VIDEO SYSTEM CENTER OLYMPUS CV-V1 (UDI 04953170356384)Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of VIDEO SYSTEM CENTER (4)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Olympus video system center is the imaging unit used with endoscopes. Reports describe no image, failure to power up, poor image quality and front panel buttons or lights not working, with many found during setup or maintenance. There are 111 reports with no deaths and no injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
111
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 111 adverse event reports that list this exact device (DI 04953170356384), received between Nov 26, 2020 and Mar 14, 2025 — classified as 111 malfunctions.

The most frequently reported specific device problem is “No Display/Image” (38 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 5 reportsDec 2023: 4 reportsJan 2024: 5 reportsFeb 2024: 6 reportsMar 2024: 4 reportsApr 2024: 4 reportsMay 2024: 4 reportsJun 2024: 1 reportsJul 2024: 3 reportsAug 2024: 5 reportsSep 2024: 4 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 1 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • No Display/Image38
  • Failure to Power Up18
  • Poor Quality Image11
  • Display or Visual Feedback Problem10
  • Erratic or Intermittent Display8
  • Defective Component7
  • No Device Output5
  • Output Problem5

Patient problems

  • No Clinical Signs, Symptoms or Conditions110
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 25 most recent of 111 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 14, 2025Failure to Power Up; Failure to Power Up

    IT WAS REPORTED THAT THE SUBJECT DEVICE WAS NOT WORKING. THIS WAS DISCOVERED DURING SET UP / INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21608722 ↗
  • MalfunctionMar 12, 2025No Display/Image; Connection Problem; No Display/Image; Connection Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE EXHIBITED MALFUNCTION OF THE SOCKET AND NO IMAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21586117 ↗
  • MalfunctionFeb 19, 2025No Device Output; No Device Output

    IT WAS REPORTED THAT, THE VIDEO SYSTEM CENTER FRONT PANEL WAS NOT WORKING. THE ISSUE WAS FOUND DURING MAINTENANCE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21413395 ↗
  • MalfunctionJan 21, 2025Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working

    IT WAS REPORTED THAT THE SUBJECT DEVICE PUSH BUTTON WAS NOT WORKING PROPERLY. THE ISSUE WAS FOUND DURING MAINTENANCE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21198249 ↗
  • MalfunctionOct 25, 2024No Display/Image; No Display/Image

    IT WAS REPORTED THE VIDEO SYSTEM CENTER HAD NO IMAGE. THE ISSUE OCCURRED DURING PREPARATION FOR USE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20536332 ↗
  • MalfunctionOct 1, 2024No Display/Image; No Display/Image

    IT WAS REPORTED THAT THE SUBJECT DEVICE HAD NO IMAGE. THE ISSUE OCCURRED DURING A DIAGNOSTIC GASTROSCOPY. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20344030 ↗
  • MalfunctionSep 26, 2024No Device Output; No Device Output

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE VIDEO SYSTEM CENTER EXHIBITED LED NOT GLOWING FROM FRONT PANEL. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20314165 ↗
  • MalfunctionSep 24, 2024No Display/Image; No Display/Image

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE VIDEO SYSTEM CENTER EXHIBITED NO IMAGE WHEN SHAKING DUE TO DEFECTIVE VIDEO CONNECTOR SOCKET. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20293473 ↗
  • MalfunctionSep 18, 2024Loose or Intermittent Connection; Loose or Intermittent Connection

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THE VIDEO SYSTEM CENTER FRONT PANEL LIGHT-EMITTING DIODE (LED) TURNED OFF AFTER SOME TIME. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20246140 ↗
  • MalfunctionSep 16, 2024Poor Quality Image; Poor Quality Image

    IT WAS REPORTED, THE VIDEO SYSTEM CENTER HAD INTERFACING PROBLEMS CAUSING NOISY IMAGE. THE ISSUE OCCURRED DURING PREPARATION FOR AN UNSPECIFIED PROCEDURE. THE PATIENT WAS NOT UNDER ANESTHESIA, THERE WAS NO DELAY AND THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20224279 ↗
  • MalfunctionAug 23, 2024No Display/Image; No Display/Image

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE VIDEO SYSTEM CENTER EXHIBITED NO IMAGE WHEN SHAKING THE CONNECTOR DUE TO A DEFECTIVE VIDEO CONNECTOR SOCKET. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20048574 ↗
  • MalfunctionAug 13, 2024Failure to Power Up; Failure to Power Up

    IT WAS REPORTED THAT THE VIDEO PROCESSOR WAS NOT TURNING ON. THE ISSUE OCCURRED DURING MAINTENANCE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19978456 ↗
  • MalfunctionAug 13, 2024No Display/Image; No Display/Image

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE VIDEO SYSTEM CENTER EXHIBITED NO IMAGE DUE TO FAULTY SCOPE SOCKET. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19977772 ↗
  • MalfunctionAug 13, 2024Erratic or Intermittent Display; Erratic or Intermittent Display

    IT WAS REPORTED, THAT THE VIDEO SYSTEM CENTER HAD EXHIBITED AN IMAGE ISSUE, WHICH RENDERED THE ENDOSCOPIC IMAGE NON-VISIBLE. THE ISSUE WAS OBSERVED DURING MAINTENANCE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19973455 ↗
  • MalfunctionAug 12, 2024No Display/Image; No Device Output; No Display/Image; No Device Output

    IT WAS REPORTED THE VIDEO SYSTEM CENTER HAD NO IMAGE, FRONT PANEL FLICKERS AND BUTTONS NOT RESPONDING. THE ISSUE OCCURRED DURING PREPARATION FOR USE BEFORE AN UNESPECIFIED PROCEDURE. THERE WAS NO DELAY AND PATIENT WAS NOT UNDER ANESTHESIA. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19962520 ↗
  • MalfunctionJul 16, 2024Loose or Intermittent Connection; Loose or Intermittent Connection

    IT WAS REPORTED THE VIDEO SYSTEM CENTER CONNECTOR WAS LOOSE, THEN THE IMAGE WOULD BECOME BLACK WHEN SHAKING THE SCOPE CONNECTOR SLIGHTLY. THE ISSUE OCCURRED DURING PREPARATION FOR USE FOR A GASTROSCOPY. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THE PATIENT WAS NOT UNDER ANESTHESIA AND THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19753749 ↗
  • MalfunctionJul 3, 2024Loss of Power; Sparking; Loss of Power; Sparking

    IT WAS REPORTED, THE VIDEO SYSTEM CENTER STOPPED WORKING AFTER A SPARK. THE ISSUE OCCURRED DURING MAINTENANCE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19667146 ↗
  • MalfunctionJul 1, 2024Loss of Power; Loss of Power

    IT WAS REPORTED THAT THE VIDEO SYSTEM CENTER PROCESSOR FRONT PANEL BUTTON GETS TURNED OFF AUTOMATICALLY. THE EVENT OCCURRED DURING MAINTENANCE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19650071 ↗
  • MalfunctionJun 4, 2024Poor Quality Image; Poor Quality Image

    IT WAS REPORTED THAT THE VIDEO SYSTEM CENTER EXHIBITED A NOISY IMAGE CAUSED BY A FAULTY PRINTED CIRCUIT BOARD AND THERE WAS ADDITIONAL INTERFERENCE IN THE IMAGE DUE TO A DEFECTIVE VIDEO CONNECTOR SOCKET. IT IS UNKNOWN WHEN THE ISSUE WAS FOUND. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19464428 ↗
  • MalfunctionMay 29, 2024Failure to Power Up; Failure to Power Up

    IT WAS REPORTED THAT THE VIDEO SYSTEM CENTER WOULD NOT POWER ON. THE ISSUE OCCURRED DURING MAINTENANCE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19414976 ↗
  • MalfunctionMay 16, 2024No Display/Image; Connection Problem; No Display/Image; Connection Problem

    IT WAS REPORTED THE VIDEO SYSTEM CENTER HAD NO IMAGE ON MONITOR, VGA PORT NOT WORKING, AND THE VIDEO CONNECTOR CANNOT BE CONNECTED. THE ISSUE OCCURRED DURING PREPARATION FOR USE FOR A DIAGNOSTIC GASTROSCOPY THAT WAS CANCELLED WITHOUT DELAY. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19333080 ↗
  • MalfunctionMay 7, 2024No Display/Image; No Display/Image

    IT WAS REPORTED THE VIDEO SYSTEM CENTER HAD A FLICKERING IMAGE. THE ISSUE OCCURRED DURING PREPARATION FOR USE ON A DIAGNOSTIC (COLONOSCOPY) PROCEDURE. THE PROCEDURE WAS CANCELLED AS THE FACILITY DID NOT HAVE ANOTHER DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19259751 ↗
  • MalfunctionMay 6, 2024Failure to Power Up; Failure to Power Up

    IT WAS OBSERVED DURING THE DEVICE INSPECTION THAT THE FRONT PANEL SWITCHES OF THE VIDEO SYSTEM CENTER WERE NOT RESPONDING DUE TO A FAULTY FRONT PANEL ASSEMBLY. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19247786 ↗
  • MalfunctionApr 23, 2024No Display/Image; No Display/Image

    IT WAS REPORTED, VIDEO SYSTEM CENTER EXHIBITED THAT THE IMAGE DISAPPEARS, AND ONLY GREEN BARS WERE VISIBLE. THE ISSUE OCCURRED DURING PREPARATION FOR A DIAGNOSTIC GASTROSCOPIC PROCEDURE THAT WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19166535 ↗
  • MalfunctionApr 8, 2024No Display/Image; Failure to Power Up; No Display/Image; Failure to Power Up

    IT WAS REPORTED THE VIDEO SYSTEM CENTER HAD NO O/P (OUTPUT) SIGNAL. THE ISSUE WAS FOUND AT MAINTENANCE. THERE WERE NO REPORTS OF A PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19068726 ↗

Other Colonoscope and accessories, flexible/rigid devices

All Colonoscope and accessories, flexible/rigid devices →