Balloon2 MH-303 (UDI 04953170355721)Recalls & reported problems
Olympus Medical Systems Corp. · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 04953170355721), received between Feb 3, 2023 and Oct 1, 2024 — classified as 12 malfunctions.
The most frequently reported specific device problem is “Device Reprocessing Problem” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Device Reprocessing Problem5
- Failure to Clean Adequately3
- Problem with Sterilization3
- Device Markings/Labelling Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions12
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 1, 2024Failure to Clean Adequately; Failure to Clean Adequately
IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE SUBJECT DEVICE EXHIBITED FOREIGN MATERIAL ON THE BALLOON AND MOISTURE CAUSED RUST TO FORM ON THE OXYGEN ABSORBER. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20343118 ↗MalfunctionDec 20, 2023Problem with Sterilization; Problem with Sterilization
AN OLYMPUS REPRESENTATIVE REPORTED TO OLYMPUS ON BEHALF OF THE CUSTOMER THAT THE BALLOON WAS USED ON PATIENTS WITHOUT THE GAS STERILIZATION PRIOR TO BEING USED. IT WAS USED ON MULTIPLE PATIENTS FROM MAY TO NOVEMBER OF THIS YEAR (2023). THE BALLOON WAS USED DURING A DIAGNOSTIC ULTRASOUND ENDOSCOPY PROCEDURE. THE PROCEDURE WAS COMPLETED WITH THE SAME DEVICE. THERE WERE NO REPORTS OF PATIENT HARM OR HEALTH DAMAGE. THIS REPORT IS RELATED TO LINKED PATIENT IDENTIFIERS: (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18370147 ↗MalfunctionDec 11, 2023Problem with Sterilization; Problem with Sterilization
THE CUSTOMER REPORTED TO OLYMPUS THAT THE BALLOON WAS USED DURING A DIAGNOSTIC ULTRASOUND ENDOSCOPY PROCEDURE. THE CUSTOMER HAD NOT PERFORMED GAS STERILIZATION OF THE BALLOON PRIOR TO USE. THE PROCEDURE WAS COMPLETED WITH THE SAME DEVICE. THERE WERE NO REPORTS OF PATIENT HARM OR HEALTH DAMAGE. THIS REPORT IS RELATED TO LINKED PATIENT IDENTIFIERS (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18301853 ↗MalfunctionDec 6, 2023Device Reprocessing Problem; Device Reprocessing Problem
THE CUSTOMER REPORTED TO OLYMPUS, THAT AN UNSTERILIZED BALLOON 2 WAS USED IN A PATIENT. THE INTENDED PROCEDURE WAS COMPLETED USING THE SAME DEVICE. THERE WERE NO REPORTS OF HARM TO PATIENTS. THE DISTRIBUTOR CHECKED WITH CUSTOMER INFORMATION CENTER TO SEE IF BALLOON 2 COULD BE STERILIZED WITH STELLAD. IT WAS DISCOVERED THAT THE FACILITY WAS UNABLE TO HANDLE GAS STERILIZATION, AND THAT AN UNSTERILIZED ULTRASONIC BALLOON WAS MISUSED. THE BALLOON HAD NOT BEEN USED SINCE THE DISCOVERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18273606 ↗MalfunctionNov 28, 2023Device Reprocessing Problem; Device Reprocessing Problem
THE CUSTOMER REPORTED TO OLYMPUS THAT THE BALLOON WAS USED DURING A DIAGNOSTIC ULTRASOUND ENDOSCOPY PROCEDURE WITHOUT GAS STERILIZATION. THE PROCEDURE WAS COMPLETED WITH THE SAME DEVICE. THERE WERE NO REPORTS OF PATIENT HARM OR HEALTH DAMAGE. RELATED PATIENT IDENTIFIERS: (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18222432 ↗MalfunctionSep 21, 2023Problem with Sterilization; Problem with Sterilization
THE CUSTOMER REPORTED TO OLYMPUS, THE BALLOON2 MAY HAVE BEEN USED NON-STERILE IN A PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM. THIS REPORT IS RELATED TO LINKED PATIENT IDENTIFIER (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17789353 ↗MalfunctionSep 19, 2023Failure to Clean Adequately; Failure to Clean Adequately
CIC SAFETY NOTIFIED OLYMPUS BY MAIL THAT DURING A DIAGNOSTIC ULTRASOUND ENDOSCOPE PROCEDURE, AN UNSTERILIZED BALLOON 2 WAS USED ON A PATIENT. THERE WERE NO REPORTS OF PATIENT HARM. RELATED TO PATIENT IDENTIFIER: (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17777555 ↗MalfunctionSep 1, 2023Device Reprocessing Problem; Device Reprocessing Problem
THE CUSTOMER REPORTED THAT HIS OLYMPUS BALLOON2 MAY HAVE INADVERTENTLY BEEN USED ON A PATIENT AFTER HAVING BEEN INSUFFICIENTLY REPROCESSED. THERE HAS BEEN NO PATIENT HARM REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17669854 ↗MalfunctionAug 30, 2023Device Reprocessing Problem; Device Reprocessing Problem
A USER FACILITY REPORTED TO OLYMPUS THE POSSIBLE USE OF NON-STERILE BALLOON2 ON A PATIENT. THERE WERE NO REPORTS OF PATIENT HARM FOR THE UNIDENTIFIED DIAGNOSTIC PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17653342 ↗MalfunctionAug 28, 2023Failure to Clean Adequately; Failure to Clean Adequately
THE CUSTOMER REPORTED TO OLYMPUS THAT A NON-STERILE BALLOON2 WAS POSSIBLY USED ON A PATIENT. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17631708 ↗MalfunctionAug 16, 2023Device Reprocessing Problem; Device Reprocessing Problem
THE CUSTOMER REPORTED THAT AN OLYMPUS BALLOON MAY HAVE BEEN INSUFFICIENTLY REPROCESSED AND USED DURING A DIAGNOSTIC ULTRASONIC ENDOSCOPY PROCEDURE. THERE WAS NO PATIENT HARM REPORTED AS A RESULT OF THIS EVENT. THIS REPORT IS RELATED TO REPORT WITH PATIENT IDENTIFIER (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17556917 ↗MalfunctionFeb 3, 2023Device Markings/Labelling Problem; Device Markings/Labelling Problem
THE CUSTOMER REPORTED THAT, WHEN THE FACILITY CHECKED THE PACKAGE INSERT, THE PACKAGE INSERT FOR THE SUBJECT DEVICE STATES THAT STERILIZATION IS REQUIRED, BUT THE PRODUCT PACKAGE LABEL STATES STERILIZE AS NEEDED. THIS WAS THE FIRST TIME THE DISCREPANCY WAS NOTED, AND THE FACILITY HAD PREVIOUSLY USED THE PRODUCT WITHOUT STERILIZATION. THERE HAVE BEEN NO REPORTS OF INFECTION AND DOCTORS HAVE NOT SEEN IT AS A PROBLEM, BUT THE CUSTOMER HAS COMPLAINED THAT THERE MAY BE A PROBLEM WITH THE WAY OLYMPUS WRITES IT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16293787 ↗
Other Colonoscope and accessories, flexible/rigid devices
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS CF-HQ190LOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls2 deaths6,804 reports
- EVIS EXERA III XENON LIGHT SOURCE OLYMPUS CLV-190Olympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid6,617 reports
- EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-HQ190Olympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls5 deaths6,244 reports
- EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-H190Olympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid2 recalls7 deaths5,572 reports
- EVIS EXERA III COLONOVIDEOSCOPE OLYMPUS PCF-H190DLOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls2,830 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS PCF-H190LOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls2 deaths2,513 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS CF-H190IOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid1 recall1,967 reports
- PENTAX Medical EC38-i10cLHoya Corporation · Colonoscope and accessories, flexible/rigid1,618 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page