outer coating peeling or flaking, parts coming loose, and cleaning or reprocessing problems, with a few mentioning a foreign body left in the patient. There are 3 injuries and no deaths. Two open recalls, started September 2023 and July 2023, concern suction flow not meeting specification and the scope fitting too loosely in the tube connector.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
70
Deaths reported
0
Injuries reported
3
Recalls naming it
2

Summary

The FDA's MAUDE database contains 70 adverse event reports that list this exact device (DI 04953170340215), received between Sep 24, 2019 and Aug 31, 2026 — classified as 3 injuries, 67 malfunctions.

The most frequently reported specific device problem is “Peeled/Delaminated” (26 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 3 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 2 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 1 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 2 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 2 reportsJan 2026: 1 reportsFeb 2026: 2 reportsMar 2026: 4 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 3 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 28, 2023Class IIOpen, Classified

    Fiberscopes do not have adequate data to support that the suction flow rate meets the device specification.

    Z-0557-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Jul 27, 2023Class IIOpen, Classified

    Complaint of endoscope model becoming lodged in the endotracheal tube connector due to diameter being too large.

    Z-2504-2023 · Root cause: Labeling Change Control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Peeled/Delaminated26
  • Detachment of Device or Device Component9
  • Failure to Clean Adequately7
  • Device Reprocessing Problem6
  • Microbial Contamination of Device6
  • Degraded4
  • Break3
  • Material Split, Cut or Torn3

Patient problems

  • No Clinical Signs, Symptoms or Conditions65
  • Foreign Body In Patient2
  • No Known Impact Or Consequence To Patient2
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 25 most recent of 70 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 31, 2026Device Reprocessing Problem; Corroded

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50523487 ↗
  • MalfunctionAug 25, 2026Degraded; Peeled/Delaminated

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50436665 ↗
  • MalfunctionAug 17, 2026Material Fragmentation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50279549 ↗
  • MalfunctionJul 30, 2026Fracture

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50087577 ↗
  • MalfunctionApr 15, 2026Detachment of Device or Device Component

    THE CUSTOMER DID NOT REPORT A MALFUNCTION. DURING INSPECTION OF THE FIBERSCOPE, GLUE WAS MISSING FROM THE OBJECTIVE AND LIGHT GUIDE LENS WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24885793 ↗
  • MalfunctionApr 6, 2026Detachment of Device or Device Component

    THE TRACHEAL INTUBATION FIBERSCOPE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT OF HOLE ON THE DISTAL END. THIS DOES NOT INDICATE AN MDR REPORTABLE EVENT. HOWEVER, MDR REPORTABLE FAILURE WAS FOUND DURING TESTING AT THE SERVICE CENTER. DURING INSPECTION OF THE DEVICE, A-RUBBER ADHESIVE WAS DETACHED AT THE DISTAL END WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24789651 ↗
  • MalfunctionMar 25, 2026Material Fragmentation

    IT WAS OBSERVED DURING OLYMPUS' DEVICE INSPECTION THAT THE FIBERSCOPE EXHIBITED A CHIPPED ADHESIVE ON THE OBJECTIVE LENS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24688026 ↗
  • MalfunctionMar 17, 2026Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE TRACHEAL INTUBATION FIBERSCOPE EXHIBITED A CHIPPED LIGHT GUIDE LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24615107 ↗
  • MalfunctionMar 3, 2026Device Reprocessing Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE TRACHEAL INTUBATION FIBERSCOPE EXHIBITED DIRT ON THE IMAGE CONDUCTOR BUNDLE COVER LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24495902 ↗
  • MalfunctionMar 2, 2026Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE TRACHEAL INTUBATION FIBERSCOPE EXHIBITED DISTAL END DETACHED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24484298 ↗
  • MalfunctionFeb 11, 2026Complete Blockage; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE TRACHEAL INTUBATION FIBERSCOPE FORCEPS PASSAGE CHANNEL CLOGGED WITH FOREIGN MATERIAL. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24311023 ↗
  • MalfunctionFeb 9, 2026Degraded

    IT WAS OBSERVED DURING THE DEVICE INSPECTION THAT THE TRACHEAL INTUBATION FIBERSCOPE HAD THE FIBERSCOPE EYEPIECE LENS COATING DETACHED. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24293923 ↗
  • MalfunctionJan 8, 2026Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE CEMENT WAS MISSING FROM THE OBJECTIVE AND LIGHT GUIDE LENS OF THE FIBERSCOPE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24005192 ↗
  • MalfunctionDec 27, 2025Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE FIBERSCOPE EXHIBITED DUST INSIDE LENS CAUSING A BLURRY IMAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23902170 ↗
  • MalfunctionDec 15, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE FLEXIBLE SCOPE ADHESIVE A-RUBBER HAD DETACHED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23802932 ↗
  • MalfunctionAug 28, 2025Device Reprocessing Problem; Device Reprocessing Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE FIBERSCOPE EXHIBITED RESIDUAL LIQUID OR FOREIGN MATERIAL CAME OUT OF CHANNEL-TUBE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22911608 ↗
  • MalfunctionAug 26, 2025Material Separation; Material Separation

    DISTAL END DAMAGED/LENS MISSING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22894291 ↗
  • MalfunctionJul 31, 2025Material Puncture/Hole; Material Puncture/Hole

    IT WAS OBSERVED DURING THE DEVICE EVALUATION, THE FIBERSCOPE EXHIBITED A TEAR INSIDE THE CHANNEL. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22673013 ↗
  • MalfunctionJul 21, 2025Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THE SUBJECT DEVICE WAS DAMAGED WITH PHYSICAL FORCE WHILE TRYING TO INTUBATE A PATIENT WITH A SWOLLEN THROAT. THIS OCCURRED DURING AN UNSPECIFIED PROCEDURE. THERE WERE NO REPORTS OF PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22553779 ↗
  • MalfunctionJun 18, 2025Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS REPORTED THE FIBERSCOPE HAD A FOREIGN OBJECT STUCK AT THE DISTAL TIP DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22288553 ↗
  • MalfunctionMay 12, 2025Leak/Splash; Detachment of Device or Device Component; Material Split, Cut or Torn; Leak/Splash; Detachment of Device or Device Component; Material Split, Cut or Torn

    IT WAS REPORTED THE TRACHEAL INTUBATION FIBERSCOPE WAS UNSHEATHED ON THE DISTAL PART. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT. THE ISSUE WAS FOUND BEFORE PERFORMING THE EXAMINATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22002685 ↗
  • MalfunctionMar 28, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS OBSERVED DURING THE DEVICE INSPECTION THAT THE FIBERSCOPE EXHIBITED THE BENDING SECTION COVER HAD COME OFF OF THE BENDING SECTION. THERE WAS NO PATIENT INVOLVEMENT. THE REPORT CAPTURES THE REPORTABLE MALFUNCTION IDENTIFIED DURING THE OLYMPUS DEVICE EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21725044 ↗
  • MalfunctionMar 17, 2025Degraded; Degraded

    NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21619983 ↗
  • MalfunctionSep 10, 2024Peeled/Delaminated; Peeled/Delaminated

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE INTUBATION FIBERSCOPE EXHIBITED THE COATING PEELING 1 TO 2MM OR MORE OFF THE IMAGE GUIDE HOSE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20184619 ↗
  • MalfunctionJul 21, 2024No Display/Image; No Display/Image

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE FLEXIBLE SCOPE EXHIBITED THE IMAGE GUIDE SNAKE TUBE DISPLAY IS MISSING (MORE THAN 1MM2). THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19787832 ↗