Skip to content

URETERO-RENO FIBERSCOPE OLYMPUS URF-P6RRecalls & reported problems

Olympus Medical Systems Corp. · "URF-P6R SUPER-SLIM FLEXIBLE FIBEROPTIC U" · Class 2

All models of URETERO-RENO FIBERSCOPE (6)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Olympus ureteroscope is a thin flexible scope used to look inside the urinary tract. Many of its 188 reports describe breaks, detaching parts, or peeling. There are 7 injury reports and no deaths. There are 2 open recalls, one about wet channels after repair and one about focus ring adjustment. Reports rose from 16 in the year before to 121 in the latest year.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
190
Deaths reported
0
Injuries reported
7
Recalls naming it
2

Summary

The FDA's MAUDE database contains 190 adverse event reports that list this exact device (DI 04953170340833), received between Jun 6, 2018 and Sep 25, 2026 — classified as 7 injuries, 183 malfunctions.

The most frequently reported specific device problem is “Break” (34 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 2 reportsSep 2024: 2 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 4 reportsSep 2025: 2 reportsOct 2025: 5 reportsNov 2025: 2 reportsDec 2025: 9 reportsJan 2026: 13 reportsFeb 2026: 13 reportsMar 2026: 9 reportsApr 2026: 12 reportsMay 2026: 23 reportsJun 2026: 8 reportsJul 2026: 11 reportsAug 2026: 18 reportsSep 2026: 5 reportsOct 2026: 0 reports20242025202623
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 25, 2023Class IIOpen, Classified

    Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

    Z-0347-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Apr 20, 2022Class IIOpen, Classified

    Incorrect adhesive used to secure the diopter ring and nameplate ring. Improperly connected diopter ring may result in the loss of adjustment function leading to a blurred field of vision and unintentional contact with the patients body and in rare cases perforation in the urinary tract/renal pelvis.

    Z-1264-2022 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break34
  • Detachment of Device or Device Component34
  • Material Fragmentation14
  • Degraded12
  • Corroded11
  • Failure to Clean Adequately11
  • Peeled/Delaminated11
  • Device Reprocessing Problem9

Patient problems

  • No Clinical Signs, Symptoms or Conditions171
  • No Patient Involvement6
  • No Consequences Or Impact To Patient4
  • No Known Impact Or Consequence To Patient3
  • Foreign Body In Patient2
  • Hepatitis2
  • Abrasion1
  • Cramp(s) /Muscle Spasm(s)1

Recent adverse event reports

The 25 most recent of 190 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 25, 2026Environmental Particulates

    The uretero-reno fiberscope was returned to the service center for a separate issue. During evaluation of the device, dirt/foreign material was found on the objective lens, likely due to environmental particulates, and was repaired. There was no patient involvement, and no harm was reported as a result of the event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50893094 ↗
  • MalfunctionSep 24, 2026Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50890059 ↗
  • MalfunctionSep 23, 2026Material Fragmentation

    The customer returned the Uretero-Reno Fiberscope to the service center for evaluation related to a separate issue. During pre-investigation/equipment intake, the service repair technician identified the bending section A-rubber glue was chipped. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50870632 ↗
  • MalfunctionSep 4, 2026Thermal Decomposition of Device; Detachment of Device or Device Component

    The uretero-reno fiberscope was returned to the service center for a separate issue. During evaluation of the device, a piece of the distal end plastic C-body was found to be missing, likely due to a high energy device thermal burn, and was replace. There was no patient involvement, and no harm was reported as a result of the event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50621196 ↗
  • MalfunctionSep 1, 2026Material Fragmentation

    The customer reported the Uretero-Reno Videoscope had a separate issue. During the initial evaluation of the device, it was observed that the adhesive on the bending section was chipped. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50536346 ↗
  • MalfunctionAug 24, 2026Corroded

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50421783 ↗
  • MalfunctionAug 24, 2026Material Puncture/Hole

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50419767 ↗
  • MalfunctionAug 20, 2026Failure to Clean Adequately

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50359292 ↗
  • MalfunctionAug 20, 2026Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50351412 ↗
  • MalfunctionAug 18, 2026Material Fragmentation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50291573 ↗
  • MalfunctionAug 13, 2026Material Fragmentation; Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50242890 ↗
  • MalfunctionAug 11, 2026Environmental Particulates

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50214023 ↗
  • MalfunctionAug 11, 2026Contamination

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50210482 ↗
  • MalfunctionAug 11, 2026Degraded

    The customer returned the uretero-reno fiberscope to the service center for a separate issue. During the device analysis, the service repair technician indicated a chipped adhesive at the bending rubber was found. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50204228 ↗
  • MalfunctionAug 10, 2026Break

    The customer reported at scope tip, the cover for the lens was broken into pieces during reprocessing involving an Olympus Uretero-Reno Fiberscope. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50196559 ↗
  • MalfunctionAug 10, 2026Degraded

    The customer returned the Uretero-Reno Fiberscope to the service center for a separate issue. During the device analysis, the service repair technician indicated adhesive on the bending section cover/distal sheath has a chip. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50196323 ↗
  • MalfunctionAug 7, 2026Loss of or Failure to Bond

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50175678 ↗
  • MalfunctionAug 6, 2026Material Deformation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50166771 ↗
  • MalfunctionAug 6, 2026Material Fragmentation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50161696 ↗
  • MalfunctionAug 5, 2026Loss of or Failure to Bond

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50145849 ↗
  • MalfunctionAug 5, 2026Material Deformation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50145297 ↗
  • MalfunctionAug 4, 2026Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50134113 ↗
  • MalfunctionAug 4, 2026Detachment of Device or Device Component

    The customer returned the Uretero-Reno Fiberscope to the service center for a separate issue. During the device analysis, the service repair technician indicates that the a-rubber glue is chipped. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50129356 ↗
  • MalfunctionJul 27, 2026Material Fragmentation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50007763 ↗
  • MalfunctionJul 21, 2026Material Fragmentation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25888278 ↗

Other Ureteroscope and accessories, flexible/rigid devices

All Ureteroscope and accessories, flexible/rigid devices →