ENDOEYE FLEX 3D DEFLECTABLE VIDEOSCOPE OLYMPUS LTF-190-10-3D (UDI 04953170340710)Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of ENDOEYE FLEX 3D DEFLECTABLE VIDEOSCOPE (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Olympus ENDOEYE FLEX 3D videoscope is a bendable scope used for viewing inside the body during surgery. Many of the 292 reports describe the image failing or looking blurry, adhesive peeling off the lens, and damage at the tip. The number of reports rose from 19 in the prior 12 months to 36 in the last 12 months. The reports include no deaths and 17 injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
293
Deaths reported
0
Injuries reported
17
Recalls naming it
0

Summary

The FDA's MAUDE database contains 293 adverse event reports that list this exact device (DI 04953170340710), received between Apr 8, 2020 and Sep 9, 2026 — classified as 17 injuries, 276 malfunctions.

The most frequently reported specific device problem is “No Display/Image” (68 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 5 reportsDec 2023: 4 reportsJan 2024: 6 reportsFeb 2024: 2 reportsMar 2024: 5 reportsApr 2024: 3 reportsMay 2024: 1 reportsJun 2024: 3 reportsJul 2024: 2 reportsAug 2024: 1 reportsSep 2024: 2 reportsOct 2024: 3 reportsNov 2024: 2 reportsDec 2024: 2 reportsJan 2025: 1 reportsFeb 2025: 2 reportsMar 2025: 1 reportsApr 2025: 2 reportsMay 2025: 2 reportsJun 2025: 2 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 4 reportsNov 2025: 2 reportsDec 2025: 2 reportsJan 2026: 16 reportsFeb 2026: 4 reportsMar 2026: 1 reportsApr 2026: 5 reportsMay 2026: 0 reportsJun 2026: 4 reportsJul 2026: 0 reportsAug 2026: 2 reportsSep 2026: 1 reportsOct 2026: 0 reports20242025202616
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • No Display/Image68
  • Peeled/Delaminated65
  • Communication or Transmission Problem32
  • Poor Quality Image29
  • Degraded23
  • Adverse Event Without Identified Device or Use Problem16
  • Break14
  • Detachment of Device or Device Component13

Patient problems

  • No Clinical Signs, Symptoms or Conditions275
  • Hemorrhage/Blood Loss/Bleeding5
  • Fistula4
  • Wound Dehiscence3
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
  • Obstruction/Occlusion2
  • Post Operative Wound Infection2
  • Thrombosis/Thrombus2

Recent adverse event reports

The 25 most recent of 293 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 9, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50694859 ↗
  • MalfunctionAug 26, 2026Degraded

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50457014 ↗
  • MalfunctionAug 5, 2026Poor Quality Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50147516 ↗
  • MalfunctionJun 25, 2026Detachment of Device or Device Component

    THE CUSTOMER RETURNED THE FLEX DEFLECTABLE VIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE SERVICE REPAIR TECHNICIAN INDICATES THAT THE ADHESIVE ON THE OBJECTIVE LENS IS CHIPPED AND PARTIALLY MISSING, AND THE BENDING SECTION COVER GLUE IS CHIPPED WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25629762 ↗
  • MalfunctionJun 25, 2026Peeled/Delaminated

    THE CUSTOMER RETURNED THE FLEX DEFLECTABLE VIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE SERVICE REPAIR TECHNICIAN INDICATES THAT THE OBJECTIVE LENS CEMENT WAS PEELING. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25620501 ↗
  • MalfunctionJun 9, 2026Break; Poor Quality Image

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25448516 ↗
  • MalfunctionJun 2, 2026Degraded

    IT WAS REPORTED THAT DURING IN-ROOM SET UP / INSPECTION FOR AN UNSPECIFIED PROCEDURE, PRIOR TO THE PATIENT BEING IN THE ROOM, THE FLEX DEFLECTABLE VIDEOSCOPE WAS FOUND TO EXHIBIT DAMAGE AT THE DEVICE TIP. THERE WAS NO PATIENT INVOLVEMENT, AND NO HARM WAS REPORTED AS A RESULT OF THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25359240 ↗
  • MalfunctionApr 23, 2026Break; Material Fragmentation

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE FLEX 3D DEFLECTABLE VIDEOSCOPE HAD ADHESIVE DETACHED FROM THE OBJECTIVE LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24962089 ↗
  • MalfunctionApr 21, 2026Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE FLEX DEFLECTABLE VIDEOSCOPE HAD CHIPPED A-RUBBER AT THE DISTAL END. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24942606 ↗
  • MalfunctionApr 15, 2026Degraded; Material Fragmentation

    IT WAS OBSERVED DURING THE DEVICE EVALUATION, THE FLEX DEFLECTABLE 3D VIDEOSCOPE EXHIBITED THE OBJECTIVE LENS GLUE WAS CHIPPED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24887999 ↗
  • MalfunctionApr 14, 2026Material Integrity Problem; Sharp Edges

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24868068 ↗
  • MalfunctionApr 6, 2026Material Split, Cut or Torn

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE ADHESIVE ON THE VIDEOSCOPE'S DISTAL END HAD A HOLE AND ITS BENDING SECTION COVER ADHESIVE WAS CHIPPED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24792202 ↗
  • MalfunctionMar 19, 2026Dent in Material; Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE FLEX 3D DEFLECTABLE VIDEOSCOPE EXHIBITED A DENT AT THE DISTAL BODY AND PEELING CEMENT IN THE LIGHT GUIDE AND OBJECTIVE LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24643185 ↗
  • MalfunctionFeb 5, 2026Break

    IT WAS REPORTED THE DURING REPROCESSING, THE FLEX 3D DEFLECTABLE VIDEOSCOPE EXHIBITED A BROKEN DISTAL TIP. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24268550 ↗
  • MalfunctionFeb 5, 2026Material Separation

    IT WAS OBSERVED DURING THE DEVICE INVESTIGATION THAT THE FLEX DEFLECTABLE VIDEOSCOPE HAD A MISSING ADHESIVE ON THE DISTAL END OF THE BENDING SECTION. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24267497 ↗
  • MalfunctionFeb 3, 2026Material Puncture/Hole

    NO ADDITIONAL INFORMATION FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24251197 ↗
  • MalfunctionFeb 2, 2026Material Split, Cut or Torn

    NO ADDITIONAL INFORMATION REC FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24235474 ↗
  • MalfunctionJan 30, 2026Peeled/Delaminated; Detachment of Device or Device Component

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE FLEX 3D DEFLECTABLE VIDEOSCOPE EXHIBITED PEELING CEMENT AROUND THE ADHESIVE AT THE OBJECTIVE LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24220312 ↗
  • MalfunctionJan 28, 2026Material Split, Cut or Torn

    NO ADDITIONAL INFORMATION REPORTED BY CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24206555 ↗
  • MalfunctionJan 22, 2026Poor Quality Image

    IT WAS REPORTED THE FLEX 3D DEFLECTABLE VIDEOSCOPE PRESENTED A VERY DIM LIGHT. THE ISSUE WAS IDENTIFIED DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24147986 ↗
  • MalfunctionJan 22, 2026Material Separation

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE FLEX 3D DEFLECTABLE VIDEOSCOPE LIGHT GUIDE PRONG AND MOUNT WAS MISSING THE CHANNEL RING. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24144972 ↗
  • MalfunctionJan 21, 2026Detachment of Device or Device Component; Material Split, Cut or Torn

    IT WAS REPORTED THE VIDEOSCOPE EXHIBITED TORN DISTAL TIP. THE EVENT OCCURRED DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24141228 ↗
  • MalfunctionJan 21, 2026Poor Quality Image

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE FLEX 3D DEFLECTABLE VIDEOSCOPE HAD A DIRTY OBJECT LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24132223 ↗
  • MalfunctionJan 9, 2026Poor Quality Image; Material Integrity Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE FLEX 3D DEFLECTABLE VIDEOSCOPE HAD A LINE IN THE IMAGE AND CRACKED A-RUBBER AT THE DISTAL END. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24027819 ↗
  • MalfunctionJan 9, 2026Degraded; Peeled/Delaminated

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE FLEX 3D DEFLECTABLE VIDEOSCOPE HAD AN UNEVEN SEAL AT THE SEAM ON THE DISTAL END AND PEELING CEMENT AROUND THE LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24027394 ↗

Other Laparoscope, general & plastic surgery devices