Skip to content

VISERA TRANSNASAL ESOPHAGOVIDEOSCOPE OLYMPUS PEF TYPE VRecalls & reported problems

Olympus Medical Systems Corp. · "PEF-V TRANSNASAL ESOPHAGO VIDE O SCOPE" · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04953170340376), received between Jan 26, 2021 and Apr 5, 2023 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Device Reprocessing Problem” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Sep 25, 2023Class IIOpen, Classified

    Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

    Z-0344-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device Reprocessing Problem1
  • Material Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Patient Involvement1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 5, 2023Device Reprocessing Problem; Device Reprocessing Problem

    THE CUSTOMER CALLED THE OLYMPUS TECHNICAL ASSISTANCE CENTER (TAC) TO INQUIRE INFORMATION ON HOW THE VISERA TRANSNASAL ESOPHAGOVIDEOSCOPE SHOULD BE REPROCESSED, AND WHICH CHEMICAL AGENTS ARE COMPATIBLE WITH THE PEF-V. IT WAS REPORTED THAT DURING THE CALL, THE CUSTOMER STATED THAT THE FACILITY HAD BEEN STORING THE REPROCESSED SCOPE IN A CONTAINER INSTEAD OF HANGING THEM IN A DRYING CABINET. THE OLYMPUS TECHNICAL ASSISTANCE CENTER (TAC) REPRESENTATIVE THEN ADVISED THE CUSTOMER TO FOLLOW THE IFU STEP ON REPROCESSING AND STORAGE WHICH RECOMMEND THE SCOPE SHOULD BE HUNG VERTICALLY DURING STORAGE TO AVOID WATER POOLING IN THE CHANNELS AND BACTERIA GROWTH. THIS MDR IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION OF INCORRECT REPROCESSING OF THE SCOPE STATED BY THE CUSTOMER DURING THE PHONE CALL TO THE OLYMPUS TECHNICAL ASSISTANCE CENTER. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16687010 ↗
  • MalfunctionJan 26, 2021Material Separation; Material Separation

    AS REPORTED, THE DISTAL END COVER OF THE VISERA TRANSNASAL ESOPHAGOVIDEOSCOPE WAS DAMAGED. ACCORDING TO THE INITIAL REPORTER, THE RUBBER AT THE DISTAL TIP WAS BEEN PULLED OFF. THIS EVENT OCCURRED DURING REPROCESSING AND THEREFORE HAD NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 11230519 ↗

Other Gastroscope and accessories, flexible/rigid devices

All Gastroscope and accessories, flexible/rigid devices →